A Passive Magnetically and Hydrodynamically Suspended Rotary Blood Pump

被引:4
|
作者
Stoiber, Martin [1 ,2 ]
Grasl, Christian [1 ,2 ]
Pirker, Stefan [3 ]
Raderer, Franz [1 ]
Schistek, Roland [4 ]
Huber, Leopold [1 ,2 ]
Gittler, Philipp [3 ]
Schima, Heinrich [1 ,2 ]
机构
[1] Med Univ Vienna, Ctr Biomed Engn & Phys, A-1090 Vienna, Austria
[2] Ludwig Boltzmann Cluster Cardiovasc Res, Vienna, Austria
[3] Johannes Kepler Univ Linz, Inst Fluid Mech & Heat Transfer, A-4040 Linz, Austria
[4] St Johns Hosp, Dept Cardiac Surg, Salzburg, Austria
关键词
Rotary blood pump; Magnetic levitation; Hydrodynamic levitation; Computational fluid dynamics; Blood trauma;
D O I
10.1111/j.1525-1594.2009.00715.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A combined hydrodynamic-magnetic bearing allows the design of rotary blood pumps that are not encumbered with mechanical bearings and magnets requiring sensors or electrical power. However, such pumps have so far needed very small and accurately manufactured gaps between rotor and housing to assure effective hydromagnetic bearing behavior. In order to use this concept in disposable pump heads, a design that allows larger rotor-housing gaps, and thus larger manufacturing tolerances, is needed. A pump with passive magnetic bearings and a gap between rotor and housing in the range of 0.5 mm was designed. Numerical simulations were performed to optimize the rotor geometry at low levels of shear stress. An experimental test stand was used to find a range of speeds and gap settings that resulted in low levels of vibration and useful pressure-flow relationships. Three different rotor geometries were tested using a viscosity-adjusted test fluid. Blood damage tests were conducted within the desirable range of speeds and gap settings. In this study stable pump performance was demonstrated at total gap widths between 0.3 and 0.7 mm at flows of 0-10 L/min, with afterloads up to 230 mm Hg. Best performance was achieved with rotors sliding on a fluid pillow between the rotor and the outer housing at a gap distance of 50 to 250 mu m. The inner gap distance, between the rotor and the inner housing, could be as great as 500 mu m. Hemolysis tests on the prototype within the chosen operating range showed lower values (NIH = 0.0029 +/- 0.0012 g/100 L) than the Biomedicus BP-80 pump (NIH = 0.0033 +/- 0.0011 g/100 L). In conclusion, it is possible to build rotary blood pumps with passive hydromagnetic bearings that have large gaps between their rotors and housings. Rotor behavior is sensitive to the position of the permanent magnetic drive unit. To minimize vibration and blood damage, the fluid gaps and the rotational speed have to be adjusted according to the desired operating point of the pump. Further study is needed to optimize the magnetic drive unit and to ascertain its ability to withstand inertial loads imposed by sudden movements and external shock.
引用
收藏
页码:250 / 257
页数:8
相关论文
共 50 条
  • [21] Modeling and control of a magnetic bearing system for the magnetically suspended centrifugal blood pump
    Kim, H
    Kim, HC
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2000, 23 (10): : 689 - 696
  • [22] Adaptive Fuzzy Control of magnetically suspended Rotary Table
    Schallschmidt, Thomas
    Draganov, Denis
    Palis, Frank
    2008 13TH INTERNATIONAL POWER ELECTRONICS AND MOTION CONTROL CONFERENCE, VOLS 1-5, 2008, : 1610 - 1613
  • [23] Investigation of the Axial Gap Clearance in a Hydrodynamic-Passive Magnetically Levitated Rotary Blood Pump Using X-Ray Radiography
    Thamsen, Bente
    Plamondon, Mathieu
    Granegger, Marcus
    Daners, Marianne Schmid
    Kaufmann, Rolf
    Neels, Antonia
    Meboldt, Mirko
    ARTIFICIAL ORGANS, 2018, 42 (05) : 510 - 515
  • [24] Design and experimental investigation of a magnetically suspended rotary table
    Schallschmidt, Thomas.
    Draganov, Denis.
    Palls, Frank.
    2007 EUROPEAN CONFERENCE ON POWER ELECTRONICS AND APPLICATIONS, VOLS 1-10, 2007, : 2583 - 2590
  • [25] Development of a centrifugal blood pump with magnetically suspended impeller and the related fluid mechanical problems
    T Akamatsu
    T Tsukiya
    Sadhana, 1998, 23 : 597 - 603
  • [26] Development of a centrifugal blood pump with magnetically suspended impeller and the related fluid mechanical problems
    Akamatsu, T
    Tsukiya, T
    SADHANA-ACADEMY PROCEEDINGS IN ENGINEERING SCIENCES, 1998, 23 (5-6): : 597 - 603
  • [27] Use of motor current in flow rate measurement for the magnetically suspended centrifugal blood pump
    Tsukiya, T
    Akamatsu, T
    Nishimura, K
    Yamada, T
    Nakazeki, T
    ARTIFICIAL ORGANS, 1997, 21 (05) : 396 - 401
  • [28] Absolute dry running magnetically coupled rotary pump
    Renner, Martin
    CHEMIE INGENIEUR TECHNIK, 2008, 80 (04) : 482 - 483
  • [29] Field Dynamic Balancing for Magnetically Suspended Turbomolecular Pump
    Chen, Qi
    Li, Jinlei
    SENSORS, 2023, 23 (13)
  • [30] Chronic animal experiment with magnetically suspended centrifugal pump
    Yamada, T
    Nishimura, K
    Park, CH
    Kono, S
    Yuasa, S
    Tsukiya, T
    Akamatsu, T
    Matsuda, K
    Ban, T
    ARTIFICIAL ORGANS, 1997, 21 (07) : 635 - 638