Study of trapping levels in doped HgI2 radiation detectors

被引:0
|
作者
Sandia Natl Lab, Livermore, United States [1 ]
机构
来源
关键词
Alpha particles - Butane - Charge carriers - Crystal defects - Crystal impurities - Crystal lattices - Deep level transient spectroscopy - Mercury compounds - Naphthalene - Paraffins - Polyethylene glycols - Polyethylenes;
D O I
暂无
中图分类号
学科分类号
摘要
The transient charge technique, TCT, was used to measure the trapping levels of pure and doped HgI2 detectors over the temperature range of 170-300 K. The dopants were excess of Hg and I2, naphthalene, butane, heavy paraffins, polyethylene and polyethylene glycol. The charge carriers were induced by α particles as well as by a UV nitrogen (337 nm) pulsed laser beam. The TCT determined trapping levels are compared with levels reported in literature from other solid state measurements.
引用
收藏
相关论文
共 50 条
  • [1] Study of trapping levels in doped HgI2 radiation detectors
    Hermon, H
    Schieber, M
    Roth, M
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1996, 380 (1-2): : 10 - 13
  • [2] Study on the performance of HgI2 semiconductor detectors
    Zhang, Lan
    Li, Yuanjing
    Zheng, Xiaocui
    Deng, Zhi
    2010 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD (NSS/MIC), 2010, : 3937 - 3941
  • [3] Study of trapping levels in α-HgI2 crystals by UV and isothermal current measurements
    Xu, Gang
    Jie, Wanqi
    Li, Gaohong
    Zha, Gangqiang
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 613 (02): : 277 - 279
  • [4] Growth of mercuric iodide (HgI2) for nuclear radiation detectors
    van den Berg, L.
    Schnepple, W.F.
    NASA Technical Memorandum, 1988, 1 (4069): : 269 - 286
  • [5] FABRICATION OF HGI2 NUCLEAR RADIATION DETECTORS BY MACHINE CLEAVING
    LEVI, A
    BURGER, A
    SCHIEBER, M
    VANDENBERG, L
    YELON, WB
    ALKIRE, RW
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1983, 213 (01): : 31 - 34
  • [6] FABRICATION OF HGI2 NUCLEAR DETECTORS
    SCHIEBER, M
    NUCLEAR INSTRUMENTS & METHODS, 1977, 144 (03): : 469 - 477
  • [7] Evaluation of pixellated HgI2 detectors
    Hitomi, K
    Meng, LJ
    He, Z
    2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 4456 - 4460
  • [8] RECENT IMPROVEMENTS IN HGI2 DETECTORS
    VANDENBERG, L
    WHITED, RC
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1978, 25 (01) : 395 - 397
  • [9] HgI2 Flat Panel Radiation Detectors for Medical Imaging Acquisition
    Oh, Kyungmin
    Yun, MinSeok
    Kim, Minwoo
    Jang, Kiwon
    Nam, Sanghee
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, VOL 25, PT 2 - DIAGNOSTIC IMAGING, 2009, 25 : 590 - 593
  • [10] FABRICATION OF HgI2 NUCLEAR RADIATION DETECTORS BY MACHINE CLEAVING.
    Levi, A.
    Burger, A.
    Schieber, M.
    van den Berg, L.
    Yelon, W.B.
    Alkire, R.W.
    1600, (213):