Optimization of the design of a single-sided magnetic resonance magnet array for improved T2 estimation accuracy

被引:0
|
作者
Miao, Ruixin [1 ,2 ,3 ]
Ren, Chunpeng [1 ,2 ,3 ]
Jiang, Chuandong [1 ,2 ,3 ]
Wang, Qingyue [1 ,2 ,3 ]
Wang, Yunzhi [1 ,2 ,3 ]
Yu, Xiangqian [4 ]
机构
[1] Minist Educ China, Key Lab Geophys Explorat Equipment, Changchun 130061, Peoples R China
[2] Jilin Univ, Coll Instrumentat & Elect Engn, Changchun 130061, Peoples R China
[3] Natl Geophys Explorat Equipment Engn Res Ctr, Changchun 130061, Peoples R China
[4] Hohai Univ, Sch Earth Sci & Engn, Nanjing 211100, Peoples R China
基金
中国国家自然科学基金;
关键词
low-field NMR; SSMR magnet array; transverse relaxation time; spin echo; SQP algorithm; UNILATERAL MAGNET; NMR SENSOR;
D O I
10.1088/1361-6501/ad9e16
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Single-sided magnetic resonance (SSMR) technology has a compact structure and great potential for use in mobile and nondestructive detection of large aqueous materials. As the detection depth increases, the magnetic field gradient generated by the magnetic resonance magnet placed on one side weakens, leading to a large distortion of T2, thereby significantly impeding accurate estimation of the pore size distribution within the detection target. In this study, to address this problem, the distortion in the region of interest (ROI) is significantly reduced by optimizing the size and arrangement of the SSMR magnet array. First, to evaluate the T2 distortion within the ROI, we developed a T2 distortion loss function based on alpha and theta, which are key parameters of the spin echo (SE) signal. We then combined the nonlinear constraints with the loss function to form a Lagrangian function, which was solved via the Newton iteration method to obtain the magnet array parameters. The magnet array is called the AT magnet array. The T2 distortion was quantitatively evaluated by calculating the SE signal generated by the AT magnet array in the ROI. Furthermore, we compared the detection results obtained with the AT and the semi-annular (SA) magnet array. The results showed that within the 3 cm x 3 cm target range of a 10 cm depth, the AT magnet array had a smaller T2 distortion of only 1.3%. After the ROI was expanded to 5 cm x 5 cm, the T2 distortion of the SE signal measured by the AT magnet array was 6%, which met the SSMR detection accuracy requirements in this range. Therefore, the optimized design of the magnet array based on alpha and theta provides a theoretical basis for obtaining SE signals with low T2 distortion at large depths and over large areas via SSMR.
引用
收藏
页数:14
相关论文
共 50 条
  • [41] Improved Interpretation of Nuclear Magnetic Resonance T1 and T2 Distributions for Permeability Prediction: Simulation of Diffusion Coupling for a Fractal Cluster of Pores
    H. Pape
    C. Clauser
    Pure and Applied Geophysics, 2009, 166 : 949 - 968
  • [42] Accuracy, repeatability, and reproducibility of T1 and T2 relaxation times measurement by 3D magnetic resonance fingerprinting with different dictionary resolutions
    Wicaksono, Krishna Pandu
    Fushimi, Yasutaka
    Nakajima, Satoshi
    Sakata, Akihiko
    Okuchi, Sachi
    Hinoda, Takuya
    Oshima, Sonoko
    Otani, Sayo
    Tagawa, Hiroshi
    Urushibata, Yuta
    Nakamoto, Yuji
    EUROPEAN RADIOLOGY, 2023, 33 (04) : 2895 - 2904
  • [43] Accuracy, repeatability, and reproducibility of T1 and T2 relaxation times measurement by 3D magnetic resonance fingerprinting with different dictionary resolutions
    Krishna Pandu Wicaksono
    Yasutaka Fushimi
    Satoshi Nakajima
    Akihiko Sakata
    Sachi Okuchi
    Takuya Hinoda
    Sonoko Oshima
    Sayo Otani
    Hiroshi Tagawa
    Yuta Urushibata
    Yuji Nakamoto
    European Radiology, 2023, 33 : 2895 - 2904
  • [44] High Efficient Reconstruction of Single-Shot Magnetic Resonance T2 Mapping Through Overlapping Echo Detachment and DenseNet
    Wang, Chao
    Wu, Yawen
    Ding, Xinghao
    Huang, Yue
    Cai, Congbo
    NEURAL INFORMATION PROCESSING (ICONIP 2018), PT VI, 2018, 11306 : 408 - 418
  • [45] The influence of reconstruction slice thickness on the accuracy of T2 sagittal sequence in the Magnetic Resonance Imaging diagnosis of optic neuropathy, in patients with multiple sclerosis
    Healy, G. M.
    Fleming, H.
    Redmond, C. E.
    Carroll, A. G.
    McGuigan, C.
    Killeen, R. P.
    MULTIPLE SCLEROSIS JOURNAL, 2016, 22 : 558 - 559
  • [46] Automatic cardiac T2*relaxation time estimation from magnetic resonance images using region growing method with automatically initialized seed points
    Wantanajittikul, Kittichai
    Theera-Umpon, Nipon
    Saekho, Suwit
    Auephanwiriyakul, Sansanee
    Phrommintikul, Arintaya
    Leemasawat, Krit
    COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2016, 130 : 76 - 86
  • [47] The Sensitivity, Specificity, and Accuracy of Contrast-Enhanced T1-Weighted Image, T2*-Weighted Image, and Magnetic Resonance Venography in Diagnosis of Cerebral Venous Sinus Thrombosis
    Jalli, Reza
    Zarei, Fatemeh
    Farahangiz, Siamak
    Khaleghi, Fatemeh
    Petramfar, Peyman
    Borhani-Haghighi, Afshin
    Yadollahikhales, Golnaz
    JOURNAL OF STROKE & CEREBROVASCULAR DISEASES, 2016, 25 (08): : 2083 - 2086
  • [48] Design and synthesis of Gd(III) magnetic resonance imaging (MRI) contrast agent ligand caps for hydroxypyridinone (HOPO) oxygen donor chelators and their T1 and T2 relaxivity
    Peretz, Ryan H.
    Pailloux, Sylvie L.
    Klemm, Piper J.
    Raymond, Kenneth N.
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2011, 241
  • [49] Estimation of liver function using T2* mapping on gadolinium ethoxybenzyl diethylenetriamine pentaacetic acid enhanced magnetic resonance imaging (vol 81, pg 1460, 2012)
    Katsube, Takashi
    Okada, Masahiro
    Kumano, Seishi
    Imaoka, Izumi
    Kagawa, Yuki
    Hori, Masatoshi
    Ishii, Kazunari
    Tanigawa, Noboru
    Imai, Yasuharu
    Kudo, Masatoshi
    Murakami, Takamichi
    EUROPEAN JOURNAL OF RADIOLOGY, 2013, 82 (06) : 1039 - 1039
  • [50] Optimization image of magnetic resonance imaging (MRI) T2 fast spin echo (FSE) with variation echo train length (ETL) on the rupture tendon achilles case
    Muzamil, Akhmad
    Firmansyah, Achmad Haries
    INTERNATIONAL CONFERENCE ON PHYSICAL INSTRUMENTATION AND ADVANCED MATERIALS, 2017, 853