Characterization of a pediatric rotary blood pump

被引:2
|
作者
de Melo T.R. [1 ]
Vasconcelos F.J.S. [1 ]
Ribeiro L.H.R.D. [1 ]
Bacht S. [2 ]
Cestari I.A. [2 ]
Neto J.S.D.R. [1 ]
Lima A.M.N. [1 ]
机构
[1] Federal University of Campina Grande, Campina Grande, PB
[2] Bioengineering Division, Heart Institute, São Paulo, SP
基金
巴西圣保罗研究基金会;
关键词
Centrifugal flow pump; Lumped parameter model; Rotary blood pump; System identification; Ventricular assist device;
D O I
10.1590/2446-4740.180046
中图分类号
学科分类号
摘要
Introduction: A ventricular assist device (VAD) is an electromechanical pump used to treat heart failures. For designing the physiological control system for a VAD, one needs a mathematical model and its related parameters. This paper presents a characterization procedure for determining the model parameter values of the electrical, mechanical, and hydraulic subsystems of a pediatric Rotary Blood Pump (pRBP). Methods: An in vitro test setup consisting of a pRBP prototype, a motor driver module, an acrylic reservoir, mechanical resistance and tubings, pressure and fluid flow sensors, and data acquisition, processing, and visualization system. The proposed procedure requires a set of experimental tests, and a parameter estimation algorithm for determining the model parameters values. Results: The operating limits of the pRBP were identified from the steady-state data. The relationship between the pressure head, flow rate, and the rotational speed of the pRBP was found from the static tests. For the electrical and mechanical subsystems, the dc motor model has a viscous friction coefficient that varies nonlinearly with the flow. For the hydraulic subsystem, the pressure head is assumed to be a sum of terms related to the resistance, the inertance, the friction coefficient, and the pump speed. Conclusion: The proposed methodology was successfully applied to the characterization of the pRBP. The combined use of static and dynamic tests provided a precise lumped parameter model for representing the pRBP dynamics. The agreement, regarding mean squared deviation, between experimental and simulated results demonstrates the correctness and feasibility of the characterization procedure. © 2018, Brazilian Society of Biomedical Engineering. All rights reserved.
引用
收藏
页码:299 / 309
页数:10
相关论文
共 50 条
  • [21] Development of Medical and Technical Requirements and Simulation of the Flow−Pressure Characteristics of the Sputnik Pediatric Rotary Blood Pump
    Denisov M.V.
    Selishchev S.V.
    Telyshev D.V.
    Frolova E.A.
    Biomedical Engineering, 2017, 50 (5) : 296 - 299
  • [22] Cardiac transplant or rotary blood pump: Contemporary evidence
    Westaby, Stephen
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2013, 145 (01): : 24 - 31
  • [23] High efficiency magnetic bearing for a rotary blood pump
    Chen, HM
    Smith, WA
    Walton, JF
    ASAIO JOURNAL, 1998, 44 (05) : M728 - M732
  • [24] Ischemic cardiomyopathy: cardiac transplant or rotary blood pump?
    Westaby, S.
    CARDIOLOGY, 2013, 126 : 53 - 53
  • [25] COMPUTATIONAL FLOW OPTIMIZATION OF ROTARY BLOOD PUMP COMPONENTS
    ANTAKI, JF
    GHATTAS, O
    BURGREEN, GW
    HE, BC
    ARTIFICIAL ORGANS, 1995, 19 (07) : 608 - 615
  • [26] Gain Scheduling Flow Control of a Rotary Blood Pump
    Bergmann, Lukas
    Rueschen, Daniel
    Amacher, Raffael
    Nelles, Oliver
    Leonhardt, Steffen
    Walter, Marian
    FORSCHUNG IM INGENIEURWESEN-ENGINEERING RESEARCH, 2018, 82 (01): : 21 - 31
  • [27] A rotary blood pump: Its design and development strategy
    Nose, Y
    ARTIFICIAL ORGANS, 1997, 21 (04) : 263 - 264
  • [28] The estimation of cardiac function from the rotary blood pump
    Nakata, K
    Shiono, M
    Akiyama, K
    Orime, Y
    Saito, A
    Tutomu, H
    Negishi, N
    Sezai, Y
    Sankai, T
    ARTIFICIAL ORGANS, 2001, 25 (09) : 709 - 712
  • [29] PULSATILE CONTROL OF ROTARY BLOOD PUMP AND CARDIAC WORKLOAD
    Arakawa, Mamoru
    Tatsumi, Eisuke
    Nishimura, Takashi
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2013, 145 (04): : 1144 - 1145
  • [30] Paradoxical effects of viscosity on the VentrAssist rotary blood pump
    Vidakovic, S
    Ayre, P
    Woodard, J
    Lingard, N
    Tansley, G
    Reizes, J
    ARTIFICIAL ORGANS, 2000, 24 (06) : 478 - 482