Use of motor current in flow rate measurement for the magnetically suspended centrifugal blood pump

被引:0
|
作者
Tsukiya, T
Akamatsu, T
Nishimura, K
Yamada, T
Nakazeki, T
机构
[1] KYOTO UNIV,DEPT CARDIOVASC SURG,KYOTO,JAPAN
[2] NTN CORP,IWATA,JAPAN
关键词
ventricular assist device; rotary blood pump; magnetic suspension; indirect flow rate measurement; viscosity;
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Indirect measurement of the flow rate of a centrifugal blood pump using the driving motor current was studied. The pump flow rate can be expressed as a function of the motor current under a given motor speed in the absence of energy loss resulting from uncertain mechanical contact friction. The magnetically suspended centrifugal blood pump (MSCP), developed by the collaboration of Kyoto University and NTN Inc., was suitable for the application of this measuring method because the impeller is suspended magnetically inside the pump housing without any mechanical contact. The effect of fluid viscosity on the pump performance was investigated in detail, and it was possible to estimate the pump flow rate and the pressure difference through the pump (from inlet port to outlet port) accurately by monitoring the motor current and speed when the kinematic viscosity of working fluids was known. The kinematic viscosity of working fluids can also be measured with the MSCP. The motor current and motor speed were monitored in a chronic animal experiment, and the estimated flow rate and pressure difference showed good correlation with directly measured data.
引用
下载
收藏
页码:396 / 401
页数:6
相关论文
共 50 条
  • [1] Magnetically suspended centrifugal blood pump with a self bearing motor
    Masuzawa, T
    Onuma, H
    Kim, SJ
    Okada, Y
    ASAIO JOURNAL, 2002, 48 (04) : 437 - 442
  • [2] Magnetically suspended centrifugal blood pump with an axially levitated motor
    Masuzawa, T
    Ezoe, S
    Kato, T
    Okada, Y
    ARTIFICIAL ORGANS, 2003, 27 (07) : 631 - 638
  • [3] CENTRIFUGAL BLOOD PUMP WITH A MAGNETICALLY SUSPENDED IMPELLER
    AKAMATSU, T
    NAKAZEKI, T
    ITOH, H
    ARTIFICIAL ORGANS, 1992, 16 (03) : 305 - 308
  • [4] Visualization of main and leakage flow in magnetically suspended centrifugal blood pump
    Masaaki Horie
    Keiichi Yamamura
    Journal of Visualization, 2012, 15 : 353 - 361
  • [5] Visualization of main and leakage flow in magnetically suspended centrifugal blood pump
    Horie, Masaaki
    Yamamura, Keiichi
    JOURNAL OF VISUALIZATION, 2012, 15 (04) : 353 - 361
  • [6] Measurement of blood hematocrit inside the magnetically suspended centrifugal pump using an optical technique: Application to assessment of pump flow
    Kaiwa, T
    Mori, T
    Kijima, T
    Nogawa, M
    Nojiri, C
    Takatani, S
    ARTIFICIAL ORGANS, 1999, 23 (06) : 490 - 495
  • [7] A magnetically suspended and hydrostatically stabilized centrifugal blood pump
    Hart, RM
    Filipenco, VG
    Kung, RTV
    ARTIFICIAL ORGANS, 1996, 20 (06) : 591 - 596
  • [8] RECENT STUDIES OF THE CENTRIFUGAL BLOOD PUMP WITH A MAGNETICALLY SUSPENDED IMPELLER
    AKAMATSU, T
    TSUKIYA, T
    NISHIMURA, K
    PARK, CH
    NAKAZEKI, T
    ARTIFICIAL ORGANS, 1995, 19 (07) : 631 - 634
  • [9] Performance and unsteady characteristics of magnetically suspended centrifugal blood pump
    Yoshino, Y
    Akamatsu, T
    JSME INTERNATIONAL JOURNAL SERIES B-FLUIDS AND THERMAL ENGINEERING, 1997, 40 (01) : 114 - 120
  • [10] Magnetically suspended centrifugal blood pump with a radial magnetic driver
    Hoshi, H
    Katakoa, K
    Ohuchi, K
    Asama, J
    Shinshi, T
    Shimokohbe, A
    Takatani, S
    ASAIO JOURNAL, 2005, 51 (01) : 60 - 64