SAFIR-I: first NEMA NU 4-2008-based performance characterization

被引:0
|
作者
Bebie, Pascal [1 ]
Lustermann, Werner [1 ]
Debus, Jan [1 ]
Ritzer, Christian [1 ,4 ]
Dissertori, Guenther [1 ]
Weber, Bruno [2 ,3 ]
机构
[1] Swiss Fed Inst Technol, Inst Particle Phys & Astrophys, Otto Stern Weg 5, CH-8093 Zurich, Switzerland
[2] Univ Zurich, Inst Pharmacol & Toxicol, Winterthurerstr 190, CH-8057 Zurich, Switzerland
[3] Univ Zurich, Neurosci Ctr Zurich, Winterthurerstr 190, CH-8057 Zurich, Switzerland
[4] Swiss Fed Inst Technol, Winterthurerstr 190, CH-8057 Zurich, Switzerland
关键词
Positron Emission Tomography; PET/MRI; Preclinical imaging; SAFIR; NEMA NU 4-2008; Performance characterization; PET;
D O I
10.1186/s40658-023-00603-1
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: Small Animal Fast Insert for MRI detector I (SAFIR-I) is a novel Positron Emission Tomography insert for a 7 T Bruker BioSpec 70/30 Ultra Shield Refrigerated Magnetic Resonance Imaging (MRI) system. It facilitates truly simultaneous quantitative imaging in mice and rats at injected activities as high as 500 MBq. Exploitation of the resulting high count rates enables quick image formation at few seconds per frame. In this investigation, key performance parameters of SAFIR-I have been determined according to the evaluations outlined in the National Electrical Manufacturers Association (NEMA) Standards Publication NU 4-2008 (NEMA-NU4) protocol. Results: Using an energy window of 391 to 601 keV and a Coincidence Timing Window of 500 ps, the following performance was observed: The average spatial resolution at 5mm radial offset (Full Width at Half Maximum) is 2.54mm when using Filtered Backprojection, 3D Reprojection reconstruction. For the mouse- and rat-like phantoms, the maximal Noise-Equivalent Count Rates (NECRs) are 1368 kcps at the highest tested average effective concentration of 14.7 MBq cc-1, and 713 kcps at the highest tested average effective concentration of 1.72 MBq cc-1, respectively. The NECR peak is not yet reached for either of these cases. The peak sensitivity is 1.46%. The Image Quality phantom uniformity standard deviation is 4.8%. The Recovery Coefficient for the 5mm rod is (1.08 +/- 0.10). The Spill-Over Ratios are (0.22 +/- 0.03) and (0.22 +/- 0.02), for the waterand air-filled cylinder, respectively. An accuracy of 4.3% was achieved for the quantitative calibration of reconstructed voxel values. Conclusions: The measured performance parameters indicate that the various design goals have been achieved. SAFIR-I offers excellent performance, especially at the high activities it was designed for. This facilitates planned experiments with fast tracer kinetics in small animals. Ways to potentially improve performance can still be explored. Simultaneously, further performance gains can be expected for a forthcoming insert featuring 2.7 times longer axial coverage named Small Animal Fast Insert for MRI detector II (SAFIR-II).
引用
收藏
页数:20
相关论文
共 50 条
  • [41] NEMA NU-04-based Performance Characteristics of the LabPET-8™ Small Animal PET Scanner
    Prasad, Rameshwar
    Ratib, Osman
    Zaidi, Habib
    2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 3654 - 3660
  • [42] NEMA NU 1-2018 performance characterization and Monte Carlo model validation of the Cubresa Spark SiPM-based preclinical SPECT scanner
    Matthew E. Strugari
    Drew R. DeBay
    Steven D. Beyea
    Kimberly D. Brewer
    EJNMMI Physics, 10
  • [43] PERFORMANCE EVALUATION OF PRECLINICAL PET SCANNERS WITH THE NEMA NU-4 IMAGE QUALITY PHANTOM USING A COLLECTION OF RADIOISOTOPES
    Nezich, Rik
    Reilhac, Anthonin
    Zahra, David
    Price, Roger
    INTERNAL MEDICINE JOURNAL, 2013, 43 : 23 - 23
  • [44] Characterization and optimization of image quality as a function of reconstruction algorithms and parameter settings in a Siemens Inveon small-animal PET scanner using the NEMA NU 4-2008 standards
    Visser, Eric P.
    Disselhorst, Jonathan A.
    van Lier, Monique G. J. T. B.
    Laverman, Peter
    de Jong, Gabie M.
    Oyen, Wim J. G.
    Boerman, Otto C.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2011, 629 (01): : 357 - 367
  • [45] Image quality assessment for several positron emitters using the NEMA NU 4 2008 standards in the Inveon small-animal PET scanner
    Disselhorst, Jonathan
    Brom, Maarten
    Laverman, Peter
    Slump, Cornelis
    Boerman, Otto
    Oyen, Wim
    Gotthardt, Martin
    Visser, Eric
    JOURNAL OF NUCLEAR MEDICINE, 2010, 51
  • [46] Monte Carlo simulations versus experimental measurements in a small animal PET system. A comparison in the NEMA NU 4-2008 framework
    Popota, F. D.
    Aguiar, P.
    Espana, S.
    Lois, C.
    Udias, J. M.
    Ros, D.
    Pavia, J.
    Gispert, J. D.
    PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (01): : 151 - 162
  • [47] Performance of 3D MAP in small animal PET: evaluation using the NEMA NU-4 standard phantom
    Khalil, M.
    Gsell, W.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2011, 38 : S262 - S262
  • [48] Performance Evaluation for 68Ga and 18F of the ARGUS Small-Animal PET Scanner Based on the NEMA NU-4 Standard
    Canadas, M.
    Romero Sanz, E.
    Oteo Vives, M.
    Vaquero, J. J.
    Desco, M.
    Vicente, E.
    Udias, J. M.
    Romero, L.
    2010 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD (NSS/MIC), 2010, : 3454 - 3457
  • [49] Preclinical QC made easy: automated analysis of imaging data from a NEMA NU 4-based phantom
    van Tiel, S.
    Hoefter, A.
    Haeck, J.
    Gremse, F.
    Bernsen, M.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2022, 49 (SUPPL 1) : S190 - S190
  • [50] Performance comparison of two commercial BGO-based PET/CT scanners using NEMA NU 2-2001
    Bolard, Gregory
    Prior, John O.
    Modolo, Luca
    Delaloye, Angelika Bischof
    Kosinski, Marek
    Wastiel, Claude
    Malterre, Jerome
    Bulling, Shelley
    Bochud, Francois
    Verdun, Francis R.
    MEDICAL PHYSICS, 2007, 34 (07) : 2708 - 2717