Design and evaluation of a hip prosthesis simulator: A technical note

被引:3
|
作者
Fanous, Amir [1 ]
Botros, Michael [1 ]
Gholizadeh, Hossein [1 ,2 ]
Baddour, Natalie [1 ]
Lemaire, Edward D. [2 ,3 ]
机构
[1] Univ Ottawa, Dept Mech Engn, Fac Engn, 505 Smyth Rd, Ottawa, ON K1H 8M2, Canada
[2] Ottawa Hosp, Ottawa Hosp Res Inst, Rehabil Ctr, Ottawa, ON, Canada
[3] Univ Ottawa, Fac Med, Ottawa, ON, Canada
关键词
amputation; prosthesis; hip disarticulation; hemipelvectomy; rehabilitation; simulator; artificial limb; ENERGY-EXPENDITURE; DISARTICULATION; WALKING; GAIT;
D O I
10.1097/PXR.0000000000000208
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
People with a limb loss at the level of the hip or pelvis have the most difficulty returning to walking compared with those with a lower amputation. This is because their prosthesis must replace the hip, knee, and ankle joints. An adjustable hip-disarticulation/hemipelvectomy prosthesis simulator that allows able-bodied individuals to wear and assess a prosthesis can help researchers and manufacturers when designing new prosthetic components (ie, hip joints). SolidWorks computer-aided design software was used to design and simulate an adapter that can connect prosthetic components to an off-the-shelf hip abduction orthosis. The adapter was made of 1020 stainless steel and aluminium 6061-T3 with a yield strength of 276 MPa. To confirm that this adapter is strong and safe for ambulation, mechanical testing was performed using an INSTRON machine. The maximum loads generated in any activity were chosen according to the International Organization for Standardization 15032:2000 standard for hip disarticulation external prostheses. The designed adapter allowed frontal, lateral, or distal mounting of different prosthetic hip joints. Mechanical testing confirmed that the new adapter can withstand forces and moments experienced during ambulation. The hip disarticulation/hemipelvectomy prosthesis simulator is easy to use and adjustable based on each person's height and pelvic width. Furthermore, this simulator would assist rehabilitation practitioners in experiencing the use of hip-level prostheses and give them a better understanding of people using such technologies. The next step in this project is to evaluate able-bodied participant gait while using this hip simulator prosthesis with different hip joints.
引用
收藏
页码:443 / 446
页数:4
相关论文
共 50 条
  • [21] Design and Fabrication of cementless Hip Prosthesis
    Sivasankar, M.
    Aravindhan, A.
    Gowthaman, K.
    Karthigeyan, T.
    Kumar, K. Sathish
    2014 International Conference on Science Engineering and Management Research (ICSEMR), 2014,
  • [22] Topological optimization in hip prosthesis design
    Fraldi, M.
    Esposito, L.
    Perrella, G.
    Cutolo, A.
    Cowin, S. C.
    BIOMECHANICS AND MODELING IN MECHANOBIOLOGY, 2010, 9 (04) : 389 - 402
  • [23] Topological optimization in hip prosthesis design
    M. Fraldi
    L. Esposito
    G. Perrella
    A. Cutolo
    S. C. Cowin
    Biomechanics and Modeling in Mechanobiology, 2010, 9 : 389 - 402
  • [24] Design of Custom Made Prosthesis of the Hip
    Tabakovic, Slobodan
    Zeljkovic, Milan
    Milojevic, Zoran
    Zivkovic, Aleksandar
    TEHNICKI VJESNIK-TECHNICAL GAZETTE, 2019, 26 (02): : 323 - 330
  • [25] Discussion on the design of a hip joint simulator
    Viceconti, M
    Cavallotti, G
    Andrisano, AO
    Toni, A
    MEDICAL ENGINEERING & PHYSICS, 1996, 18 (03) : 234 - 240
  • [26] Technical note: Rainfall simulator runoff hydrograph analysis
    Frasier, GW
    Weltz, M
    Weltz, L
    JOURNAL OF RANGE MANAGEMENT, 1998, 51 (05): : 531 - 535
  • [27] Development of a brain simulator for intracranial targeting: Technical note
    Carolus, A.
    Hesse, M.
    Rudak, B.
    Weihe, S.
    Brenke, C.
    JOURNAL OF CLINICAL NEUROSCIENCE, 2019, 59 : 378 - 383
  • [28] Patient-specific design process and evaluation of a hip prosthesis femoral stem
    Abdelaa, Osama
    Darwish, Saied
    El-Hofy, Hassan
    Saito, Yoshio
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2019, 42 (06): : 271 - 290
  • [29] TECHNICAL NOTE - A MAINS DROP-OUT SIMULATOR
    NORWOOD, HM
    JOURNAL OF MEDICAL ENGINEERING & TECHNOLOGY, 1988, 12 (05) : 224 - 225
  • [30] Technical note: A mains drop-out simulator
    Norwood, H.M.
    Journal of Medical Engineering and Technology, 1988, 12 (05): : 224 - 225