The energy dependence of BaSO4: Eu nanophosphors for thermoluminescence dosimetry of orthovoltage X-rays and low

被引:3
|
作者
Nattudurai, Ravikumar [1 ,4 ]
Arous, Delmon [1 ,2 ]
Edin, Nina F. J. [1 ]
Pandey, Anant [3 ]
Malinen, Eirik [1 ,2 ]
机构
[1] Univ Oslo, Dept Phys, Oslo, Norway
[2] Oslo Univ Hosp, Dept Med Phys, Oslo, Norway
[3] Univ Delhi, Sri Venkateswara Coll, Dept Phys, New Delhi, India
[4] Univ Oslo, Dept Phys, Sect Biophys & Med Phys, Oslo, Norway
关键词
Thermoluminescence; X-rays; Protons; Dosimetry; Energy dependence; TRACK STRUCTURE-THEORY; RADIATION-THERAPY; TL; SPECTRA;
D O I
10.1016/j.radphyschem.2023.110928
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
Accurate dosimetry of ionizing radiation requires careful consideration of the absorption properties of the dosimeter material in question. The energy dependence of a dosimeter signifies the variation in dosimeter signal per absorbed dose to water with radiation type and energy. The nanophosphor BaSO4: Eu is a radiosensitive material for thermoluminescence (TL) dosimetry. However, the energy dependence of the material for orthovoltage X-rays and low energy protons has not previously been investigated, which is the purpose of the current work. We irradiated BaSO4 doped with 0.5 mol% of europium with 100, 160 and 225 kV X-rays and 2 and 9 MeV protons. Co-60 gamma-rays were used for reference irradiation. Absolute dose-to-water dosimetry was ensured with an ion chamber. Delivered calibration doses ranged from 0.01 to about 2 Gy. TL output was measured at a heating rate of 5 K/s, where the maximum intensity of the dominant glow peak at about 212 degrees C defined the dosimeter signal. The signal was found to increase linearly with absorbed dose for both radiation types and all energies. The measured energy response, defined as the ratio of the TL calibration slope at a given radiation type and energy to that following Co-60 gamma-irradiation, varied between 47 and 62 for X-rays and between 0.41 and 0.61 for protons. Theoretical modelling of the energy response and proton detector efficiency gave similar trends. The findings show that BaSO4 is an energy-dependent material, especially for X-rays, but may still prove valuable for accurate dose determination if the dosimeter is calibrated in the user beam.
引用
收藏
页数:7
相关论文
共 50 条
  • [21] THE APPLICATION OF MICRODOSIMETRIC MEASUREMENTS TO THE STUDY OF THE DOSIMETRY OF LOW-ENERGY X-RAYS
    WAKER, AJ
    EIVAZI, MT
    RADIATION PROTECTION DOSIMETRY, 1989, 29 (1-2) : 75 - 79
  • [22] MEASUREMENT OF THE RESTRICTED LET OF ORTHOVOLTAGE AND LOW-ENERGY X-RAYS IN TISSUE-EQUIVALENT PLASTIC
    MAKRIGIORGOS, GM
    CHIN, LM
    PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (06): : 835 - 841
  • [23] Study of induced thermoluminescence in CVD diamond film by low-energy X-rays
    Liu, CC
    Lin, JP
    Chu, TC
    APPLIED RADIATION AND ISOTOPES, 2003, 59 (01) : 79 - 85
  • [24] Effect of thermal annealing on the thermoluminescence glow curve of X-irradiated BaSO4:Pr
    Singh, M. Deben
    Devi, Y. Rangeela
    Singh, S. Dorendrajit
    INDIAN JOURNAL OF PURE & APPLIED PHYSICS, 2009, 47 (06) : 432 - 434
  • [25] DOSIMETRY OF IRREGULAR FIELDS FOR 4 MV X-RAYS
    GOLDEN, R
    PETERSON, MD
    PHYSICS IN MEDICINE AND BIOLOGY, 1972, 17 (01): : 116 - &
  • [26] Thermoluminescence properties of nanocrystalline BaSO4 doped with Eu2+ produced by solid state combustion synthesis
    Hussain, Taqmeem
    Asfora, Viviane K.
    da Nobrega, Boisguilbert P. A. G.
    de Barros, Vinicius S. M.
    de Azevedo, Walter M.
    Khoury, Helen J.
    RADIATION PHYSICS AND CHEMISTRY, 2021, 186
  • [27] Investigation of the radiological properties of various phantoms for their application in low energy X-rays dosimetry
    Langaroodi, Roghayeh Kargar Shaker
    Abtahi, Seyed Mohammad Mahdi
    Akbari, Mohammad Esmaeil
    RADIATION PHYSICS AND CHEMISTRY, 2019, 157 : 33 - 39
  • [28] TLD-700 for proton dosimetry in the presence of low-energy X-rays
    Miljanic, S
    Ranogajec-Komor, M
    Blagus, S
    Miljanic, D
    Osvay, M
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2004, 519 (03): : 667 - 673
  • [29] Energy dependence of TLD-900 dosimeters exposed to low energy X-rays
    Avila, O.
    Ramirez-Barbosa, E.
    Gamboa-deBuen, I.
    RADIATION MEASUREMENTS, 2014, 71 : 127 - 132
  • [30] ABSORBED DOSE DETERMINATION OF ORTHOVOLTAGE X-RAYS USING PHOTON ENERGY FLUENCE CALORIMETER
    RAO, ISS
    NAIK, SB
    INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES, 1984, 35 (01): : 15 - 20