摘要
Introduction/Aims In healthy subjects, we observed high amplitude motor unit potential (MUP) waveforms that resembled the cannula potential (CP) with a positive sharp wave (PSW)-like waveform. We analyzed the source of this signal, its prevalence, and its effects on the analysis of electromyographic waveforms. Methods Three channel recordings were performed to explore the contribution of the needle core and cannula to the MUP. In 7 control subjects with no neuromuscular disease, the EMG interference pattern was quantified for the amplitude asymmetry (AASM) of its positive and negative going peaks. Results The large amplitude MUP resembling a CP with a PSW-type waveform is recorded by the needle core, not the cannula. With a slight change in needle position, its amplitude decreased and the waveform had a normal appearance. Thus, it is an artifact. Such potentials (AASM > 150%) are commonly encountered (frequency 7%) and occur in all muscles and subjects in this study. Discussion The artifact is an enigma and appears related to the interaction between the needle and muscle fiber(s) similar to that in PSW signals. Failure to recognize these waveforms may result in a false impression of abnormality. AASM calculations can help confirm this artifact in automated analysis algorithms to preclude false positive neurogenic patterns.
引用
收藏
页码:216 / 222
页数:7
相关论文
共 50 条
  • [21] SIMULATION OF CONCENTRIC NEEDLE EMG MOTOR UNIT ACTION-POTENTIALS
    NANDEDKAR, SD
    SANDERS, DB
    STALBERG, EV
    ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1985, 60 (05): : P103 - P103
  • [22] DECOMPOSITION OF CONCENTRIC NEEDLE EMG MOTOR UNIT ACTION-POTENTIALS
    NANDEDKAR, SD
    SANDERS, DB
    MUSCLE & NERVE, 1989, 12 (09) : 752 - 753
  • [23] SIMULATION OF CONCENTRIC NEEDLE EMG MOTOR UNIT ACTION-POTENTIALS
    NANDEDKAR, SD
    SANDERS, DB
    STALBERG, EV
    ANDREASSEN, S
    MUSCLE & NERVE, 1988, 11 (02) : 151 - 159
  • [24] Surface EMG measurements during fMRI at 3T: Accurate EMG recordings after artifact correction
    van Duinen, H
    Zijdewind, I
    Hoogduin, H
    Maurits, N
    NEUROIMAGE, 2005, 27 (01) : 240 - 246
  • [25] QUANTITATIVE COMPARISON OF MOTOR UNIT ACTION-POTENTIALS BETWEEN MONOPOLAR AND CONCENTRIC NEEDLE RECORDINGS UNDER DIFFERENT FILTER SETTINGS OF EMG AMPLIFIER
    CHU, J
    CHEN, RC
    ARCHIVES OF PHYSICAL MEDICINE AND REHABILITATION, 1983, 64 (10): : 512 - 512
  • [26] THE CALCULATED INFLUENCE OF A METAL-ELECTRODE SHAFT IN NEEDLE EMG RECORDINGS
    GOOTZEN, THJM
    THEEUWEN, MMHJ
    STEGEMAN, DF
    IMAGES OF THE TWENTY-FIRST CENTURY, PTS 1-6, 1989, 11 : 986 - 987
  • [27] Filter design for cancellation of baseline-fluctuation in needle EMG recordings
    Rodríguez-Carreño, I
    Malanda-Trigueros, A
    Gila-Useros, L
    Navallas-Irujo, J
    Rodríguez-Falces, J
    COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2006, 81 (01) : 79 - 93
  • [28] Discriminant analysis of various concentric needle EMG and macro-EMG parameters in detecting myopathic abnormality
    Cengiz, B
    Ozdag, F
    Ulas, UH
    Odabasi, Z
    Vural, O
    CLINICAL NEUROPHYSIOLOGY, 2002, 113 (09) : 1423 - 1428
  • [29] An automated ECG-artifact removal method for trunk muscle surface EMG recordings
    Mak, Joseph N. F.
    Hu, Yong
    Luk, Keith D. K.
    MEDICAL ENGINEERING & PHYSICS, 2010, 32 (08) : 840 - 848
  • [30] Single-fiber EMG with a concentric needle electrode: Validation in myasthenia gravis
    Sarrigiannis, PG
    Kennett, RP
    Read, S
    Farrugia, ME
    MUSCLE & NERVE, 2006, 33 (01) : 61 - 65