Wear estimation of hip implants with varying chamfer geometry at the trunnion junction: a finite element analysis

被引:24
|
作者
Gutmann, Celine [1 ]
Shaikh, Numa [2 ]
Shenoy, B. Satish [3 ]
Bhat, N. Shaymasunder [4 ]
Keni, Laxmikant G. [3 ]
Chethan, K. N. [3 ]
机构
[1] Univ Appl Sci Northwestern Switzerland, Inst Med Engn & Med Informat, Sch Life Sci HLS, FHNW, Windisch, Switzerland
[2] Manipal Acad Higher Educ, Manipal Inst Technol, Dept Mech & Ind Engn, Manipal, Karnataka, India
[3] Manipal Acad Higher Educ, Manipal Inst Technol, Dept Aeronaut & Automobile Engn, Manipal, Karnataka, India
[4] Manipal Acad Higher Educ, Kasturba Med Coll, Dept Orthopaed, Manipal, Karnataka, India
关键词
hip implant; finite element analysis; ansys; linear wear; pressure; TAPER JUNCTION; FRETTING WEAR; ARTHROPLASTY; STEM; REPLACEMENTS; FAILURE; RATES;
D O I
10.1088/2057-1976/acb710
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The hip joint helps the upper body to transfer its weight to lower body. Along with age, there are various reasons for the degeneration of the hip joint. The artificial hip implant replaces the degenerated hip. Wear between the joints is the primary cause of the hip implant becoming loose. The wear can occur due to various reasons. Due to this revision surgery are most common in young and active patients. In the design phase of the implant if this is taken care then life expectancy of the implant can be improved. Small design changes can significantly enhance the implant's life. In this work, elliptical-shaped hip implant stem is designed, and linear wear is estimated at trunnion junction. In this work, a 28 mm diameter femoral head with a 4 mm thick acetabular cup and a 2 mm thick backing cup is used. The top surface taper radiuses are changed. Solid works was used to create the models. Ansys was used to perform the analysis. It was found that as the radius of the TTR decreased, the wear rate decreased. The least wear rate was found in 12/14 mm taper with a value of 1.15E(-02) mm year(-1) for the first material combination and with a value of 1.23E (-02) mm year(-1) for the second material combination. In the comparison between the models with 1 mm chamfer and no chamfer, it was found that the wear rate was lower for the models with 1 mm chamfer. When the chamfer was increased (more than 1 mm), the linear wear increased. Wear is the main reason for the loosening of hip implants, which leads to a revision of an implant. It was found that with a decrease in TTR, there was a small increase in the linear wear rate. Overall, the implant with TTR 6 mm and a chamfer of 1 mm was found to have the least wear rate. To validate these results, the implant can be 3D printed and tested on a hip simulator.
引用
收藏
页数:11
相关论文
共 50 条
  • [1] Wear estimation of trapezoidal and circular shaped hip implants along with varying taper trunnion radiuses using finite element method
    Chethan, K. N.
    Ogulcan, Guldeniz
    Bhat, Shyamasunder N.
    Zuber, Mohammad
    Shenoy, Satish B.
    COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2020, 196
  • [2] Wear estimation at the contact surfaces of oval shaped hip implants using finite element analysis
    Shaikh, Numa
    Shenoy, B. Satish
    Bhat, N. Shaymasunder
    Shetty, Sawan
    Chethan, K. N.
    COGENT ENGINEERING, 2023, 10 (01):
  • [3] A review on the finite element simulation of fretting wear and corrosion in the taper junction of hip replacement implants
    Feyzi, Mohsen
    Fallahnezhad, Khosro
    Taylor, Mark
    Hashemi, Reza
    COMPUTERS IN BIOLOGY AND MEDICINE, 2021, 130
  • [4] A finite element methodology for wear-fatigue analysis for modular hip implants
    Zhang, T.
    Harrison, N. M.
    McDonnell, P. F.
    McHugh, P. E.
    Leen, S. B.
    TRIBOLOGY INTERNATIONAL, 2013, 65 : 113 - 127
  • [5] Design and Performance Evaluation of a Novel Spiral Head-Stem Trunnion for Hip Implants Using Finite Element Analysis
    Soliman, Md Mohiuddin
    Chowdhury, Muhammad E. H.
    Islam, Mohammad Tariqul
    Musharavati, Farayi
    Mahmud, Sakib
    Hafizh, Muhammad
    Ayari, Mohamed Arselene
    Khandakar, Amith
    Alam, Mohammad Kaosar
    Nezhad, Erfan Zal
    MATERIALS, 2023, 16 (04)
  • [6] Contact Pressure, sliding distance and wear rate analysis at trunnion of hip implant for daily Activities: A finite element approach
    Soliman, Md Mohiuddin
    Islam, Mohammad Tariqul
    Kirawanich, Phumin
    Chowdhury, Muhammad E. H.
    Alam, Touhidul
    Alrashdi, Ayed M.
    Misran, Norbahiah
    Soliman, Mohamed S.
    BIOCYBERNETICS AND BIOMEDICAL ENGINEERING, 2025, 45 (02) : 137 - 153
  • [7] Finite Element Analysis of the Influence of Chamfer Hub Geometry on the Stress Concentrations of Shrink Fits
    Izard, Eulalia
    Garcia-Martin, Roberto
    Rodriguez-Martin, Manuel
    Lorenzo, Miguel
    APPLIED SCIENCES-BASEL, 2023, 13 (06):
  • [8] FINITE ELEMENT APPROACH TO DESIGN OF MODULAR HIP IMPLANTS MINIMIZING FRETTING WEAR
    Mech, Dhruba Jyoti
    Chakraborty, Arindam
    Chowdhury, Amit Roy
    Datta, Pallab
    JOURNAL OF MECHANICS IN MEDICINE AND BIOLOGY, 2022, 22 (07)
  • [9] The wear of the taper junction of the stem trunnion with the femoral head of a total modular hip prosthesis
    Capitanu, Lucian
    Badita, Liliana-Laura
    Tiganesteanu, Constantin
    Florescu, Virgil
    JURNAL TRIBOLOGI, 2019, 23 : 61 - 75
  • [10] Finite element based simulation of tribocorrosion at the head-neck junction of hip implants
    Fallahnezhad, Khosro
    Feyzi, Mohsen
    Ghadirinejad, Khashayar
    Hashemi, Reza
    Taylor, Mark
    TRIBOLOGY INTERNATIONAL, 2022, 165