Accuracy of Nodule Volume and Airway Wall Thickness Measurement Using Low-Dose Chest CT on a Photon-Counting Detector CT Scanner

被引:10
|
作者
Dunning, Chelsea A. S. [1 ]
Marsh, Jeffrey F., Jr. [1 ]
Winfree, Timothy [1 ]
Rajendran, Kishore [1 ]
Leng, Shuai [1 ]
Levin, David L. [1 ]
Johnson, Tucker F. [1 ]
Fletcher, Joel G. [1 ]
McCollough, Cynthia H. [1 ]
Yu, Lifeng [1 ]
机构
[1] Mayo Clin, Dept Radiol, Rochester, MN USA
关键词
computed tomography; x-ray; lung cancer; pulmonary nodules; dose reduction; chronic obstructive pulmonary disease; OBSTRUCTIVE PULMONARY-DISEASE; NOISE-REDUCTION METHOD; IMAGE; IMPACT;
D O I
10.1097/RLI.0000000000000933
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
ObjectivesA comparison of high-resolution photon-counting detector computed tomography (PCD-CT) versus energy-integrating detector (EID) CT via a phantom study using low-dose chest CT to evaluate nodule volume and airway wall thickness quantification.Materials and MethodsTwelve solid and ground-glass lung nodule phantoms with 3 diameters (5 mm, 8 mm, and 10 mm) and 2 shapes (spherical and star-shaped) and 12 airway tube phantoms (wall thicknesses, 0.27-1.54 mm) were placed in an anthropomorphic chest phantom. The phantom was scanned with EID-CT and PCD-CT at 5 dose levels (CTDIvol = 0.1-0.8 mGy at Sn-100 kV, 7.35 mGy at 120 kV). All images were iteratively reconstructed using matched kernels for EID-CT and medium-sharp kernel (MK) PCD-CT and an ultra-sharp kernel (USK) PCD-CT kernel, and image noise at each dose level was quantified. Nodule volumes were measured using semiautomated segmentation software, and the accuracy was expressed as the percentage error between segmented and reference volumes. Airway wall thicknesses were measured, and the root-mean-square error across all tubes was evaluated.ResultsMK PCD-CT images had the lowest noise. At 0.1 mGy, the mean volume accuracy for the solid and ground-glass nodules was improved in USK PCD-CT (3.1% and 3.3% error) compared with MK PCD-CT (9.9% and 10.2% error) and EID-CT images (11.4% and 9.2% error), respectively. At 0.2 mGy and 0.8 mGy, the wall thickness root-mean-square error values were 0.42 mm and 0.41 mm for EID-CT, 0.54 mm and 0.49 mm for MK PCD-CT, and 0.23 mm and 0.16 mm for USK PCD-CT.ConclusionsUSK PCD-CT provided more accurate lung nodule volume and airway wall thickness quantification at lower radiation dose compared with MK PCD-CT and EID-CT.
引用
收藏
页码:283 / 292
页数:10
相关论文
共 50 条
  • [1] Low-dose performance of a whole-body research photon-counting CT scanner
    Yu, Zhicong
    Leng, Shuai
    Kappler, Steffen
    Hahn, Katharina
    Li, Zhoubo
    Halaweish, Ahmed F.
    Henning, Andre
    Ritman, Erik L.
    McCollough, Cynthia H.
    MEDICAL IMAGING 2016: PHYSICS OF MEDICAL IMAGING, 2016, 9783
  • [2] Low-dose CT of the abdomen: Initial experience on a novel photon-counting detector CT and comparison with energy-integrating detector CT
    Decker, Josua A.
    Bette, Stefanie
    Lubina, Nora
    Rippel, Katharina
    Braun, Franziska
    Risch, Franka
    Wollny, Claudia
    Scheurig-Muenkler, Christian
    Kroencke, Thomas J.
    Schwarz, Florian
    EUROPEAN JOURNAL OF RADIOLOGY, 2022, 148
  • [3] Noise performance of low-dose CT: comparison between an energy integrating detector and a photon counting detector using a whole-body research photon counting CT scanner
    Yu, Zhicong
    Leng, Shuai
    Kappler, Steffen
    Hahn, Katharina
    Li, Zhoubo
    Halaweish, Ahmed F.
    Henning, Andre
    McCollough, Cynthia H.
    JOURNAL OF MEDICAL IMAGING, 2016, 3 (04)
  • [4] CT Angiography of the Aorta Using Photon-counting Detector CT with Reduced Contrast Media Volume
    Higashigaito, Kai
    Mergen, Victor
    Eberhard, Matthias
    Jungblut, Lisa
    Hebeisen, Monika
    Ratzer, Susan
    Zanini, Bettina
    Kobe, Adrian
    Martini, Katharina
    Euler, Andre
    Alkadhi, Hatem
    RADIOLOGY-CARDIOTHORACIC IMAGING, 2023, 5 (01):
  • [5] Low-dose high-resolution chest CT in adults with cystic fibrosis: intraindividual comparison between photon-counting and energy-integrating detector CT
    Frings, Marko
    Welsner, Matthias
    Mousa, Christin
    Zensen, Sebastian
    Salhoefer, Luca
    Meetschen, Mathias
    Beck, Nikolas
    Bos, Denise
    Westhoelter, Dirk
    Wienker, Johannes
    Taube, Christian
    Umutlu, Lale
    Schaarschmidt, Benedikt M.
    Forsting, Michael
    Haubold, Johannes
    Sutharsan, Sivagurunathan
    Opitz, Marcel
    EUROPEAN RADIOLOGY EXPERIMENTAL, 2024, 8 (01)
  • [6] Low-dose lung cancer screening with photon-counting CT: a feasibility study
    Symons, Rolf
    Cork, Tyler E.
    Sahbaee, Pooyan
    Fuld, Matthew K.
    Kappler, Steffen
    Folio, Les R.
    Bluemke, David A.
    Pourmorteza, Amir
    PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (01): : 202 - 213
  • [7] Performance of calcium quantifications on low-dose photon-counting detector CT with high-pitch: A phantom study
    Zhou, Shanshui
    Liu, Peng
    Dong, Haipeng
    Li, Jiqiang
    Xu, Zhihan
    Schmidt, Bernhard
    Lin, Shushen
    Yang, Wenjie
    Yan, Fuhua
    Qin, Le
    HELIYON, 2024, 10 (12)
  • [8] Optimization of the Reconstruction Settings for Low-Dose Ultra-High-Resolution Photon-Counting Detector CT of the Lungs
    Graafen, Dirk
    Halfmann, Moritz C.
    Emrich, Tilman
    Yang, Yang
    Kreuter, Michael
    Dueber, Christoph
    Kloeckner, Roman
    Mueller, Lukas
    Jorg, Tobias
    DIAGNOSTICS, 2023, 13 (23)
  • [9] Quantum Iterative Reconstruction for Low-Dose Ultra-High-Resolution Photon-Counting Detector CT of the Lung
    Sartoretti, Thomas
    Racine, Damien
    Mergen, Victor
    Jungblut, Lisa
    Monnin, Pascal
    Flohr, Thomas G.
    Martini, Katharina
    Frauenfelder, Thomas
    Alkadhi, Hatem
    Euler, Andre
    DIAGNOSTICS, 2022, 12 (02)
  • [10] Low-dose chest CT and the impact on nodule visibility
    Tugwell-Allsup, J.
    Owen, B. W.
    England, A.
    RADIOGRAPHY, 2021, 27 (01) : 24 - 30