SAFIR-I: first NEMA NU 4-2008-based performance characterization

被引:0
|
作者
Bebie, Pascal [1 ]
Lustermann, Werner [1 ]
Debus, Jan [1 ]
Ritzer, Christian [1 ,4 ]
Dissertori, Guenther [1 ]
Weber, Bruno [2 ,3 ]
机构
[1] Swiss Fed Inst Technol, Inst Particle Phys & Astrophys, Otto Stern Weg 5, CH-8093 Zurich, Switzerland
[2] Univ Zurich, Inst Pharmacol & Toxicol, Winterthurerstr 190, CH-8057 Zurich, Switzerland
[3] Univ Zurich, Neurosci Ctr Zurich, Winterthurerstr 190, CH-8057 Zurich, Switzerland
[4] Swiss Fed Inst Technol, Winterthurerstr 190, CH-8057 Zurich, Switzerland
关键词
Positron Emission Tomography; PET/MRI; Preclinical imaging; SAFIR; NEMA NU 4-2008; Performance characterization; PET;
D O I
10.1186/s40658-023-00603-1
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: Small Animal Fast Insert for MRI detector I (SAFIR-I) is a novel Positron Emission Tomography insert for a 7 T Bruker BioSpec 70/30 Ultra Shield Refrigerated Magnetic Resonance Imaging (MRI) system. It facilitates truly simultaneous quantitative imaging in mice and rats at injected activities as high as 500 MBq. Exploitation of the resulting high count rates enables quick image formation at few seconds per frame. In this investigation, key performance parameters of SAFIR-I have been determined according to the evaluations outlined in the National Electrical Manufacturers Association (NEMA) Standards Publication NU 4-2008 (NEMA-NU4) protocol. Results: Using an energy window of 391 to 601 keV and a Coincidence Timing Window of 500 ps, the following performance was observed: The average spatial resolution at 5mm radial offset (Full Width at Half Maximum) is 2.54mm when using Filtered Backprojection, 3D Reprojection reconstruction. For the mouse- and rat-like phantoms, the maximal Noise-Equivalent Count Rates (NECRs) are 1368 kcps at the highest tested average effective concentration of 14.7 MBq cc-1, and 713 kcps at the highest tested average effective concentration of 1.72 MBq cc-1, respectively. The NECR peak is not yet reached for either of these cases. The peak sensitivity is 1.46%. The Image Quality phantom uniformity standard deviation is 4.8%. The Recovery Coefficient for the 5mm rod is (1.08 +/- 0.10). The Spill-Over Ratios are (0.22 +/- 0.03) and (0.22 +/- 0.02), for the waterand air-filled cylinder, respectively. An accuracy of 4.3% was achieved for the quantitative calibration of reconstructed voxel values. Conclusions: The measured performance parameters indicate that the various design goals have been achieved. SAFIR-I offers excellent performance, especially at the high activities it was designed for. This facilitates planned experiments with fast tracer kinetics in small animals. Ways to potentially improve performance can still be explored. Simultaneously, further performance gains can be expected for a forthcoming insert featuring 2.7 times longer axial coverage named Small Animal Fast Insert for MRI detector II (SAFIR-II).
引用
收藏
页数:20
相关论文
共 50 条
  • [31] Comparative performance measurement of 124I and 18F PET on Siemens Inveon PET using NEMA NU4-2008: Toward optimal parameters in 124I PET
    Yu, A. Ram
    Kim, Jin Su
    Kim, Kyeong Min
    Lee, Young Sub
    Kim, Jong Guk
    Park, JiAe
    Woo, Sang-Keun
    Kim, Hee-Joung
    Cheon, Gi Jeong
    JOURNAL OF NUCLEAR MEDICINE, 2010, 51
  • [32] NEMA NU 4-2008 performance evaluation and MR compatibility tests of an APD-based small animal PET-insert for simultaneous PET/MR imaging
    Disselhorst, Jonathan A.
    Newport, Danny F.
    Schmid, Andreas M.
    Schmidt, Fabian P.
    Parl, Christoph
    Liu, Chih-Chieh
    Pichler, Bernd J.
    Mannheim, Julia G.
    PHYSICS IN MEDICINE AND BIOLOGY, 2022, 67 (04):
  • [33] Performance characteristics of a new development dedicated breast PET scanner evaluated with the NEMA NU 2-2008 standard
    Inoue, Mika
    Nakamoto, Yuji
    Matsumoto, Keiichi
    Kawase, Shigeto
    Yamakawa, Yoshiyuki
    Ohi, Junichi
    Kitamura, Keishi
    Togashi, Kaori
    Higashimura, Kyoji
    Onoguchi, Masahisa
    JOURNAL OF NUCLEAR MEDICINE, 2011, 52
  • [34] NEMA NU-4 performance evaluation of PETbox4, a high sensitivity dedicated PET preclinical tomograph
    Gu, Z.
    Taschereau, R.
    Vu, N. T.
    Wang, H.
    Prout, D. L.
    Silverman, R. W.
    Bai, B.
    Stout, D. B.
    Phelps, M. E.
    Chatziioannou, A. F.
    PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (11): : 3791 - 3814
  • [35] NEMA NU 4-2008 and in vivo imaging performance of RAYCAN trans-PET/CT X5 small animal imaging system
    Teuho, J.
    Han, C.
    Riehakainen, L.
    Honkaniemi, A.
    Tirri, M.
    Liljenback, H.
    Virta, J.
    Gu, S.
    Liu, S.
    Wan, L.
    Teras, M.
    Roivainen, A.
    Xie, Q.
    Knuuti, J.
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (11):
  • [36] PET performance evaluation of the small-animal Hyperion IID PET/MRI insert based on the NEMA NU-4 standard
    Hallen, Patrick
    Schug, David
    Weissler, Bjoern
    Gebhardt, Pierre
    Salomon, Andre
    Kiessling, Fabian
    Schulz, Volkmar
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2018, 4 (06):
  • [37] Performance Test of the MiniPET-II Small Animal Scanner According to the NEMA NU-4 Standard
    Kis, S. A.
    Lajtos, I.
    Emri, M.
    Tron, L.
    Opposits, G.
    Buekki, T.
    Hegyesi, Gy.
    Imrek, J.
    Valastyan, I.
    Molnar, J.
    Novak, D.
    Balkay, L.
    2009 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2009, : 3185 - +
  • [38] NEMA NU 1-2018 performance characterization and Monte Carlo model validation of the Cubresa Spark SiPM-based preclinical SPECT scanner
    Strugari, Matthew E. E.
    DeBay, Drew R. R.
    Beyea, Steven D. D.
    Brewer, Kimberly D. D.
    EJNMMI PHYSICS, 2023, 10 (01)
  • [39] Performance tests for ray-scan 64 PET/CT based on NEMA NU-2 2007
    Li, Suying
    Zhou, Kun
    Zhang, Qiushi
    Zhang, Jinming
    Yang, Kun
    Xu, Baixuan
    Ren, Qiushi
    JOURNAL OF INSTRUMENTATION, 2015, 10
  • [40] NEMA NU-04-based performance characteristics of the LabPET-8™ small animal PET scanner
    Prasad, Rameshwar
    Ratib, Osman
    Zaidi, Habib
    PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (20): : 6649 - 6664