Evaluation and Comparison of Dosimetric Characteristics of Semiflex®3D and Microdiamond in Relative Dosimetry under 6 and 15 MV Photon Beams in Small Fields

被引:1
|
作者
Harzanji Z.M. [1 ]
Larizadeh M.H. [2 ]
Namiranian N. [3 ]
Nickfarjam A. [1 ]
机构
[1] Department of Medical Physics, School of Medicine, Shahid Sad-oughi University of Medical Sciences, Yazd
[2] Department of Radiation Oncology, School of Medicine, Kerman University of Medical Sciences, Kerman
[3] Yazd Diabetes Research Center, Shahid Sadoughi University of Medical Sciences, Yazd
来源
关键词
Radiation Dosimeters; Radiotherapy; Conformal; High-Energy; Small Field Dosimetry;
D O I
10.31661/jbpe.v0i0.2008-1160
中图分类号
学科分类号
摘要
Background: In modern radiotherapy techniques, the frequently small and non-uniformed fields can increase treatment efficiency due to their highly conformal dose distribution. Particular features including lack of Lateral Charge Particle Equilibrium (LCPE) lead to detectors with high resolution since any error in obtained dosimetric data could cause patient mistreatments. Objective: This study aims to evaluate and compare two small detectors (Semiflex®3D and microdiamond) dosimetric characteristics in small field relative do-simetry. Material and Methods: In this experimental study, the dosimetric properties of Semiflex®3D and microdiamond were assessed under 6 and 15 MV photon beams. The linearity and stability of the detector’s response and dose rate were measured. Square-field sizes ranging from 0.6×0.6-5×5 cm2 were used for obtaining percentage depth dose curves (PDDs) and in-plane profiles. The angular and temperature dependence of both detectors’ responses were also studied. Results: The detector response shows good stability, no deviation from linearity, and low dose rate dependence (≤1.6%). PDDs and in-plan profiles of both detectors are in good agreement and no significant difference was observed except for the high dose gradient regions (P-value≤0.017). Both detectors demonstrated low angular dependence (<0.3%) with temperature dependence lower than 1% for both detectors. Conclusion: The results indicate both investigated detectors were well performed in small field relative dosimetry and for measuring penumbra, it is better to use micro-diamond detector. © Journal of Biomedical Physics and Engineering.
引用
收藏
页码:477 / 488
页数:11
相关论文
共 18 条
  • [1] Polarity and ion recombination corrections for SemiFlex 3D 31021 IC in 6MV and 6MV FFF photon beams
    Castellet Garcia, M. C.
    Monasor Denia, P.
    Manjon Garcia, C.
    Lopez Tarjuelo, J.
    RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S921 - S921
  • [2] Dosimetric effect of jaws for small MLC fields in 6 and 15 MV photon beam
    Chow, J
    Seguin, M
    Alexander, A
    MEDICAL PHYSICS, 2005, 32 (06) : 1981 - 1982
  • [3] Evaluation of a liquid ionization chamber for relative dosimetry in small and large fields of radiotherapy photon beams
    Benitez, E. M.
    Casado, F. J.
    Garcia-Pareja, S.
    Martin-Viera, J. A.
    Moreno, C.
    Parra, V.
    RADIATION MEASUREMENTS, 2013, 58 : 79 - 86
  • [4] Comparison of dosimetric parameters of small-field electron beams between Advanced Markus, Semiflex 3D, and Diode E responses
    Bayatiani, Mohammad Reza
    Aliasgharzadeh, Akbar
    Seif, Fatemeh
    Mohaghegh, Fatholah
    Fallahi, Fatemeh
    RADIOLOGICAL PHYSICS AND TECHNOLOGY, 2020, 13 (03) : 296 - 305
  • [5] Comparison of dosimetric parameters of small-field electron beams between Advanced Markus, Semiflex 3D, and Diode E responses
    Mohammad Reza Bayatiani
    Akbar Aliasgharzadeh
    Fatemeh Seif
    Fatholah Mohaghegh
    Fatemeh Fallahi
    Radiological Physics and Technology, 2020, 13 : 296 - 305
  • [6] Feasibility of using PRESAGE® for relative 3D dosimetry of small proton fields
    Zhao, Li
    Newton, Joseph
    Oldham, Mark
    Das, Indra J.
    Cheng, Chee-Wai
    Adamovics, John
    PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (22): : N431 - N443
  • [7] Evaluation of the Relative Output Ratio for Collimator Jaw and MLC Defined Small Static 6MV Photon Fields
    Cho, G.
    Thwaites, D.
    MEDICAL PHYSICS, 2014, 41 (06) : 262 - 263
  • [8] Can Gafchromic EBT3 films effectively characterize small fields of 6 MV unflattened photon beams of Cyberknife system?
    Amalraj, Jerrin
    Velayudham, Ramasubramanian
    Anchineyan, Pichandi
    POLISH JOURNAL OF MEDICAL PHYSICS AND ENGINEERING, 2018, 24 (04): : 181 - 187
  • [9] Evaluation of dosimetric parameters of small fields of 6 MV flattening filter free photon beam measured using various detectors against Monte Carlo simulation
    Mamballikalam, Gopinath
    Senthilkumar, S.
    Jayadevan, P. M.
    Jaon Bos, R. C.
    Ahamed Basith, P. M.
    Inippully, Rohit
    Shine, N. S.
    Clinto, C. O.
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2021, 20 (03) : 332 - 339
  • [10] Dosimetric and geometric characteristics of a small animal image-guided irradiator using 3D dosimetry/optical CT scanner
    Na, Yong Hum
    Wang, Yi-Fang
    Black, Paul J.
    Velten, Christian
    Qian, Xin
    Lin, Shih-Chi
    Adamovics, John
    Wuu, Cheng-Shie
    MEDICAL PHYSICS, 2018, 45 (07) : 3330 - 3339