Effect of metallic materials on magnetic resonance image uniformity: a quantitative experimental study

被引:0
|
作者
Shimamoto, Hiroaki [1 ]
Felemban, Doaa [2 ]
Uchimoto, Yuka [1 ]
Matsuda, Nobuhiko [1 ]
Takagawa, Naoko [1 ]
Takeshita, Ami [1 ]
Iwamoto, Yuri [1 ]
Okahata, Ryoko [1 ]
Tsujimoto, Tomomi [1 ]
Kreiborg, Sven [3 ,4 ,5 ]
Mallya, Sanjay M. [6 ]
Yang, Fan-pei Gloria [1 ,7 ,8 ]
机构
[1] Osaka Univ, Grad Sch Dent, Dept Oral & Maxillofacial Radiol, 1-8 Yamadaoka, Suita, Osaka 5650871, Japan
[2] Taibah Univ, Coll Dent, Dept Oral & Maxillofacial Diagnost Sci, Div Oral Radiol, POB 2898, Madinah 43353, Saudi Arabia
[3] Univ Copenhagen, Fac Hlth & Med Sci, Sch Dent, Dept Pediat Dent & Clin Genet, Norre 20, DK-2200 Copenhagen, Denmark
[4] Univ Copenhagen, Dept Appl Math & Comp Sci, Sch Dent, 3D Craniofacial Image Res Lab, Norre 20, DK-2200 Copenhagen, Denmark
[5] Tech Univ Denmark, Copenhagen Univ Hosp, Rigshosp, Ctr Head & Orthoped, Norre 20, DK-2200 Copenhagen, Denmark
[6] UCLA Sch Dent, Sect Oral & Maxillofacial Radiol, 10833 Conte Ave, Los Angeles, CA 90095 USA
[7] Natl Tsing Hua Univ, Dept Foreign Languages & Literature, 101,Sect 2,Guangfu Rd, Hsinchu 300013, Taiwan
[8] Natl Tsing Hua Univ, Ctr Cognit & Mind Sci, 101,Sect 2,Guangfu Rd, Hsinchu 300013, Taiwan
关键词
MRI; Metallic materials; Image uniformity; Quantitative study; COIL;
D O I
10.1007/s11282-024-00778-y
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
ObjectiveTo assess quantitatively the effect of metallic materials on MR image uniformity using a standardized method.MethodsSix types of 1 cm cubic metallic materials (i.e., Au, Ag, Al, Au-Ag-Pd alloy, Ti, and Co-Cr alloy) embedded in a glass phantom filled were examined and compared with no metal condition inserted as a reference. The phantom was scanned five times under each condition using a 1.5-T MR superconducting magnet scanner with an 8-channel phased-array brain coil and head and neck coil. For each examination, the phantom was scanned in three planes: axial, coronal, and sagittal using T1-weighted spin echo (SE) and gradient echo (GRE) sequences in accordance with the American Society for Testing and Materials (ASTM) F2119-07 standard. Image uniformity was assessed using the non-uniformity index (NUI), which was developed by the National Electrical Manufacturers Association (NEMA), as an appropriate standardized measure for investigating magnetic field uniformity.ResultsT1-GRE images with Co-Cr typically elicited the lowest uniformity, followed by T1-GRE images with Ti, while all other metallic materials did not affect image uniformity. In particular, T1-GRE images with Co-Cr showed significantly higher NUI values as far as 6.6 cm at maximum equivalent to 11 slices centering around it in comparison with the measurement uncertainty from images without metallic materials.ConclusionWe found that MR image uniformity was influenced by the scanning sequence and coil type when Co-Cr and Ti were present. It is assumed that the image non-uniformity in Co-Cr and Ti is caused by their high magnetic susceptibility.
引用
收藏
页码:78 / 87
页数:10
相关论文
共 50 条
  • [41] THE EFFECT OF IMAGE DIGITIZATION AND COMPRESSION ON IMAGE QUALITY - A QUANTITATIVE STUDY
    SIKES, RA
    HINZ, TA
    ALBERTH, DP
    ADAMS, DE
    KINSEY, JA
    INVESTIGATIVE RADIOLOGY, 1985, 20 (06) : S38 - S38
  • [42] Experimental investigation of the effect of metallic interstitial materials on thermal contact conductance
    Villanueva, E
    Madhusudana, C
    HEAT AND MASS TRANSFER AUSTRALASIA 1996, 1998, : 161 - 168
  • [43] Spin resonance in soft magnetic composite materials: a surprising effect of the magnetic load
    Chevalier, A
    Le Floc'h, F
    Mattei, JL
    Talbot, P
    Le Floc'h, M
    JOURNAL DE PHYSIQUE IV, 1998, 8 (P2): : 417 - 420
  • [44] Spin resonance in soft magnetic composite materials: a surprising effect of the magnetic load
    Chevalier, A.
    Le Floc'h, F.
    Mattei, J.L.
    Talbot, Ph.
    Le Floc'h, M.
    Journal De Physique. IV : JP, 1998, 8 (02): : 417 - 420
  • [45] Nuclear magnetic resonance as a quantitative tool to study interactions in biomacromolecules
    Grzesiek, S.
    Allan, M.
    Cordier, F.
    Haeussinger, D.
    Jensen, P.
    Kahmann, J.
    Meier, S.
    Sass, H. J.
    PURE AND APPLIED CHEMISTRY, 2005, 77 (08) : 1409 - 1424
  • [46] A QUANTITATIVE MAGNETIC-RESONANCE-IMAGING STUDY OF PATIENTS WITH SCHIZOPHRENIA
    KAWASAKI, Y
    MAEDA, Y
    URATA, K
    HIGASHIMA, M
    YAMAGUCHI, N
    SUZUKI, M
    TAKASHIMA, T
    IDE, Y
    EUROPEAN ARCHIVES OF PSYCHIATRY AND CLINICAL NEUROSCIENCE, 1993, 242 (05) : 268 - 272
  • [47] Septum Pellucidum in schizophrenia: A quantitative magnetic resonance imaging study
    Bersani, G
    Iannitelli, A
    Pacitti, F
    Pancheri, P
    SCHIZOPHRENIA RESEARCH, 2001, 49 (1-2) : 151 - 151
  • [48] The spectrum of hippocampal sclerosis: A quantitative magnetic resonance imaging study
    VanPaesschen, W
    Connelly, A
    King, MD
    Jackson, GD
    Duncan, JS
    ANNALS OF NEUROLOGY, 1997, 41 (01) : 41 - 51
  • [49] QUANTITATIVE NUCLEAR-MAGNETIC-RESONANCE IMAGING - CHARACTERIZATION OF EXPERIMENTAL CEREBRAL EDEMA
    BARNES, D
    MCDONALD, WI
    JOHNSON, G
    TOFTS, PS
    LANDON, DN
    JOURNAL OF NEUROLOGY NEUROSURGERY AND PSYCHIATRY, 1987, 50 (02): : 125 - 133
  • [50] AN EXPERIMENTAL STUDY ON THE EFFECT OF COMPRESSION FORCE UNIFORMITY IN ELECTROSURGICAL TISSUE WELDING
    Li, Xiaoran
    Borduin, Russell
    Chen, Roland
    Li, Wei
    PROCEEDINGS OF THE ASME 11TH INTERNATIONAL MANUFACTURING SCIENCE AND ENGINEERING CONFERENCE, 2016, VOL 2, 2016,