Optimization of the design of a single-sided magnetic resonance magnet array for improved T2 estimation accuracy

被引:0
|
作者
Miao, Ruixin [1 ,2 ,3 ]
Ren, Chunpeng [1 ,2 ,3 ]
Jiang, Chuandong [1 ,2 ,3 ]
Wang, Qingyue [1 ,2 ,3 ]
Wang, Yunzhi [1 ,2 ,3 ]
Yu, Xiangqian [4 ]
机构
[1] Minist Educ China, Key Lab Geophys Explorat Equipment, Changchun 130061, Peoples R China
[2] Jilin Univ, Coll Instrumentat & Elect Engn, Changchun 130061, Peoples R China
[3] Natl Geophys Explorat Equipment Engn Res Ctr, Changchun 130061, Peoples R China
[4] Hohai Univ, Sch Earth Sci & Engn, Nanjing 211100, Peoples R China
基金
中国国家自然科学基金;
关键词
low-field NMR; SSMR magnet array; transverse relaxation time; spin echo; SQP algorithm; UNILATERAL MAGNET; NMR SENSOR;
D O I
10.1088/1361-6501/ad9e16
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Single-sided magnetic resonance (SSMR) technology has a compact structure and great potential for use in mobile and nondestructive detection of large aqueous materials. As the detection depth increases, the magnetic field gradient generated by the magnetic resonance magnet placed on one side weakens, leading to a large distortion of T2, thereby significantly impeding accurate estimation of the pore size distribution within the detection target. In this study, to address this problem, the distortion in the region of interest (ROI) is significantly reduced by optimizing the size and arrangement of the SSMR magnet array. First, to evaluate the T2 distortion within the ROI, we developed a T2 distortion loss function based on alpha and theta, which are key parameters of the spin echo (SE) signal. We then combined the nonlinear constraints with the loss function to form a Lagrangian function, which was solved via the Newton iteration method to obtain the magnet array parameters. The magnet array is called the AT magnet array. The T2 distortion was quantitatively evaluated by calculating the SE signal generated by the AT magnet array in the ROI. Furthermore, we compared the detection results obtained with the AT and the semi-annular (SA) magnet array. The results showed that within the 3 cm x 3 cm target range of a 10 cm depth, the AT magnet array had a smaller T2 distortion of only 1.3%. After the ROI was expanded to 5 cm x 5 cm, the T2 distortion of the SE signal measured by the AT magnet array was 6%, which met the SSMR detection accuracy requirements in this range. Therefore, the optimized design of the magnet array based on alpha and theta provides a theoretical basis for obtaining SE signals with low T2 distortion at large depths and over large areas via SSMR.
引用
收藏
页数:14
相关论文
共 50 条
  • [11] Design optimization of a low-speed single-sided linear induction motor for improved efficiency and power factor
    Isfahani, A. Hassanpour
    Ebrahimi, B. M.
    Lesani, H.
    IEEE TRANSACTIONS ON MAGNETICS, 2008, 44 (02) : 266 - 272
  • [12] Electromagnetic Design Optimization of Single-Sided Linear Induction Motor for Improved Drive Performance Based on Linear Metro Application
    Hu, Dong
    Xu, Wei
    Qu, Ronghai
    2014 Australasian Universities Power Engineering Conference (AUPEC), 2014,
  • [13] Improved Detection of Cardiac Sarcoidosis Using Magnetic Resonance with Myocardial T2 Mapping
    Crouser, Elliott D.
    Ono, Chikako
    Tam Tran
    He, Xin
    Raman, Subha V.
    AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2014, 189 (01) : 109 - 112
  • [14] Improved survival of thalassaemia major in the UK and relation to T2* cardiovascular magnetic resonance
    Bernadette Modell
    Maren Khan
    Matthew Darlison
    Mark A Westwood
    David Ingram
    Dudley J Pennell
    Journal of Cardiovascular Magnetic Resonance, 10
  • [15] Improved survival of thalassaemia major in the UK and relation to T2* cardiovascular magnetic resonance
    Modell, Bernadette
    Khan, Maren
    Darlison, Matthew
    Westwood, Mark A.
    Ingram, David
    Pennell, Dudley J.
    JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE, 2008, 10 (1)
  • [16] Improved. T2*assessment in liver iron overload by magnetic resonance imaging
    Positano, Vincenzo
    Salani, Benedetta
    Pepe, Alessia
    Santarelli, Maria Filomena
    De Marchi, Daniele
    Ramazzotti, Anna
    Favilli, Brunella
    Cracolici, Eliana
    Midiri, Massimo
    Clanciulli, Paolo
    Lombardi, Massimo
    Landini, Luigi
    MAGNETIC RESONANCE IMAGING, 2009, 27 (02) : 188 - 197
  • [17] ITERATIVE ESTIMATION OF T2 TO CORRECT ECHO PLANAR MAGNETIC-RESONANCE IMAGES
    SINGH, M
    OSHIO, K
    BRECHNER, RR
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1990, 37 (02) : 795 - 799
  • [18] Renal Iron Load Estimation in Thalassemia Patients Using T2* Magnetic Resonance Imaging
    Afshan, Shirkavand
    Parisa, Mokhtari Hesari
    Shahram, Akhlaghpoor
    Azita, Azarkeivan
    Hashemieh, Mozhgan
    INTERNATIONAL JOURNAL OF MEDICAL RESEARCH & HEALTH SCIENCES, 2019, 8 (04): : 182 - 189
  • [19] B1-insensitive T2 preparation for improved coronary magnetic resonance angiography at 3 T
    Nezafat, R
    Stuber, M
    Ouwerkerk, R
    Gharib, AM
    Desai, MY
    Pettigrew, RI
    MAGNETIC RESONANCE IN MEDICINE, 2006, 55 (04) : 858 - 864
  • [20] A phenomenological approach on the influence of paramagnetic iron in cement stone on 2D T1-T2 relaxation in single-sided 1H nuclear magnetic resonance
    Holthausen, Robert Schulte
    Raupach, Michael
    CEMENT AND CONCRETE RESEARCH, 2019, 120 : 279 - 293