Eye lens dose in spine surgeons during myelography procedures: a dosimetry study

被引:0
|
作者
Nagamoto, Keisuke [1 ]
Kawachino, Tomonori [1 ]
Suetsugu, Yoshiki [1 ]
Saruwatari, Hiromi [1 ]
Nihei, Shun-ichi [2 ]
Kunugita, Naoki [3 ]
机构
[1] Hosp Univ Occupat & Environm Hlth, Dept Radiol, 1-1 Iseigaoka,Yahatanishi Ku, Fukuoka, Japan
[2] Univ Occupat & Environm Hlth, Dept Emergency & Intens Care Med, 1-1 Iseigaoka,Yahatanishi Ku, Fukuoka, Japan
[3] Univ Occupat & Environm Hlth, Dept Occupat & Community Hlth, Nursing Sch Hlth Sci, 1-1 Iseigaoka,Yahatanishi Ku, Fukuoka, Japan
关键词
3 mm dose equivalent; Hp(3); lens dose; myelography; spine surgeons; radio-photoluminescence glass dosimeters; x-ray tube system; INTERVENTIONAL RADIOLOGY; RADIATION-EXPOSURE; FLUOROSCOPY; TRIAL; CT;
D O I
10.1088/1361-6498/ad0b3a
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
To determine the eye lens dose (3 mm dose equivalent [Hp(3)]) received by spine surgeons during myelography and evaluate the effectiveness of radiation-protective glasses and x-ray tube system positioning in reducing radiation exposure. This study included spine surgeons who performed myelography using over- or under-table x-ray tube systems. Hp(3) was measured for each examination using a radio-photoluminescence glass dosimeter (GD-352M) mounted on radiation-protective glass. This study identified significantly high Hp(3) levels, especially in the right eye lens in spinal surgeons. The median Hp(3) values in the right eye were 524 (391-719) and 58 (42-83) mu Sv/examination for over- and under-table x-ray tube systems, respectively. Further, Hp(3)AK, which was obtained by dividing the cumulative air kerma from Hp(3), was 8.09 (6.69-10.21) and 5.11 (4.06-6.31) mu Sv mGy-1 for the over- and under-table x-ray tube systems, respectively. Implementing radiation-protective glasses resulted in dose reduction rates of 54% (50%-57%) and 54% (51%-60%) for the over- and under-table x-ray tube systems, respectively. The use of radiation protection glasses significantly reduced the radiation dose in the eye lens during myelography, with the most effective measures being the combination of using radiation protection glasses and an under-table x-ray tube system.
引用
收藏
页数:10
相关论文
共 50 条
  • [31] IRRADIATION DOSE TO THE LENS OF THE EYE DURING CT OF THE HEAD
    LUND, E
    HALABURT, H
    NEURORADIOLOGY, 1982, 22 (04) : 181 - 184
  • [32] Estimation of equivalent dose to the eye lens and hands in wrist surgery using OSL dosimetry
    Patino, G.
    Alaman, C.
    Luquero, N.
    Ballester, F.
    Vijande, J.
    Diez, S.
    2016 GLOBAL MEDICAL ENGINEERING PHYSICS EXCHANGES/PAN AMERICAN HEALTH CARE EXCHANGES (GMEPE/PAHCE), 2016,
  • [33] Practical guidelines for personal monitoring and estimation of effective dose and dose to the lens of the eye in interventional procedures
    Askounis, Panagiotis
    Gonzalez, Anna Torras
    Ginjaume, Merce
    Carinou, Eleftheria
    JOURNAL OF RADIOLOGICAL PROTECTION, 2022, 42 (03)
  • [34] MONTE CARLO CALCULATIONS ON EXTREMITY AND EYE LENS DOSIMETRY FOR MEDICAL STAFF AT INTERVENTIONAL RADIOLOGY PROCEDURES
    Carinou, E.
    Ferrari, P.
    Koukorava, C.
    Krim, S.
    Struelens, L.
    RADIATION PROTECTION DOSIMETRY, 2011, 144 (1-4) : 492 - 496
  • [35] Efficiency evaluation of leaded glasses and visors for eye lens dose reduction during fluoroscopy guided interventional procedures
    Samara, Eleni Theano
    Cester, Davide
    Furlan, Miha
    Pfammatter, Thomas
    Frauenfelder, Thomas
    Stuessi, Anja
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2022, 100 : 129 - 134
  • [36] Radiation dose to the operator during fluoroscopically guided spine procedures
    Roccatagliata, Luca
    Presilla, Stefano
    Pravata, Emanuele
    Cianfoni, Alessandro
    NEURORADIOLOGY, 2017, 59 (09) : 885 - 892
  • [37] Radiation dose to the operator during fluoroscopically guided spine procedures
    Luca Roccatagliata
    Stefano Presilla
    Emanuele Pravatà
    Alessandro Cianfoni
    Neuroradiology, 2017, 59 : 885 - 892
  • [38] Surgeon eye lens dose monitoring in interventional neuroradiology, cardiovascular and radiology procedures
    Busoni, S.
    Bruzzi, M.
    Giomi, S.
    Poggiali, C.
    Quattrocchi, M.
    Betti, M.
    D'Urso, D.
    Fedeli, L.
    Mazzoni, L. N.
    Paolucci, M.
    Rossi, F.
    Taddeucci, A.
    Bettarini, S.
    Tortoli, P.
    Belli, G.
    Bernardi, L.
    Gasperi, C.
    Campanella, F.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2022, 104 : 123 - 128
  • [39] Eye Lens Radiation Dose to Nurses during Cardiac Interventional Radiology: An Initial Study
    Yamada, Ayumi
    Haga, Yoshihiro
    Sota, Masahiro
    Abe, Mitsuya
    Kaga, Yuji
    Inaba, Yohei
    Suzuki, Masatoshi
    Tada, Norio
    Zuguchi, Masayuki
    Chida, Koichi
    DIAGNOSTICS, 2023, 13 (18)
  • [40] ASSESSMENT OF EYE LENS DOSES FOR WORKERS DURING INTERVENTIONAL RADIOLOGY PROCEDURES
    Urboniene, A.
    Sadzeviciene, E.
    Ziliukas, J.
    RADIATION PROTECTION DOSIMETRY, 2015, 165 (1-4) : 299 - 303