Organ dose in CT: Comparison between measurements and computational methods

被引:2
|
作者
Pace, Martina [1 ]
Bonanno, Elisa [1 ]
Borzi, Giuseppina Rita [1 ]
Cavalli, Nina [1 ]
D'Anna, Alessia [2 ]
Gueli, Anna Maria [2 ]
Stella, Giuseppe [2 ]
Zirone, Lucia [1 ,2 ]
Marino, Camelo [1 ]
机构
[1] Human Ist Clin Catanese, Med Phys Dept, Misterbianco, CT, Italy
[2] Univ Catania, Dept Phys & Astron E Majorana, Catania, Italy
来源
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS | 2023年 / 112卷
关键词
DoseWatch TM; EBT-3; XR-QA2; Patient dosimetry; Diagnostic radiology; RADIOCHROMIC FILMS; DOSIMETRY; PROTOCOL; QUALITY; SCANNER; PHANTOM;
D O I
10.1016/j.ejmp.2023.102627
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: This study aims to compare two methods for the organ dose evaluation in computed tomography (CT) in the head- and thorax regions: an experimental method, using radiochromic films, and a computational one, using a commercial software.Methods: Gafchromic & REG; XR-QA2 and EBT-3 were characterized in terms of energetic, angular, and irradiation configurations dependence. Two free-in-air irradiation calibration configurations were employed using a CT scanner: with the sensitive surface of the film orthogonal (OC) and parallel (PC) to the beam axis. Different dose-response curves were obtained by varying the irradiation configurations and the beam quality (BQ). Subsequently, films were irradiated within an anthropomorphic phantom using CT-thorax and -head protocols, and the organ dose values obtained were compared with those provided by the commercial software.Results: At different configurations, an unchanged dose response was achieved with EBT-3, while a dose response of 15% was obtained with XR-QA2. By varying BQ, XR-QA2 showed a different response below 10%, while EBT-3 showed a variation below 5% for dose values >20 mGy. For films irradiation angle equal to 90 degrees, the normalized to 0 degrees relative response was 41% for the XR-QA2 model and 83% for the EBT-3 one. Organ dose values obtained with EBT-3 for both configurations and with XR-QA2 for OC were in agreement with the DW values, showing percentage discrepancies of less than 25%. Conclusions: The obtained results showed the potential of EBT-3 in CT patient dosimetry since the lower angular dependence, compared to XR-QA2, compensates for low sensitivity in the diagnostic dose range.
引用
收藏
页数:9
相关论文
共 50 条
  • [1] A comparison of pediatric and adult CT organ dose estimation methods
    Yiming Gao
    Brian Quinn
    Usman Mahmood
    Daniel Long
    Yusuf Erdi
    Jean St. Germain
    Neeta Pandit-Taskar
    X. George Xu
    Wesley E. Bolch
    Lawrence T. Dauer
    BMC Medical Imaging, 17
  • [2] A comparison of pediatric and adult CT organ dose estimation methods
    Gao, Yiming
    Quinn, Brian
    Mahmood, Usman
    Long, Daniel
    Erdi, Yusuf
    Germain, Jean St.
    Pandit-Taskar, Neeta
    Xu, X. George
    Bolch, Wesley E.
    Dauer, Lawrence T.
    BMC MEDICAL IMAGING, 2017, 17
  • [3] Comparison of organ dosimetry methods and effective dose calculation methods for paediatric CT
    Brady, Z.
    Cain, T. M.
    Johnston, P. N.
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2012, 35 (02) : 117 - 134
  • [4] Comparison of organ dosimetry methods and effective dose calculation methods for paediatric CT
    Z. Brady
    T. M. Cain
    P. N. Johnston
    Australasian Physical & Engineering Sciences in Medicine, 2012, 35 : 117 - 134
  • [5] Estimation and comparison of effective dose (E) in standard chest CT by organ dose measurements and dose-length-product methods and assessment of the influence of CT tube potential (energy dependency) on effective dose in a dual-source CT
    Paul, Jijo
    Banckwitz, Rosemarie
    Krauss, Bernhard
    Vogl, Thomas J.
    Maentele, Werner
    Bauer, Ralf W.
    EUROPEAN JOURNAL OF RADIOLOGY, 2012, 81 (04) : E507 - E512
  • [6] Comparison of organ and effective dose estimations from different Monte Carlo simulation-based software methods in infant CT and comparison with direct phantom measurements
    Lawson, Michael
    Berk, Kemal
    Badawy, Mohamed
    Qi, Yujin
    Kuganesan, Ahilan
    Metcalfe, Peter
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2022, 23 (06):
  • [7] Advanced Computational Methods for Radiation Dose Optimization in CT
    Rao, Shreekripa
    Sharan, Krishna
    Chandraguthi, Srinidhi Gururajarao
    Dsouza, Rechal Nisha
    David, Leena R.
    Ravichandran, Sneha
    Mustapha, Mubarak Taiwo
    Shettigar, Dilip
    Uzun, Berna
    Kadavigere, Rajagopal
    Sukumar, Suresh
    Ozsahin, Dilber Uzun
    DIAGNOSTICS, 2024, 14 (09)
  • [8] Pediatric organ dose measurements in axial and helical multislice CT
    McDermott, Alanna
    White, R. Allen
    Mc-Nitt-Gray, Mike
    Angel, Erin
    Cody, Dianna
    MEDICAL PHYSICS, 2009, 36 (05) : 1494 - 1499
  • [9] Comparison between effective dose in PET/CT and CT
    Coulot, J.
    Magnier, F.
    Chauchat, P.
    Guilabert, N.
    Ricard, M.
    Dreuil, S.
    Lumbroso, J.
    MEDECINE NUCLEAIRE-IMAGERIE FONCTIONNELLE ET METABOLIQUE, 2011, 35 (03): : 156 - 162
  • [10] Comparison of CT Lung Density Measurements between Standard Full-Dose and Reduced-Dose Protocols
    Hatt, Charles R.
    Oh, Andrea S.
    Obuchowski, Nancy A.
    Charbonnier, Jean-Paul
    Lynch, David A.
    Humphries, Stephen M.
    RADIOLOGY-CARDIOTHORACIC IMAGING, 2021, 3 (02):