Characterization of a pediatric rotary blood pump

被引:2
|
作者
de Melo T.R. [1 ]
Vasconcelos F.J.S. [1 ]
Ribeiro L.H.R.D. [1 ]
Bacht S. [2 ]
Cestari I.A. [2 ]
Neto J.S.D.R. [1 ]
Lima A.M.N. [1 ]
机构
[1] Federal University of Campina Grande, Campina Grande, PB
[2] Bioengineering Division, Heart Institute, São Paulo, SP
基金
巴西圣保罗研究基金会;
关键词
Centrifugal flow pump; Lumped parameter model; Rotary blood pump; System identification; Ventricular assist device;
D O I
10.1590/2446-4740.180046
中图分类号
学科分类号
摘要
Introduction: A ventricular assist device (VAD) is an electromechanical pump used to treat heart failures. For designing the physiological control system for a VAD, one needs a mathematical model and its related parameters. This paper presents a characterization procedure for determining the model parameter values of the electrical, mechanical, and hydraulic subsystems of a pediatric Rotary Blood Pump (pRBP). Methods: An in vitro test setup consisting of a pRBP prototype, a motor driver module, an acrylic reservoir, mechanical resistance and tubings, pressure and fluid flow sensors, and data acquisition, processing, and visualization system. The proposed procedure requires a set of experimental tests, and a parameter estimation algorithm for determining the model parameters values. Results: The operating limits of the pRBP were identified from the steady-state data. The relationship between the pressure head, flow rate, and the rotational speed of the pRBP was found from the static tests. For the electrical and mechanical subsystems, the dc motor model has a viscous friction coefficient that varies nonlinearly with the flow. For the hydraulic subsystem, the pressure head is assumed to be a sum of terms related to the resistance, the inertance, the friction coefficient, and the pump speed. Conclusion: The proposed methodology was successfully applied to the characterization of the pRBP. The combined use of static and dynamic tests provided a precise lumped parameter model for representing the pRBP dynamics. The agreement, regarding mean squared deviation, between experimental and simulated results demonstrates the correctness and feasibility of the characterization procedure. © 2018, Brazilian Society of Biomedical Engineering. All rights reserved.
引用
收藏
页码:299 / 309
页数:10
相关论文
共 50 条
  • [41] A Passive Magnetically and Hydrodynamically Suspended Rotary Blood Pump
    Stoiber, Martin
    Grasl, Christian
    Pirker, Stefan
    Raderer, Franz
    Schistek, Roland
    Huber, Leopold
    Gittler, Philipp
    Schima, Heinrich
    ARTIFICIAL ORGANS, 2009, 33 (03) : 250 - 257
  • [42] Numerical analysis of the flow characteristics of rotary blood pump
    Morsi Y.S.
    Yang W.
    Witt P.J.
    Ahmed A.M.
    Umezu M.
    Journal of Artificial Organs, 2001, 4 (1) : 54 - 60
  • [43] Impact of Hyperthermal Rotary Blood Pump Surfaces on Blood Clotting Behavior: An Approach
    Hamilton, Kathrin F.
    Schlanstein, Peter C.
    Mager, Ilona
    Schmitz-Rode, Thomas
    Steinseifer, Ulrich
    ARTIFICIAL ORGANS, 2009, 33 (09) : 740 - 748
  • [44] Viscoelasticity of pediatric blood and its implications for the testing of a Pulsatile pediatric blood pump
    Long, Jennifer A.
    Uendar, Akif
    Manning, Keefe B.
    Deutsch, Steven
    ASAIO JOURNAL, 2005, 51 (05) : 563 - 566
  • [45] Impeller behavior and displacement of the VentrAssist implantable rotary blood pump
    Chung, MKH
    Zhang, N
    Tansley, TD
    Woodard, JC
    ARTIFICIAL ORGANS, 2004, 28 (03) : 287 - 297
  • [46] Analysis of the preload and afterload sensitivity of the sputnik rotary blood pump
    Petukhov D.S.
    Telyshev D.V.
    Biomedical Engineering, 2016, 49 (6) : 362 - 365
  • [47] Noninvasive pulsatile flow estimation for an implantable rotary blood pump
    Karantonis, Dean M.
    Cloherty, Shaun L.
    Mason, David G.
    Ayre, Peter J.
    Lovell, Nigel H.
    2007 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-16, 2007, : 1018 - +
  • [48] Magnetic Drive System for a New Centrifugal Rotary Blood Pump
    Hilton, Andrew
    Tansley, Geoff
    ARTIFICIAL ORGANS, 2008, 32 (10) : 772 - 777
  • [49] Destination therapy with a rotary blood pump and novel power delivery
    Westaby, Stephen
    Siegenthaler, Michael
    Beyersdorf, Friedhelm
    Massetti, Massimo
    Pepper, John
    Khayat, Andre
    Hetzer, Roland
    Frazier, Oscar Howard
    EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2010, 37 (02) : 350 - 356
  • [50] Investigation on the effects of blade thickness for a centrifugal rotary blood pump
    Li, Haoyuan
    Zhang, Mingkui
    Luo, Xianwu
    Wu, Qingyu
    JOURNAL OF MECHANICAL SCIENCE AND TECHNOLOGY, 2023, 37 (02) : 767 - 778