Wear estimation of hip implants with varying chamfer geometry at the trunnion junction: a finite element analysis

被引:24
|
作者
Gutmann, Celine [1 ]
Shaikh, Numa [2 ]
Shenoy, B. Satish [3 ]
Bhat, N. Shaymasunder [4 ]
Keni, Laxmikant G. [3 ]
Chethan, K. N. [3 ]
机构
[1] Univ Appl Sci Northwestern Switzerland, Inst Med Engn & Med Informat, Sch Life Sci HLS, FHNW, Windisch, Switzerland
[2] Manipal Acad Higher Educ, Manipal Inst Technol, Dept Mech & Ind Engn, Manipal, Karnataka, India
[3] Manipal Acad Higher Educ, Manipal Inst Technol, Dept Aeronaut & Automobile Engn, Manipal, Karnataka, India
[4] Manipal Acad Higher Educ, Kasturba Med Coll, Dept Orthopaed, Manipal, Karnataka, India
关键词
hip implant; finite element analysis; ansys; linear wear; pressure; TAPER JUNCTION; FRETTING WEAR; ARTHROPLASTY; STEM; REPLACEMENTS; FAILURE; RATES;
D O I
10.1088/2057-1976/acb710
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The hip joint helps the upper body to transfer its weight to lower body. Along with age, there are various reasons for the degeneration of the hip joint. The artificial hip implant replaces the degenerated hip. Wear between the joints is the primary cause of the hip implant becoming loose. The wear can occur due to various reasons. Due to this revision surgery are most common in young and active patients. In the design phase of the implant if this is taken care then life expectancy of the implant can be improved. Small design changes can significantly enhance the implant's life. In this work, elliptical-shaped hip implant stem is designed, and linear wear is estimated at trunnion junction. In this work, a 28 mm diameter femoral head with a 4 mm thick acetabular cup and a 2 mm thick backing cup is used. The top surface taper radiuses are changed. Solid works was used to create the models. Ansys was used to perform the analysis. It was found that as the radius of the TTR decreased, the wear rate decreased. The least wear rate was found in 12/14 mm taper with a value of 1.15E(-02) mm year(-1) for the first material combination and with a value of 1.23E (-02) mm year(-1) for the second material combination. In the comparison between the models with 1 mm chamfer and no chamfer, it was found that the wear rate was lower for the models with 1 mm chamfer. When the chamfer was increased (more than 1 mm), the linear wear increased. Wear is the main reason for the loosening of hip implants, which leads to a revision of an implant. It was found that with a decrease in TTR, there was a small increase in the linear wear rate. Overall, the implant with TTR 6 mm and a chamfer of 1 mm was found to have the least wear rate. To validate these results, the implant can be 3D printed and tested on a hip simulator.
引用
收藏
页数:11
相关论文
共 50 条
  • [21] Estimation of Endurance Performance of Hip Stems Using Finite Element Analysis
    Kang, Yu-Bong
    Jung, Duk-Young
    Tsutsumi, Sadami
    BIOMATERIALS IN ASIA: IN COMMEMORATION OF THE 1ST ASIAN BIOMATERIALS CONGRESS, 2008, : 517 - +
  • [22] A large taper mismatch is one of the key factors behind high wear rates and failure at the taper junction of total hip replacements: A finite element wear analysis
    Ashkanfar, Ariyan
    Langton, David J.
    Joyce, Thomas J.
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2017, 69 : 257 - 266
  • [23] A 3D Finite Element Analysis Of Dental Implants With Varying Thread Angles
    NarendraKumar, U.
    Mathew, Arun Tom
    Iyer, Nikhil
    Rahman, Febin
    Manjubala, I.
    MATERIALS TODAY-PROCEEDINGS, 2018, 5 (05) : 11900 - 11905
  • [24] Influence of bearing geometry and material on the contact of dual mobility implants: a finite element analysis
    Uddin, M. S.
    INTERNATIONAL JOURNAL OF BIOMEDICAL ENGINEERING AND TECHNOLOGY, 2014, 15 (01) : 45 - 58
  • [25] Comparative Finite Element Analysis of the Debonding Process in Different Concepts of Cemented Hip Implants
    Perez, M. A.
    Palacios, J.
    ANNALS OF BIOMEDICAL ENGINEERING, 2010, 38 (06) : 2093 - 2106
  • [26] Comparative Finite Element Analysis of the Debonding Process in Different Concepts of Cemented Hip Implants
    M. A. Pérez
    J. Palacios
    Annals of Biomedical Engineering, 2010, 38 : 2093 - 2106
  • [27] How Does the Micro-Groove Profile Influence the Mechanics of Taper Junction in Hip Implants? A Finite Element Study
    Kalwar, Akash
    Feyzi, Mohsen
    Hashemi, Reza
    BIOMECHANICS, 2023, 3 (04): : 596 - 607
  • [28] Finite Element Analysis of Wear in Piercing
    Yang Yu
    Liu Cunping
    Yuan Yongfu
    Guo Sheng
    APPLIED MECHANICS AND MATERIALS I, PTS 1-3, 2013, 275-277 : 2156 - 2160
  • [29] Chamfer drill geometric parameters optimisation by finite element simulation and experimental analysis
    Puoza J.C.
    Appiah S.T.
    Zhang T.
    Aboagye B.
    International Journal of Machining and Machinability of Materials, 2020, 22 (05) : 406 - 429
  • [30] A biomechanical evaluation of orthopaedic implants for hip fractures by finite element analysis and in-vitro tests
    Eberle, S.
    Gerber, C.
    von Oldenburg, G.
    Hoegel, F.
    Augat, P.
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2010, 224 (H10) : 1141 - 1152