摘要
Flow rate estimation for ventricular assist devices without additional flow sensors can improve the quality of life of patients. In this article, a novel flow estimation method using the passively stabilized displacement of a magnetically levitated impeller is developed to achieve sufficient accuracy and precision of flow estimation for ventricular assist devices in a simple manner. The magnetically levitated impeller used is axially suspended by a magnetic bearing in a centrifugal blood pump that has been developed by our group. The radial displacement of the impeller, which is restricted by passive stability, can be correlated with the flow rate because the radial hydraulic force on the impeller varies according to the flow rate. To obtain the correlation with various blood viscosities, the relationships between the radial displacements of the magnetically levitated impeller and the pressure head-flow rate characteristics of the pump were determined simultaneously using aqueous solutions of glycerol with a potential blood viscosity range. The measurement results showed that accurate steady flow rates could be estimated with a coefficient of determination of approximately 0.97 and mean absolute error of approximately 0.22 L/min without fluid viscosity measurements by using the relationships between the impeller displacement and the flow rate. Moreover, a precision of approximately 0.01 (L/min)/mu m was obtained owing to a strong estimation indicator signal provided by the large displacement of the passively stabilized impeller; thus, the proposed estimation method can help ensure sufficient accuracy and precision for ventricular assist devices in a simple manner, even if the blood viscosity is unknown.
引用
收藏
页码:291 / 298
页数:8
相关论文
共 50 条
  • [41] Visualization of main and leakage flow in magnetically suspended centrifugal blood pump
    Masaaki Horie
    Keiichi Yamamura
    Journal of Visualization, 2012, 15 : 353 - 361
  • [42] Numerical investigation on the effect of impeller axial position on hemodynamics of an extracorporeal centrifugal blood pump
    Lv, Shen
    He, Zhi-Peng
    Liu, Guang-Mao
    Hu, Sheng-Shou
    COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2024, 27 (13) : 1744 - 1755
  • [43] Visualization of main and leakage flow in magnetically suspended centrifugal blood pump
    Horie, Masaaki
    Yamamura, Keiichi
    JOURNAL OF VISUALIZATION, 2012, 15 (04) : 353 - 361
  • [44] Improvement of washout flow in a centrifugal blood pump by a semi-open impeller
    Tsukiya, T.
    Taenaka, Y.
    Tatsumi, E.
    Takano, H.
    ASAIO Journal, 2000, 46 (02)
  • [45] Establishment of flow estimation for an implantable centrifugal blood pump
    Wakisaka, Y
    Okuzono, Y
    Taenaka, Y
    Chikanari, K
    Masuzawa, T
    Takano, H
    ASAIO JOURNAL, 1997, 43 (05) : M659 - M662
  • [46] Improvement of washout flow in a centrifugal blood pump by a semi-open impeller
    Tsukiya, T
    Taenaka, Y
    Tatsumi, E
    Takano, H
    ASAIO JOURNAL, 2002, 48 (01) : 76 - 82
  • [47] Effect of a bearing gap on hemolytic property in a hydrodynamically levitated centrifugal blood pump with a semi-open impeller
    Kosaka, Ryo
    Nishida, Masahiro
    Maruyama, Osamu
    Yambe, Tomoyuki
    Imachi, Kou
    Yamane, Takashi
    BIO-MEDICAL MATERIALS AND ENGINEERING, 2013, 23 (1-2) : 37 - 47
  • [48] A Low Cost Linear Position Measurement System for Magnetically Levitated Rotor in Axial Flow Pump
    Yang, Sheng-Ming
    Huang, Chien-Lung
    2008 IEEE INDUSTRY APPLICATIONS SOCIETY ANNUAL MEETING, VOLS 1-5, 2008, : 2519 - 2523
  • [49] Flow Rate Estimation of a Centrifugal Blood Pump Using the Balance of Control Currents Through the Electromagnets in the Magnetic Bearing
    Shida, Shuya
    Masuzawa, Toru
    Osa, Masahiro
    Suzuki, Yutaka
    ADVANCED BIOMEDICAL ENGINEERING, 2023, 12 : 244 - 252
  • [50] Measurement of blood hematocrit inside the magnetically suspended centrifugal pump using an optical technique: Application to assessment of pump flow
    Kaiwa, T
    Mori, T
    Kijima, T
    Nogawa, M
    Nojiri, C
    Takatani, S
    ARTIFICIAL ORGANS, 1999, 23 (06) : 490 - 495