A method to derive appropriate exposure parameters from target exposure index and patient thickness in pediatric digital radiography

被引:11
|
作者
Zhang, Menglong [1 ]
Liu, Kai [1 ]
Niu, Xuecai [2 ]
Liu, Xinli [2 ]
机构
[1] Shandong Med Imaging Res Inst, Dept Radiog, Jinan 250021, Shangdong, Peoples R China
[2] Fourth Hosp Jinan, Dept Radiat Oncol, Jinan, Shandong, Peoples R China
关键词
Radiation dose; Pediatrics; Digital radiography; Exposure index; Radiation protection; COMPUTED RADIOGRAPHY; IMAGE QUALITY; X-RAY; OPTIMIZATION; CHILDREN; CHEST;
D O I
10.1007/s00247-012-2555-3
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Little information exists concerning appropriate exposure and measuring overall patient dose in pediatric digital radiography. To establish a convenient method of appropriate exposure from target exposure index (EI) and thickness in pediatric digital radiography and estimate patient entrance-surface dose (ESD) and dose-area product (DAP) associated with chest, abdomen and pelvis radiography. A formula was deduced to calculate appropriate mAs changed with children's weight and height. EI was used to control image quality. With this formula, dose-optimized procedures were carried out. Data were collected from 180 pediatric examinations, including chest, abdomen and pelvis anterior-posterior (AP) projections. The children were divided into the following age bands: newborns (0-28 days), infants (28 days-2 years) and older children (2-7 years). In each age band, ten children were exposed with the calculated appropriate mAs and EI values were kept steady in appropriate range (referred as target group) and ten children were exposed to the factors routinely used in practice (referred to as the routine group). DAP to children was measured with a DAP meter, and ESD was calculated using measured DAP and data from the National Radiological Protection Board. Data were compared between groups. ESD ranges in the target group were 32-202 mu Gy (chest AP), 57-333 mu Gy (abdomen AP) and 52-372 mu Gy (pelvis AP). For every radiographic procedure, chi-square Student's t tests showed a significant difference in average ESD and DAP between the two groups (P < 0.005). Most ESD values from the routine group were two times higher than those from the target group. The study established a convenient method to set appropriate exposure parameters (mAs) to reach a target EI using the child's weight and height in pediatric radiography. By this method, ESD and DAP can be significantly reduced in children.
引用
收藏
页码:568 / 574
页数:7
相关论文
共 50 条
  • [41] Using Exposure Index to Develop Thickness-Based Techniques for the Skull Views of a Pediatric Skeletal Survey
    Li, X.
    Hulme, K.
    Park, E.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [42] Comparison of organ dose from chest radiography with varying beam quality and constant exposure index
    Nakanishi, Kohei
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2023, 46 (03) : 1143 - 1151
  • [43] Comparison of organ dose from chest radiography with varying beam quality and constant exposure index
    Kohei Nakanishi
    Physical and Engineering Sciences in Medicine, 2023, 46 : 1143 - 1151
  • [44] Effect of Lead Shielding on Patient Dose in Digital Radiography with Automatic Exposure Control (AEC) Driven Imaging Systems
    Gautam, A.
    Kroger, L.
    Barton, K.
    Hodge, B.
    Seibert, J.
    MEDICAL PHYSICS, 2020, 47 (06) : E258 - E258
  • [45] A Method for Deriving Exam-Specific Target Exposure Indices (TEI) in Computed Radiography as a Function of a Reference TEI
    Hulme, K.
    Kubicki, V.
    Dong, F.
    Davros, W.
    Johnson, P.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [46] Implementation of exposure index for optimize image quality and patient dose estimation with computed radiography (a clinical study of adult posteroanterior chest and anteroposterior abdomen radiography)
    Rochmayanti, Dwi
    Wibowo, Gatot Murti
    Fatimah
    Setiawan, Agung Nugroho
    9TH INTERNATIONAL CONFERENCE ON PHYSICS AND ITS APPLICATIONS (ICOPIA), 2019, 1153
  • [47] RADIATION EXPOSURE TO THE PEDIATRIC-PATIENT FROM CARDIAC-CATHETERIZATION AND ANGIOCARDIOGRAPHY
    MARTIN, EC
    OLSON, A
    BRITISH JOURNAL OF RADIOLOGY, 1980, 53 (626): : 100 - 106
  • [48] Investigation of State-Of-Practice Using Exposure Index (EI) Data Following Digital Radiography (DR) Retrofits of Eight Radiographic Units
    Hulme, K.
    Scilla, K.
    Fisher, R.
    Li, X.
    MEDICAL PHYSICS, 2018, 45 (06) : E694 - E694
  • [49] Fabrication of thin film x-ray sensor using particle-in-binder method for automatic exposure control in digital radiography
    Oh, K.
    Kim, G.
    Park, J.
    Song, Y.
    Heo, S.
    Cho, S.
    Kim, J.
    Park, S.
    Nam, S.
    JOURNAL OF INSTRUMENTATION, 2015, 10
  • [50] Microscopic Studies of the Influence of Main Exposure Time on Parameters of Flexographic Printing Plate Produced by Digital Thermal Method
    Harri, Liliya
    MICROSCOPY RESEARCH AND TECHNIQUE, 2009, 72 (10) : 707 - 716