摘要
Biodegradable stent coatings were recently introduced as a potential solution to overcome sustained inflammatory responses observed with permanent polymer-based drug-eluting stents. In a preliminary study, selected biodegradable or permanent polymer-based sirolimus-eluting stent (SES) formulations were screened for effectiveness in comparison to bare metal stents (BMS) at 28 days. Subsequently, the most favourable SES formulation was compared to commercially available SES (Cypher (TM)) at 28,90 and 180 days to investigate the histopathologic response as well as tissue, blood and organ pharmacokinetics. Overlapping SES implantation was conducted to evaluate vascular healing at 28 days in this particular setting. SES with biodegradable poly (L-lactide) polymer (PLLA) or poly(lactide-co-glycolide) showed the most favourable outcome with regards to reductions in neointimal area in comparison to BMS at 28 days. The PLLA SES showed a similar reduction in neointimal area compared to Cypher (TM) at 28 days, with significant greater reductions at 90 and 180 days (1.7 +/- 0.7 mm(2) vs. 3.1 +/- 1.5 mm(2), p=0.03 and 1.8 +/- 1.2 mm(2) vs. 3.0 +/- 1.5 mm(2), p=0.01, respectively). Sirolimus vascular tissue concentrations were detectable up to 90 days following implantation. Overlapping stented segments showed favourable histopathologic results with respect to fibrin deposition and endothelialisation at 28 days. In conclusion, the use of PLLA as drug-eluting matrix resulted in mild inflammatory responses in the presence of effective sirolimus tissue concentrations. The greater efficacy observed at long-term follow-up in PLLA SES compared to Cypher (TM) may be a multifactorial result of stent design, polymer biocompatibility and improved release kinetics.
引用
收藏
页码:1161 / 1171
页数:11
相关论文
共 50 条
  • [21] SYNERGY Biodegradable Polymer Everolimus Eluting Coronary Stent: Porcine Vascular Compatibility and Polymer Safety Study
    Wilson, Gregory
    Marks, Angela
    Eppihimer, Michael
    Knapp, David
    Pennington, Doug
    Huibregtse, Barbara
    Dawkins, Keith
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2012, 60 (17) : B164 - B164
  • [22] The SYNERGY biodegradable polymer everolimus eluting coronary stent: Porcine vascular compatibility and polymer safety study
    Wilson, Gregory J.
    Marks, Angela
    Berg, Kimberly J.
    Eppihimer, Michael
    Sushkova, Natalia
    Hawley, Steve P.
    Robertson, Kimberly A.
    Knapp, David
    Pennington, Douglas E.
    Chen, Yen-Lane
    Foss, Aaron
    Huibregtse, Barbara
    Dawkins, Keith D.
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2015, 86 (06) : E247 - E257
  • [23] COMPARISONS OF SAFETY PROFILE BETWEEN DURABLE AND BIODEGRADABLE POLYMER BASED DRUG ELUTING STENTS IN PORCINE CORONARY ARTERY MODEL
    Ijichi, Takeshi
    Nakazawa, Gaku
    Shinke, Toshiro
    Matsumoto, Daisuke
    Otake, Hiromasa
    Torii, Sho
    Yoshikawa, Ayako
    Shite, Junya
    Ikari, Yuji
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2012, 59 (13) : E315 - E315
  • [24] Long-Term Effects of Novel Biodegradable, Polymer-Coated, Sirolimus-Eluting Stents on Neointimal Formation in a Porcine Coronary Model
    Peng, Hong-Yu
    Chen, Ming
    Zheng, Bo
    Wang, Xin-Gang
    Huo, Yong
    INTERNATIONAL HEART JOURNAL, 2009, 50 (06) : 811 - 822
  • [25] Safety and Efficacy of New Biodegradable Polymer-based Sirolimus-Eluting Stents in a Preclinical Model
    Perez de Prado, Armando
    Perez Martinez, Claudia
    Cuellas Ramon, Carlos
    Regueiro Purrinos, Marta
    Lopez Benito, Maria
    Gonzalo Orden, Jose Manuel
    Rodriguez Altonaga, Jose Antonio
    Estevez Loureiro, Rodrigo
    Benito Gonzalez, Tomas
    Vinuela Baragano, David
    Molina Crisol, Maria
    Amoros Aguilar, Marc
    Perez Serranos, Isabel
    Vidal Parreu, Arnau
    Benavides Montegordo, Ana
    Duocastella Codina, Luis
    Fernandez Vazquez, Felipe
    REVISTA ESPANOLA DE CARDIOLOGIA, 2017, 70 (12): : 1059 - 1066
  • [26] The twelve-month outcomes of a biolimus eluting stent with a biodegradable polymer compared with a sirolimus eluting stent with a durable polymer
    Garg, Scot
    Sarno, Giovanna
    Serruys, Patrick W.
    de Vries, Ton
    Buszman, Pawel
    Linke, Axel
    Ischinger, Thomas
    Klauss, Volker
    Eberli, Franz
    Corti, Roberto
    Wijns, William
    Morice, Marie-Claude
    Di Mario, Carlo
    van Geuns, Robert Jan
    Eerdmans, Pedro
    van Es, Gerrit-Anne
    Meier, Bernhard
    Jueni, Peter
    Windecker, Stephan
    EUROINTERVENTION, 2010, 6 (02) : 233 - 239
  • [27] Comparisons of the outcomes of the two drug eluting stents in small vessel: abluminal biolimus-eluting biodegradable polymer stent and zotarolimus-eluting permanent polymer stent
    Yoon, Y. W.
    Yoon, J. H.
    Chung, H. M.
    Hong, B. K.
    Lee, B. K.
    Kwon, H. M.
    Choi, E. Y.
    EUROPEAN HEART JOURNAL, 2013, 34 : 877 - 877
  • [29] Comparison of abluminal biodegradable polymer biolimus-eluting stents and durable polymer everolimus-eluting stents in the treatment of coronary bifurcations
    Costopoulos, Charis
    Latib, Azeem
    Naganuma, Toru
    Sticchi, Alessandro
    Ferrarello, Santo
    Regazzoli, Damiano
    Chieffo, Alaide
    Figini, Filippo
    Carlino, Mauro
    Montorfano, Matteo
    Naim, Charbel
    Kawaguchi, Masanori
    Gerasimou, Argyrios
    Giannini, Francesco
    Godino, Cosmo
    Colombo, Antonio
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2014, 83 (06) : 889 - 895
  • [30] Unselected use of ultrathin strut biodegradable polymer sirolimus-eluting stent versus durable polymer everolimus-eluting stent for coronary revascularization
    Yamaji, Kyohei
    Zanchin, Thomas
    Zanchin, Christian
    Stortecky, Stefan
    Pilgrim, Thomas
    Heg, Dik
    Windecker, Stephan
    Raber, Lorenz
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2018, 72 (13) : B265 - B266