Design and evaluation of a hip prosthesis simulator: A technical note

被引:3
|
作者
Fanous, Amir [1 ]
Botros, Michael [1 ]
Gholizadeh, Hossein [1 ,2 ]
Baddour, Natalie [1 ]
Lemaire, Edward D. [2 ,3 ]
机构
[1] Univ Ottawa, Dept Mech Engn, Fac Engn, 505 Smyth Rd, Ottawa, ON K1H 8M2, Canada
[2] Ottawa Hosp, Ottawa Hosp Res Inst, Rehabil Ctr, Ottawa, ON, Canada
[3] Univ Ottawa, Fac Med, Ottawa, ON, Canada
关键词
amputation; prosthesis; hip disarticulation; hemipelvectomy; rehabilitation; simulator; artificial limb; ENERGY-EXPENDITURE; DISARTICULATION; WALKING; GAIT;
D O I
10.1097/PXR.0000000000000208
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
People with a limb loss at the level of the hip or pelvis have the most difficulty returning to walking compared with those with a lower amputation. This is because their prosthesis must replace the hip, knee, and ankle joints. An adjustable hip-disarticulation/hemipelvectomy prosthesis simulator that allows able-bodied individuals to wear and assess a prosthesis can help researchers and manufacturers when designing new prosthetic components (ie, hip joints). SolidWorks computer-aided design software was used to design and simulate an adapter that can connect prosthetic components to an off-the-shelf hip abduction orthosis. The adapter was made of 1020 stainless steel and aluminium 6061-T3 with a yield strength of 276 MPa. To confirm that this adapter is strong and safe for ambulation, mechanical testing was performed using an INSTRON machine. The maximum loads generated in any activity were chosen according to the International Organization for Standardization 15032:2000 standard for hip disarticulation external prostheses. The designed adapter allowed frontal, lateral, or distal mounting of different prosthetic hip joints. Mechanical testing confirmed that the new adapter can withstand forces and moments experienced during ambulation. The hip disarticulation/hemipelvectomy prosthesis simulator is easy to use and adjustable based on each person's height and pelvic width. Furthermore, this simulator would assist rehabilitation practitioners in experiencing the use of hip-level prostheses and give them a better understanding of people using such technologies. The next step in this project is to evaluate able-bodied participant gait while using this hip simulator prosthesis with different hip joints.
引用
收藏
页码:443 / 446
页数:4
相关论文
共 50 条
  • [1] DESIGN AND DEVELOPMENT OF A CARDIAC SIMULATOR FOR EVALUATION OF HEART-VALVE PROSTHESIS
    STARKEY, WL
    HAGAN, BT
    SIRAK, HD
    COLLINS, JA
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1963, 46 (02): : 207 - &
  • [2] Evaluation of new hip prosthesis design with finite element analysis
    Talip Çelik
    Yasin Kişioğlu
    Australasian Physical & Engineering Sciences in Medicine, 2019, 42 : 1033 - 1038
  • [3] Evaluation of new hip prosthesis design with finite element analysis
    Talip, Celik
    Kisioglu, Yasin
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2019, 42 (04) : 1033 - 1038
  • [4] HIPS - A Virtual Reality Hip Prosthesis Implantation Simulator
    Kaluschke, Maximilian
    Weller, Rene
    Zachmann, Gabriel
    Pelliccia, Luigi
    Lorenz, Mario
    Klimant, Philipp
    Knopp, Sebastian
    Atze, Johannes P. G.
    Moeckel, Falk
    25TH 2018 IEEE CONFERENCE ON VIRTUAL REALITY AND 3D USER INTERFACES (VR), 2018, : 591 - 592
  • [5] Wear comparison between a dual mobility total hip prosthesis and a typical modular design using a hip joint simulator
    Saikko, Vesa
    Shen, Ming
    WEAR, 2010, 268 (3-4) : 617 - 621
  • [6] NOTE ON RINGS METHODS FOR TOTAL HIP PROSTHESIS
    CROZZOLI, N
    LASI, C
    MINERVA ORTOPEDICA, 1974, 25 (09): : 407 - 415
  • [7] Evaluation of a hip joint simulator
    Smith, SL
    Burgess, IC
    Unsworth, A
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 1999, 213 (H6) : 469 - 473
  • [8] New Hip Prosthesis Design and Evaluation with Using Finite Element Analysis
    Celik, Talip
    Kisioglu, Yasin
    JOURNAL OF POLYTECHNIC-POLITEKNIK DERGISI, 2022, 25 (02): : 563 - 567
  • [9] Thermoelastic femoral stress imaging for experimental evaluation of hip prosthesis design
    Hyodo, K
    Inomoto, M
    Ma, W
    Miyakawa, S
    Tateishi, T
    JSME INTERNATIONAL JOURNAL SERIES C-MECHANICAL SYSTEMS MACHINE ELEMENTS AND MANUFACTURING, 2001, 44 (04) : 1065 - 1071
  • [10] Hip Spica Table: A Technical Note
    Jain, Meet
    Ponugoti, Godhasiri
    Bhansal, Sheenam
    Shah, Shalin
    Sangole, Chinmay
    Bohra, Vikas
    Rabbi, Qaisur
    Gupta, Gaurav
    Shah, Maulin
    INDIAN JOURNAL OF ORTHOPAEDICS, 2023, 57 (08) : 1318 - 1322