Evaluation of a liquid ionization chamber for relative dosimetry in small and large fields of radiotherapy photon beams

被引:9
|
作者
Benitez, E. M. [1 ]
Casado, F. J. [1 ]
Garcia-Pareja, S. [1 ]
Martin-Viera, J. A. [1 ]
Moreno, C. [1 ]
Parra, V. [1 ]
机构
[1] Hosp Reg Univ Carlos Haya, Serv Radiofis Hosp, Malaga 29010, Spain
关键词
Liquid ionization chamber; Relative dosimetry; Isooctane; Uncertainty analysis; GENERAL RECOMBINATION CORRECTION; ION-CHAMBER; NONSTANDARD FIELDS; 2-DOSE-RATE METHOD; SILICON DIODE; ACCELERATORS; RADIATION; ENERGY; ARRAY;
D O I
10.1016/j.radmeas.2013.08.009
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
Commissioning and quality assurance of radiotherapy linear accelerators require measurement of the absorbed dose to water, and a wide range of detectors are available for absolute and relative dosimetry in megavoltage beams. In this paper, the PTW microLion isooctane-filled ionization chamber has been tested to perform relative measurements in a 6 MV photon beam from a linear accelerator. Output factors, percent depth dose and dose profiles have been obtained for small and large fields. These quantities have been compared with those from usual detectors in the routine practice. In order to carry out a more realistic comparison, an uncertainty analysis has been developed, taking type A and B uncertainties into account. The results present microLion as a good option when high spatial resolution is needed, thanks to its reduced sensitive volume. The liquid filling also provides a high signal compared to other detectors, like that based on air filling. Furthermore, the relative response of microLion when field size is varied suggests that this detector has energy dependence, since it is appreciated an over-response for small fields and an under-response for the large ones. This effect is more obvious for field sizes wider than 20 x 20 cm(2), where the differences in percent depth dose at great depths exceed the uncertainties estimated in this study. (C) 2013 Elsevier Ltd. All rights reserved.
引用
收藏
页码:79 / 86
页数:8
相关论文
共 50 条
  • [1] Use of a liquid ionization chamber for stereotactic radiotherapy dosimetry
    Wagner, A.
    Crop, F.
    Lacornerie, T.
    Vandevelde, F.
    Reynaert, N.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (08): : 2445 - 2459
  • [2] Ionization chamber dosimetry of small photon fields:: a Monte Carlo study on stopping-power ratios for radiosurgery and IMRT beams
    Sánchez-Doblado, F
    Andreo, P
    Capote, R
    Leal, A
    Perucha, M
    Arráns, R
    Núñez, L
    Mainegra, E
    Lagares, JI
    Carrasco, E
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (14): : 2081 - 2099
  • [3] On the orientation of ionization chambers in dosimetry of small photon fields
    Casar, B.
    Gershkevitsh, E.
    Mendez, I.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S471 - S472
  • [4] Dosimetry of pulsed clinical proton beams by a small ionization chamber
    Nohtomi, A
    Sakae, T
    Tsunashima, Y
    Kohno, R
    [J]. MEDICAL PHYSICS, 2001, 28 (07) : 1431 - 1435
  • [5] Comments on 'Ionization chamber dosimetry of small photon fields: a Monte Carlo study on stopping-power ratios for radiosurgery and IMRT beams'
    Seuntjens, J
    Verhaegen, F
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (21): : L43 - L45
  • [6] DOSIMETRIC CHARACTERISATION OF SMALL RADIOTHERAPY PHOTON FIELDS: MICRO LIQUID-FILLED IONIZATION CHAMBER COMPARED WITH CONVENTIONAL DETECTORS
    Andreoli, S.
    Colleoni, P.
    Fortunato, M.
    Moretti, R.
    Galelli, M.
    Kaiser, S. Ren
    [J]. RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S421 - S422
  • [7] Liquid ionization chamber measurements of dose distributions in small 6 MV photon beams
    Dasu, A
    Lofroth, PO
    Wickman, G
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (01): : 21 - 36
  • [8] Reply to 'Comments on 'Ionization chamber dosimetry of small photon fields:: a Monte Carlo study on stopping-power ratios for radiosurgery and IMRT beams"
    Andreo, P
    Sánchez-Doblado, F
    Capote, R
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (21): : L46 - L48
  • [9] IONIZATION-CHAMBER DOSIMETRY FOR PHOTON AND ELECTRON-BEAMS - THEORETICAL CONSIDERATIONS
    ALMOND, PR
    SVENSSON, H
    [J]. ACTA RADIOLOGICA-THERAPY PHYSICS BIOLOGY, 1977, 16 (02): : 177 - 186
  • [10] The Magic Cube and the Pixef Ionization Chamber: detectors for monitor and dosimetry of radiotherapy beams
    Amerio, S
    Boriano, A
    Bourhaleb, F
    Cirio, R
    Donetti, M
    Garelli, E
    Giordanengo, S
    Madon, E
    Marchetto, F
    Nastasi, U
    Peroni, C
    Freire, CJS
    Sardo, A
    Trevisiol, E
    [J]. NUCLEAR PHYSICS B-PROCEEDINGS SUPPLEMENTS, 2003, 125 : 154 - 158