Evaluation and Comparison of Dosimetric Characteristics of Semiflex®3D and Microdiamond in Relative Dosimetry under 6 and 15 MV Photon Beams in Small Fields

被引:1
|
作者
Harzanji Z.M. [1 ]
Larizadeh M.H. [2 ]
Namiranian N. [3 ]
Nickfarjam A. [1 ]
机构
[1] Department of Medical Physics, School of Medicine, Shahid Sad-oughi University of Medical Sciences, Yazd
[2] Department of Radiation Oncology, School of Medicine, Kerman University of Medical Sciences, Kerman
[3] Yazd Diabetes Research Center, Shahid Sadoughi University of Medical Sciences, Yazd
来源
关键词
Radiation Dosimeters; Radiotherapy; Conformal; High-Energy; Small Field Dosimetry;
D O I
10.31661/jbpe.v0i0.2008-1160
中图分类号
学科分类号
摘要
Background: In modern radiotherapy techniques, the frequently small and non-uniformed fields can increase treatment efficiency due to their highly conformal dose distribution. Particular features including lack of Lateral Charge Particle Equilibrium (LCPE) lead to detectors with high resolution since any error in obtained dosimetric data could cause patient mistreatments. Objective: This study aims to evaluate and compare two small detectors (Semiflex®3D and microdiamond) dosimetric characteristics in small field relative do-simetry. Material and Methods: In this experimental study, the dosimetric properties of Semiflex®3D and microdiamond were assessed under 6 and 15 MV photon beams. The linearity and stability of the detector’s response and dose rate were measured. Square-field sizes ranging from 0.6×0.6-5×5 cm2 were used for obtaining percentage depth dose curves (PDDs) and in-plane profiles. The angular and temperature dependence of both detectors’ responses were also studied. Results: The detector response shows good stability, no deviation from linearity, and low dose rate dependence (≤1.6%). PDDs and in-plan profiles of both detectors are in good agreement and no significant difference was observed except for the high dose gradient regions (P-value≤0.017). Both detectors demonstrated low angular dependence (<0.3%) with temperature dependence lower than 1% for both detectors. Conclusion: The results indicate both investigated detectors were well performed in small field relative dosimetry and for measuring penumbra, it is better to use micro-diamond detector. © Journal of Biomedical Physics and Engineering.
引用
收藏
页码:477 / 488
页数:11
相关论文
共 18 条
  • [11] Feasibility study of using Stereotactic Field Diode for field output factors measurement and evaluating three new detectors for small field relative dosimetry of 6 and 10 MV photon beams
    Gul, Attia
    Fukuda, Shigekazu
    Mizuno, Hideyuki
    Taku, Nakaji
    Kakakhel, M. Basim
    Mirza, Sikander M.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2020, 21 (11): : 23 - 36
  • [12] Volumetric modulated arc planning for lung stereotactic body radiotherapy using conventional and unflattened photon beams: a dosimetric comparison with 3D technique
    Zhang, Geoffrey G.
    Ku, Lichung
    Dilling, Thomas J.
    Stevens, Craig W.
    Zhang, Ray R.
    Li, Weiqi
    Feygelman, Vladimir
    RADIATION ONCOLOGY, 2011, 6
  • [13] Volumetric modulated arc planning for lung stereotactic body radiotherapy using conventional and unflattened photon beams: a dosimetric comparison with 3D technique
    Geoffrey G Zhang
    Lichung Ku
    Thomas J Dilling
    Craig W Stevens
    Ray R Zhang
    Weiqi Li
    Vladimir Feygelman
    Radiation Oncology, 6
  • [14] Comparison of Flattening Filter (FF) and Flattening-Filter-Free (FFF) 6 MV photon beam characteristics for small field dosimetry using EGSnrc Monte Carlo code
    Sangeetha, S.
    Sureka, C. S.
    RADIATION PHYSICS AND CHEMISTRY, 2017, 135 : 63 - 75
  • [15] Validation of GEANT4 simulations for percentage depth dose calculations in heterogeneous media by using small photon beams from the 6-MV Cyberknife: Comparison with photon beam dosimetry with EBT2 film
    Lee, Chung Il
    Yoon, Sei-Chul
    Shin, Jae Won
    Hong, Seung-Woo
    Suh, Tae Suk
    Min, Kyung Joo
    Lee, Sang Deok
    Chung, Su Mi
    Jung, Jae-Yong
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2015, 66 (08) : 1308 - 1316
  • [16] Validation of GEANT4 simulations for percentage depth dose calculations in heterogeneous media by using small photon beams from the 6-MV Cyberknife: Comparison with photon beam dosimetry with EBT2 film
    Chung Il Lee
    Sei-Chul Yoon
    Jae Won Shin
    Seung-Woo Hong
    Tae Suk Suh
    Kyung Joo Min
    Sang Deok Lee
    Su Mi Chung
    Jae-Yong Jung
    Journal of the Korean Physical Society, 2015, 66 : 1308 - 1316
  • [17] Evaluation of dosimetric parameters of small fields of 6 MV flattening filter free photon beam measured using various detectors against Monte Carlo simulation (Mar, 10.1017/S1460396920000114, 2020)
    Mamballikalam, Gopinath
    Senthilkumar, S.
    Jayadevan, P. M.
    Jaon Bos, R. C.
    Basith, P. M. Ahamed
    Inippully, Rohit
    Shine, N. S.
    Clinto, C. O.
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2021, 20 (03) : 369 - 369
  • [18] A dosimetric evaluation of the IAEA-AAPM TRS483 code of practice for dosimetry of small static fields used in conventional linac beams and comparison with IAEA TRS-398, AAPM TG51, and TG51 Addendum protocols
    Huq, M. Saiful
    Hwang, Min-Sig
    Teo, Troy P.
    Jang, Si Young
    Heron, Dwight E.
    Lalonde, Ronald J.
    MEDICAL PHYSICS, 2018, 45 (09) : 4257 - 4273