Can Gafchromic EBT3 films effectively characterize small fields of 6 MV unflattened photon beams of Cyberknife system?

被引:1
|
作者
Amalraj, Jerrin [1 ,2 ]
Velayudham, Ramasubramanian [2 ]
Anchineyan, Pichandi [1 ]
机构
[1] HCG Enterprises Ltd, Dept Radiat Oncol, Cent Phys, 8 P Kalingarao Rd, Bangalore 560027, Karnataka, India
[2] Vellore Inst Technol, Sch Adv Sci, Vellore 632014, Tamil Nadu, India
来源
关键词
diode detectors; GafChromic EBT3; small fields; output factor; off centered ratio; percentage depth dose;
D O I
10.2478/pjmpe-2018-0025
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Shielded silicon diodes are commonly employed in commissioning of Cyberknife 6 MV photon beams. This study aims to measure output factors, off centered ratio (OCR), percentage depth dose (PDD) of 6 MV photons using shielded and unshielded diodes and to compare with Gafchromic EBT3 film measurements to investigate whether EBT3 could effectively characterize small 6 MV photon beams. Output factors, OCR and PDD were measured with shielded and unshielded silicon detectors in a radiation field analyzer system at reference condition. Water equivalent solid phantom were used while irradiating EBT3 films. From multiuser data, diodes underestimated output factor by 3% for collimator fields <= 10 mm, while EBT3 underestimated the output factor by 3.9% for 5 mm collimator. 1D Gamma analysis of OCR between diode and film, results in gamma <= 1 for all measured points with 1 mm distance to agreement (DTA) and 1% relative dose difference (DD). Dose at surface is overestimated with diodes compared to EBT3. PDD results were within 2% relative dose values between diode and EBT3 except for 5 mm collimator. Except for small collimator fields of up to 10 mm, results of output factor, OCR, PDD of all detectors used in this study exhibited similar results. Relative dose measurements with Gafchromic EBT3 in this work show that EBT3 films can be used effectively as an independent tool to verify commissioning beam data of small fields only after careful verification of methodology for any systematic errors with appropriate readout procedure.
引用
收藏
页码:181 / 187
页数:7
相关论文
共 50 条
  • [1] Comparison of Gafchromic EBT2 and EBT3 films for clinical photon and proton beams
    Reinhardt, S.
    Hillbrand, M.
    Wilkens, J. J.
    Assmann, W.
    [J]. MEDICAL PHYSICS, 2012, 39 (08) : 5257 - 5262
  • [2] Dosimetric Characteristics of Gafchromic EBT3 Films for Megavoltage Photon and Proton Beams
    Hayashi, N.
    Kato, H.
    Yada, R.
    Matsunaga, T.
    Ehara, I.
    Adachi, Y.
    [J]. MEDICAL PHYSICS, 2012, 39 (06) : 3732 - 3732
  • [3] Evaluation of Gafchromic® EBT3 films characteristics in therapy photon, electron and proton beams
    Sorriaux, J.
    Kacperek, A.
    Rossomme, S.
    Lee, J. A.
    Bertrand, D.
    Vynckier, S.
    Sterpin, E.
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2013, 29 (06): : 599 - 606
  • [4] Evaluation of the accuracy of Cyberknife tumor tracking using gafchromic EBT3 films
    Mastella, E.
    Vigorito, S.
    Rondi, E.
    Cattani, F.
    Piperno, G.
    Ferrari, A. M.
    Rozza, D.
    Strata, E.
    Orecchia, R.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S415 - S415
  • [5] Evaluation of Dose Rate and Photon Energy Dependence of Gafchromic EBT3 Film Irradiating with 6 MV and Co-60 Photon Beams
    Ataei, Gholamreza
    Rezaei, Maral
    Gorji, Ebrahimnejad
    Banaei, Amin
    Goushbolagh, Nouraddin Abdi
    Farhood, Bagher
    Bagheri, Mahmoud
    Firouzjah, Razzagh Abedi
    [J]. JOURNAL OF MEDICAL SIGNALS & SENSORS, 2019, 9 (03): : 204 - 210
  • [6] Characterization of EBT3 Films in Photon and Proton Beams
    Sorriaux, J.
    Rossomme, S.
    Kacperek, A.
    Bertrand, D.
    Vynckier, S.
    Sterpin, E.
    [J]. MEDICAL PHYSICS, 2012, 39 (06) : 3730 - 3730
  • [7] In Vivo Peripheral Dose Measurements in IMRT and CyberKnife Treatments Using Gafchromic EBT3 Films
    Ganapathy, K.
    Murali, V
    Muthukumaran, M.
    Velmurugan, J.
    [J]. MEDICAL PHYSICS, 2017, 44 (06) : 2899 - 2899
  • [8] Surface dose measurements with GafChromic EBT film for 6 and 18 MV photon beams
    Bilge, Hatice
    Cakir, Aydin
    Okutan, Murat
    Acar, Hilal
    [J]. PHYSICA MEDICA, 2009, 25 (02) : 101 - 104
  • [9] Commissioning and Validation of EBT3 Gafchromic Films for Measurements of Dose Profiles and Integrals of Dose in Small Fields
    Dufreneix, S.
    Bordy, J.
    Delaunay, F.
    Daures, J.
    Gouriou, J.
    Le Roy, M.
    Ostrowsky, A.
    Rapp, B.
    Sommier, L.
    [J]. MEDICAL PHYSICS, 2014, 41 (06) : 547 - 548
  • [10] Investigation of Gafchromic EBT3 Film Energy Dependence Using Proton, Photon, and Electron Beams
    Ferreira, C.
    Schnell, E.
    Ahmad, S.
    Herman, T. De La Fuente
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3403 - 3403