A Passive Magnetically and Hydrodynamically Suspended Rotary Blood Pump

被引:4
|
作者
Stoiber, Martin [1 ,2 ]
Grasl, Christian [1 ,2 ]
Pirker, Stefan [3 ]
Raderer, Franz [1 ]
Schistek, Roland [4 ]
Huber, Leopold [1 ,2 ]
Gittler, Philipp [3 ]
Schima, Heinrich [1 ,2 ]
机构
[1] Med Univ Vienna, Ctr Biomed Engn & Phys, A-1090 Vienna, Austria
[2] Ludwig Boltzmann Cluster Cardiovasc Res, Vienna, Austria
[3] Johannes Kepler Univ Linz, Inst Fluid Mech & Heat Transfer, A-4040 Linz, Austria
[4] St Johns Hosp, Dept Cardiac Surg, Salzburg, Austria
关键词
Rotary blood pump; Magnetic levitation; Hydrodynamic levitation; Computational fluid dynamics; Blood trauma;
D O I
10.1111/j.1525-1594.2009.00715.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A combined hydrodynamic-magnetic bearing allows the design of rotary blood pumps that are not encumbered with mechanical bearings and magnets requiring sensors or electrical power. However, such pumps have so far needed very small and accurately manufactured gaps between rotor and housing to assure effective hydromagnetic bearing behavior. In order to use this concept in disposable pump heads, a design that allows larger rotor-housing gaps, and thus larger manufacturing tolerances, is needed. A pump with passive magnetic bearings and a gap between rotor and housing in the range of 0.5 mm was designed. Numerical simulations were performed to optimize the rotor geometry at low levels of shear stress. An experimental test stand was used to find a range of speeds and gap settings that resulted in low levels of vibration and useful pressure-flow relationships. Three different rotor geometries were tested using a viscosity-adjusted test fluid. Blood damage tests were conducted within the desirable range of speeds and gap settings. In this study stable pump performance was demonstrated at total gap widths between 0.3 and 0.7 mm at flows of 0-10 L/min, with afterloads up to 230 mm Hg. Best performance was achieved with rotors sliding on a fluid pillow between the rotor and the outer housing at a gap distance of 50 to 250 mu m. The inner gap distance, between the rotor and the inner housing, could be as great as 500 mu m. Hemolysis tests on the prototype within the chosen operating range showed lower values (NIH = 0.0029 +/- 0.0012 g/100 L) than the Biomedicus BP-80 pump (NIH = 0.0033 +/- 0.0011 g/100 L). In conclusion, it is possible to build rotary blood pumps with passive hydromagnetic bearings that have large gaps between their rotors and housings. Rotor behavior is sensitive to the position of the permanent magnetic drive unit. To minimize vibration and blood damage, the fluid gaps and the rotational speed have to be adjusted according to the desired operating point of the pump. Further study is needed to optimize the magnetic drive unit and to ascertain its ability to withstand inertial loads imposed by sudden movements and external shock.
引用
收藏
页码:250 / 257
页数:8
相关论文
共 50 条
  • [1] DESIGN OF TOTALLY HYDRODYNAMICALLY SUSPENDED IMPELLERS FOR MINIATURIZED ROTARY BLOOD PUMP
    Amaral, F. K.
    Timms, D.
    Bohning, F.
    Graefe, R.
    Schmitz-Rode, T.
    Steinseifer, U.
    [J]. INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2010, 33 (07): : 425 - 425
  • [2] An ultradurable and compact rotary blood pump with a magnetically suspended impeller in the radial direction
    Masuzawa, T
    Kita, T
    Okada, Y
    [J]. ARTIFICIAL ORGANS, 2001, 25 (05) : 395 - 399
  • [3] A hydrodynamically suspended, magnetically sealed mechanically noncontact axial flow blood pump: Design of a hydrodynamic bearing
    Mitamura, Yoshinori
    Kido, Kazuyuki
    Yano, Tetsuya
    Sakota, Daisuke
    Yambe, Tomoyuki
    Sekine, Kazumitsu
    OKamoto, Eiji
    [J]. ARTIFICIAL ORGANS, 2007, 31 (03) : 221 - 224
  • [4] VentrAssist hydrodynamically suspended, open, centrifugal blood pump
    Watterson, PA
    Woodard, JC
    Ramsden, VS
    Reizes, JA
    [J]. ARTIFICIAL ORGANS, 2000, 24 (06) : 475 - 477
  • [5] Magnetically suspended rotary blood pump with radial type combined motor-bearing
    Masuzawa, T
    Kita, T
    Matsuda, K
    Okada, Y
    [J]. ARTIFICIAL ORGANS, 2000, 24 (06) : 468 - 474
  • [6] CENTRIFUGAL BLOOD PUMP WITH A MAGNETICALLY SUSPENDED IMPELLER
    AKAMATSU, T
    NAKAZEKI, T
    ITOH, H
    [J]. ARTIFICIAL ORGANS, 1992, 16 (03) : 305 - 308
  • [7] A magnetically suspended and hydrostatically stabilized centrifugal blood pump
    Hart, RM
    Filipenco, VG
    Kung, RTV
    [J]. ARTIFICIAL ORGANS, 1996, 20 (06) : 591 - 596
  • [8] RECENT STUDIES OF THE CENTRIFUGAL BLOOD PUMP WITH A MAGNETICALLY SUSPENDED IMPELLER
    AKAMATSU, T
    TSUKIYA, T
    NISHIMURA, K
    PARK, CH
    NAKAZEKI, T
    [J]. ARTIFICIAL ORGANS, 1995, 19 (07) : 631 - 634
  • [9] Magnetically suspended centrifugal blood pump with a self bearing motor
    Masuzawa, T
    Onuma, H
    Kim, SJ
    Okada, Y
    [J]. ASAIO JOURNAL, 2002, 48 (04) : 437 - 442
  • [10] Magnetically suspended centrifugal blood pump with an axially levitated motor
    Masuzawa, T
    Ezoe, S
    Kato, T
    Okada, Y
    [J]. ARTIFICIAL ORGANS, 2003, 27 (07) : 631 - 638