Eye lens dose in spine surgeons during myelography procedures: a dosimetry study

被引:0
|
作者
Nagamoto, Keisuke [1 ]
Kawachino, Tomonori [1 ]
Suetsugu, Yoshiki [1 ]
Saruwatari, Hiromi [1 ]
Nihei, Shun-ichi [2 ]
Kunugita, Naoki [3 ]
机构
[1] Hosp Univ Occupat & Environm Hlth, Dept Radiol, 1-1 Iseigaoka,Yahatanishi Ku, Fukuoka, Japan
[2] Univ Occupat & Environm Hlth, Dept Emergency & Intens Care Med, 1-1 Iseigaoka,Yahatanishi Ku, Fukuoka, Japan
[3] Univ Occupat & Environm Hlth, Dept Occupat & Community Hlth, Nursing Sch Hlth Sci, 1-1 Iseigaoka,Yahatanishi Ku, Fukuoka, Japan
关键词
3 mm dose equivalent; Hp(3); lens dose; myelography; spine surgeons; radio-photoluminescence glass dosimeters; x-ray tube system; INTERVENTIONAL RADIOLOGY; RADIATION-EXPOSURE; FLUOROSCOPY; TRIAL; CT;
D O I
10.1088/1361-6498/ad0b3a
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
To determine the eye lens dose (3 mm dose equivalent [Hp(3)]) received by spine surgeons during myelography and evaluate the effectiveness of radiation-protective glasses and x-ray tube system positioning in reducing radiation exposure. This study included spine surgeons who performed myelography using over- or under-table x-ray tube systems. Hp(3) was measured for each examination using a radio-photoluminescence glass dosimeter (GD-352M) mounted on radiation-protective glass. This study identified significantly high Hp(3) levels, especially in the right eye lens in spinal surgeons. The median Hp(3) values in the right eye were 524 (391-719) and 58 (42-83) mu Sv/examination for over- and under-table x-ray tube systems, respectively. Further, Hp(3)AK, which was obtained by dividing the cumulative air kerma from Hp(3), was 8.09 (6.69-10.21) and 5.11 (4.06-6.31) mu Sv mGy-1 for the over- and under-table x-ray tube systems, respectively. Implementing radiation-protective glasses resulted in dose reduction rates of 54% (50%-57%) and 54% (51%-60%) for the over- and under-table x-ray tube systems, respectively. The use of radiation protection glasses significantly reduced the radiation dose in the eye lens during myelography, with the most effective measures being the combination of using radiation protection glasses and an under-table x-ray tube system.
引用
收藏
页数:10
相关论文
共 50 条
  • [41] Influence of standing positions and beam projections on effective dose and eye lens dose of anaesthetists in interventional procedures
    Kong, Y.
    Struelens, L.
    Vanhavere, F.
    Vargas, C. S.
    Schoonjans, W.
    Zhuo, W. H.
    RADIATION PROTECTION DOSIMETRY, 2015, 163 (02) : 181 - 187
  • [42] Radiation dose to physicians' eye lens during interventional radiology
    Bahruddin, N. A.
    Hashim, S.
    Karim, M. K. A.
    Sabarudin, A.
    Ang, W. C.
    Salehhon, N.
    Bakar, K. A.
    13TH SOUTH-EAST ASIAN CONGRESS OF MEDICAL PHYSICS 2015 (SEACOMP), 2016, 694
  • [43] EYE LENS EXPOSURE TO MEDICAL STAFF PERFORMING ELECTROPHYSIOLOGY PROCEDURES: DOSE ASSESSMENT AND CORRELATION TO PATIENT DOSE
    Ciraj-Bjelac, Olivera
    Antic, Vojislav
    Selakovic, Jovana
    Bozovic, Predrag
    Arandjic, Danijela
    Pavlovic, Sinisa
    RADIATION PROTECTION DOSIMETRY, 2016, 172 (04) : 475 - 482
  • [44] RADIATION DOSE EMISSION DURING SPINE SURGERY WITH INSTRUMENTATION: RAISING AWARENESS IN SUBSAHARAN AFRICAN SPINE SURGEONS
    Konan, Meleine Landry
    Drogba, Kporou Landry
    N'Da, Hermann Adonis
    Meuga, Munkado Wilfried Innocent
    Tsopmene, Dongmo Marvin Richie
    Djonde, Grace Ella Armande
    Brou, N'guessan Joel
    Oka N'Dri, Dominique
    AFRICAN JOURNAL OF NEUROLOGICAL SCIENCES, 2022, 41 (02): : 85 - 93
  • [45] Estimating the whole-body effective dose and health risks as well as introducing a new easy method for eye lens dosimetry in interventional cardiology procedures
    Hatami, Alireza
    Bagheri, Mahmoud
    Falahati, Farzaneh
    Banaei, Amin
    Abedi-Firouzjah, Razzagh
    Zamani, Hamed
    Kiapour, Mohammad
    Momeni, Farideh
    METHODSX, 2020, 7
  • [46] Occupational Exposure of the Eye Lens in Interventional Procedures: How to Assess and Manage Radiation Dose
    Ciraj-Bjelac, Olivera
    Carinou, Eleftheria
    Ferrari, Paolo
    Gingaume, Merce
    Merce, Marta Sans
    O'Connor, Una
    JOURNAL OF THE AMERICAN COLLEGE OF RADIOLOGY, 2016, 13 (11) : 1347 - 1353
  • [47] Lens of the eye dose calculation for neuro-interventional procedures and CBCT scans of the head
    Xiong, Zhenyu
    Vijayan, Sarath
    Rana, Vijay
    Jain, Amit
    Rudin, Stephen
    Bednarek, Daniel R.
    MEDICAL IMAGING 2016: PHYSICS OF MEDICAL IMAGING, 2016, 9783
  • [48] Eye lens dose of medical personnel involved in fluoroscopy and interventional procedures at a Malaysian Hospital
    Wong, Jeannie Hsiu Ding
    Anem, Lydia Esther Andrew
    Tan, Suzet
    Tan, Sock Keow
    Ng, Kwan Hoong
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2019, 68 : 47 - 51
  • [49] A correlation study of eye lens dose and personal dose equivalent for interventional cardiologists
    Farah, J.
    Struelens, L.
    Dabin, J.
    Koukorava, C.
    Donadille, L.
    Jacob, S.
    Schnelzer, M.
    Auvinen, A.
    Vanhavere, F.
    Clairand, I.
    RADIATION PROTECTION DOSIMETRY, 2013, 157 (04) : 561 - 569
  • [50] Variation of eye-lens dose with variation of the location of the beam isocenter in the head during neuro-interventional fluoroscopic procedures
    Guo, Chao
    Troville, Jonathan
    Sun, Sheng-Hsuan
    Rudin, Stephen
    Bednarek, Daniel R.
    MEDICAL IMAGING 2020: PHYSICS OF MEDICAL IMAGING, 2020, 11312