Optimization of the design of a single-sided magnetic resonance magnet array for improved T2 estimation accuracy

被引:0
|
作者
Miao, Ruixin [1 ,2 ,3 ]
Ren, Chunpeng [1 ,2 ,3 ]
Jiang, Chuandong [1 ,2 ,3 ]
Wang, Qingyue [1 ,2 ,3 ]
Wang, Yunzhi [1 ,2 ,3 ]
Yu, Xiangqian [4 ]
机构
[1] Minist Educ China, Key Lab Geophys Explorat Equipment, Changchun 130061, Peoples R China
[2] Jilin Univ, Coll Instrumentat & Elect Engn, Changchun 130061, Peoples R China
[3] Natl Geophys Explorat Equipment Engn Res Ctr, Changchun 130061, Peoples R China
[4] Hohai Univ, Sch Earth Sci & Engn, Nanjing 211100, Peoples R China
基金
中国国家自然科学基金;
关键词
low-field NMR; SSMR magnet array; transverse relaxation time; spin echo; SQP algorithm; UNILATERAL MAGNET; NMR SENSOR;
D O I
10.1088/1361-6501/ad9e16
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Single-sided magnetic resonance (SSMR) technology has a compact structure and great potential for use in mobile and nondestructive detection of large aqueous materials. As the detection depth increases, the magnetic field gradient generated by the magnetic resonance magnet placed on one side weakens, leading to a large distortion of T2, thereby significantly impeding accurate estimation of the pore size distribution within the detection target. In this study, to address this problem, the distortion in the region of interest (ROI) is significantly reduced by optimizing the size and arrangement of the SSMR magnet array. First, to evaluate the T2 distortion within the ROI, we developed a T2 distortion loss function based on alpha and theta, which are key parameters of the spin echo (SE) signal. We then combined the nonlinear constraints with the loss function to form a Lagrangian function, which was solved via the Newton iteration method to obtain the magnet array parameters. The magnet array is called the AT magnet array. The T2 distortion was quantitatively evaluated by calculating the SE signal generated by the AT magnet array in the ROI. Furthermore, we compared the detection results obtained with the AT and the semi-annular (SA) magnet array. The results showed that within the 3 cm x 3 cm target range of a 10 cm depth, the AT magnet array had a smaller T2 distortion of only 1.3%. After the ROI was expanded to 5 cm x 5 cm, the T2 distortion of the SE signal measured by the AT magnet array was 6%, which met the SSMR detection accuracy requirements in this range. Therefore, the optimized design of the magnet array based on alpha and theta provides a theoretical basis for obtaining SE signals with low T2 distortion at large depths and over large areas via SSMR.
引用
收藏
页数:14
相关论文
共 50 条
  • [21] An expansion of the field modulus suitable for the description of strong field gradients in axisymmetric magnetic fields: Application to single-sided magnet design, field mapping and STRAFI
    Hugon, Cedric
    Aubert, Guy
    Sakellariou, Dimitris
    JOURNAL OF MAGNETIC RESONANCE, 2012, 214 : 124 - 134
  • [22] High Accuracy of Magnetic Resonance T2*- Weighted Sequences for Detection of Intracranial Arterial Thrombus.
    Liberato, Afonso P.
    Shah, Santosh
    Maza, Noor
    Barnaure, Isabelle
    Gonzalez, Ramon G.
    Romero, Javier M.
    STROKE, 2016, 47
  • [23] Accuracy of T2 magnetic resonance assays as point-of-care methods in the intensive care unit
    Vrettou, C. S.
    Douka, E.
    Perivolioti, E. P.
    Vassiliou, A. G.
    Sarri, A.
    Giannopoulou, V.
    Trigkidis, K. K.
    Jahaj, E.
    Dimopoulou, I.
    Kotanidou, A.
    JOURNAL OF HOSPITAL INFECTION, 2023, 139 : 240 - 248
  • [24] Simultaneous estimation of PD, T1, T2, T2*, and ΔB0 using magnetic resonance fingerprinting with background gradient compensation
    Hong, Taehwa
    Han, Dongyeob
    Kim, Dong-Hyun
    MAGNETIC RESONANCE IN MEDICINE, 2019, 81 (04) : 2614 - 2623
  • [25] On the accuracy and precision of cardiac magnetic resonance T2 mapping: A high-resolution radial study using adiabatic T2 preparation at 3T
    Bano, Wajiha
    Feliciano, Helene
    Coristine, Andrew J.
    Stuber, Matthias
    van Heeswijk, Ruud B.
    MAGNETIC RESONANCE IN MEDICINE, 2017, 77 (01) : 159 - 169
  • [26] Improved Detection of Parenchymal Cysticercal Lesions in Neurocysticercosis with T2*-weighted Angiography Magnetic Resonance Imaging
    Verma, Avantika
    Awasthi, Rishi
    Prasad, Kashi N.
    Soni, Neetu
    Singh, Kamini
    Paliwal, Vimal K.
    Lal, Hira
    Pandey, Chandra M.
    Gupta, Rakesh K.
    ACADEMIC RADIOLOGY, 2012, 19 (08) : 958 - 964
  • [27] Improved initial value estimation for short echo time magnetic resonance spectroscopy spectral analysis using short T2 signal attenuation
    Steinberg, Jeffrey
    Soher, Brian J.
    MAGNETIC RESONANCE IN MEDICINE, 2012, 67 (05) : 1195 - 1202
  • [28] The Influence of the Analysis Technique on Myocardial T2 Estimation Using Cardiac Magnetic Resonance Imaging (CMR)
    Hammad, Muhammed S.
    Khalifa, Ayman M.
    Al-Atabany, W.
    Ibrahim, El-Sayed H.
    2017 IEEE 17TH INTERNATIONAL CONFERENCE ON BIOINFORMATICS AND BIOENGINEERING (BIBE), 2017, : 317 - 320
  • [29] Magnetic resonance fingerprinting for simultaneous renal T1 and T2*mapping in a single breath-hold
    Hermann, Ingo
    Chacon-Caldera, Jorge
    Brumer, Irene
    Rieger, Benedikt
    Weingartner, Sebastian
    Schad, Lothar R.
    Zoellner, Frank G.
    MAGNETIC RESONANCE IN MEDICINE, 2020, 83 (06) : 1940 - 1948
  • [30] Improved T2 and diffusion maps from wavelet de-noised magnetic resonance imaging data
    Sarty, GE
    Kendall, EJ
    PROCEEDINGS OF THE 18TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOL 18, PTS 1-5, 1997, 18 : 1113 - 1114