Accuracy of Nodule Volume and Airway Wall Thickness Measurement Using Low-Dose Chest CT on a Photon-Counting Detector CT Scanner

被引:10
|
作者
Dunning, Chelsea A. S. [1 ]
Marsh, Jeffrey F., Jr. [1 ]
Winfree, Timothy [1 ]
Rajendran, Kishore [1 ]
Leng, Shuai [1 ]
Levin, David L. [1 ]
Johnson, Tucker F. [1 ]
Fletcher, Joel G. [1 ]
McCollough, Cynthia H. [1 ]
Yu, Lifeng [1 ]
机构
[1] Mayo Clin, Dept Radiol, Rochester, MN USA
关键词
computed tomography; x-ray; lung cancer; pulmonary nodules; dose reduction; chronic obstructive pulmonary disease; OBSTRUCTIVE PULMONARY-DISEASE; NOISE-REDUCTION METHOD; IMAGE; IMPACT;
D O I
10.1097/RLI.0000000000000933
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
ObjectivesA comparison of high-resolution photon-counting detector computed tomography (PCD-CT) versus energy-integrating detector (EID) CT via a phantom study using low-dose chest CT to evaluate nodule volume and airway wall thickness quantification.Materials and MethodsTwelve solid and ground-glass lung nodule phantoms with 3 diameters (5 mm, 8 mm, and 10 mm) and 2 shapes (spherical and star-shaped) and 12 airway tube phantoms (wall thicknesses, 0.27-1.54 mm) were placed in an anthropomorphic chest phantom. The phantom was scanned with EID-CT and PCD-CT at 5 dose levels (CTDIvol = 0.1-0.8 mGy at Sn-100 kV, 7.35 mGy at 120 kV). All images were iteratively reconstructed using matched kernels for EID-CT and medium-sharp kernel (MK) PCD-CT and an ultra-sharp kernel (USK) PCD-CT kernel, and image noise at each dose level was quantified. Nodule volumes were measured using semiautomated segmentation software, and the accuracy was expressed as the percentage error between segmented and reference volumes. Airway wall thicknesses were measured, and the root-mean-square error across all tubes was evaluated.ResultsMK PCD-CT images had the lowest noise. At 0.1 mGy, the mean volume accuracy for the solid and ground-glass nodules was improved in USK PCD-CT (3.1% and 3.3% error) compared with MK PCD-CT (9.9% and 10.2% error) and EID-CT images (11.4% and 9.2% error), respectively. At 0.2 mGy and 0.8 mGy, the wall thickness root-mean-square error values were 0.42 mm and 0.41 mm for EID-CT, 0.54 mm and 0.49 mm for MK PCD-CT, and 0.23 mm and 0.16 mm for USK PCD-CT.ConclusionsUSK PCD-CT provided more accurate lung nodule volume and airway wall thickness quantification at lower radiation dose compared with MK PCD-CT and EID-CT.
引用
收藏
页码:283 / 292
页数:10
相关论文
共 50 条
  • [11] High-Pitch Multienergy Coronary CT Angiography in Dual-Source Photon-Counting Detector CT Scanner at Low Iodinated Contrast Dose
    Rajiah, Prabhakar Shantha
    Dunning, Chelsea A. S.
    Rajendran, Kishore
    Tandon, Yasmeen K.
    Ahmed, Zaki
    Larson, Nicholas B.
    Collins, Jeremy D.
    Thorne, Jamison
    Williamson, Eric
    Fletcher, Joel G.
    McCollough, Cynthia
    Leng, Shuai
    INVESTIGATIVE RADIOLOGY, 2023, 58 (09) : 681 - 690
  • [12] Pediatric contrast-enhanced chest CT on a photon-counting detector CT: radiation dose and image quality compared to energy-integrated detector CT
    Alexander M. El-Ali
    Naomi Strubel
    Lynne Pinkney
    Christine Xue
    Bari Dane
    Shailee V. Lala
    Pediatric Radiology, 2024, 54 (12) : 1984 - 1995
  • [13] Pulmonary nodule visualization and evaluation of AI-based detection at various ultra-low-dose levels using photon-counting detector CT
    Jungblut, Lisa
    Euler, Andre
    Landsmann, Anna
    Englmaier, Vanessa
    Mergen, Victor
    Sefirovic, Medina
    Frauenfelder, Thomas
    ACTA RADIOLOGICA, 2024, : 1238 - 1245
  • [14] Editorial Comment: Better Images of the Pelvis With Less Dose Using Photon-Counting Detector CT
    Roth, Trenton D.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2024, 222 (01)
  • [15] Technical feasibility proof for high-resolution low-dose photon-counting CT of the breast
    Kalender, Willi A.
    Kolditz, Daniel
    Steiding, Christian
    Ruth, Veikko
    Lueck, Ferdinand
    Rossler, Ann-Christin
    Wenkel, Evelyn
    EUROPEAN RADIOLOGY, 2017, 27 (03) : 1081 - 1086
  • [16] Technical feasibility proof for high-resolution low-dose photon-counting CT of the breast
    Willi A. Kalender
    Daniel Kolditz
    Christian Steiding
    Veikko Ruth
    Ferdinand Lück
    Ann-Christin Rößler
    Evelyn Wenkel
    European Radiology, 2017, 27 : 1081 - 1086
  • [17] Comparison of Image Quality and Radiation Dose in Pediatric Temporal Bone CT Using Photon-Counting Detector CT and Energy-Integrating Detector CT
    Lee, Jeong Sub
    Kim, John
    Bapuraj, Jayapalli R.
    Srinivasan, Ashok
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2024, 45 (09) : 1322 - 1326
  • [18] Comparison of Low Dose Performance of Photon-Counting and Energy Integrating CT
    Rajagopal, Jayasai R.
    Farhadi, Faraz
    Solomon, Justin
    Sahbaee, Pooyan
    Saboury, Babak
    Pritchard, William F.
    Jones, Elizabeth C.
    Samei, Ehsan
    ACADEMIC RADIOLOGY, 2021, 28 (12) : 1754 - 1760
  • [19] Low Dose Ultra-High Resolution Sinus Imaging Using Photon-Counting Detector CT with Tin Filter
    Rajendran, K.
    Zhou, W.
    Tao, S.
    Voss, B.
    Bruesewitz, M.
    Weaver, J.
    DeLone, D.
    Fletcher, J.
    McCollough, C.
    Leng, S.
    MEDICAL PHYSICS, 2019, 46 (06) : E409 - E409
  • [20] Reducing contrast media and radiation dose in CT angiography at low tube voltage: animal study with photon-counting detector CT
    Konstantin Klambauer
    Thomas Flohr
    Lukas Jakob Moser
    Victor Mergen
    Matthias Eberhard
    Andreas Prokein
    Hatem Alkadhi
    Hubertus Pietsch
    Gregor Jost
    European Radiology Experimental, 9 (1)