Corneal densitometry measurements comparison between anterior segment OCT and scheimpflug imaging

被引:0
|
作者
Lupardi, Enrico [1 ,2 ]
Moramarco, Antonio [1 ,2 ]
Cassini, Federico [1 ,2 ]
Febbraro, Simone [1 ,2 ]
Savini, Giacomo [3 ]
Fontana, Luigi [1 ,2 ]
机构
[1] Azienda Ospedaliero Univ Bologna, IRCCS, Bologna, Italy
[2] Alma Mater Studiorum Univ Bologna IT, Ophthalmol Unit, DIMEC, Bologna, Italy
[3] IRCCS Bietti Fdn, Rome, Italy
关键词
Corneal densitometry; Corneal transparency; Corneal opacity; AS-OCT; Scheimpflug Imaging; LIGHT-SCATTERING; ENDOTHELIAL DYSTROPHY; OBJECTIVE MEASUREMENT; HEALTHY CORNEAS; REPEATABILITY; BACKSCATTER; VALUES; CAMERA;
D O I
10.1007/s10792-024-03309-0
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PurposeTo evaluate and compare the repeatability of corneal densitometry (CD) measurements obtained using both an anterior-segment optical coherence tomography (AS-OCT) device and a Scheimpflug camera system, while also assessing the level of agreement. The study also sought to investigate the correlation of CD with age, gender, and central corneal thickness (CCT) in normal eyes.MethodsCD measurements were obtained using the Casia 2 and the Pentacam AXL Wave. Data were collected on Total Corneal Densitometry and 4 concentric corneal annular areas, these are referred to as zone 1, denoting the central area, through to zone 4, designating the outermost peripheral region. Repeatability was assessed using intra-session test-retest variability, coefficient of variation (CoV), and intraclass correlation coefficient (ICC). The agreement was evaluated using Bland-Altman plots. Correlation analysis was performed between CD, age, gender, and CCT.ResultsThe study included 96 healthy volunteers. The Casia 2 demonstrated high repeatability with ICC values exceeding 0.9 in all the corneal zones and lower CoV values compared to the Pentacam AXL Wave (ranging from 1.07% to 2.25% for Casia 2 and from 1.91% to 6.89% for Pentacam).95% LoA were within +/- 2 standard deviation from the average mean except from zone 1 (+/- 2.42).However, the measurements showed a consistent bias among all the corneal zones. CD values were positively correlated with age, except for zone 1 with the Pentacam (p = 0.083).ConclusionsThe findings suggest that the Casia 2 can be a reliable tool for assessing corneal transparency in healthy individuals, however its measurements are not interchangeable with those provided by the Pentacam. The AS-OCT device may be more sensitive in detecting subtle age-related changes in CD within the central zone.
引用
收藏
页数:10
相关论文
共 50 条
  • [31] Repeatability assessment of anterior segment measurements in myopic patients using an anterior segment OCT with placido corneal topography and agreement with a swept-source OCT
    Wang, Hao
    Zhu, Li-Shuang
    Pang, Chen-Jiu
    Fan, Qi
    BMC OPHTHALMOLOGY, 2024, 24 (01)
  • [32] Pattern of Scheimpflug imaging in anterior segment foreign bodies
    R Arora
    S Mehta
    J L Goyal
    S Pahuja
    D Gupta
    R Gupta
    Eye, 2010, 24 : 1304 - 1306
  • [33] Comparison of corneal dynamic parameters and tomographic measurements using Scheimpflug imaging in keratoconus
    Chan, Tommy C. Y.
    Wang, Yu Meng
    Yu, Marco
    Jhanji, Vishal
    BRITISH JOURNAL OF OPHTHALMOLOGY, 2018, 102 (01) : 42 - 47
  • [34] Pattern of Scheimpflug imaging in anterior segment foreign bodies
    Arora, R.
    Mehta, S.
    Goyal, J. L.
    Pahuja, S.
    Gupta, D.
    Gupta, R.
    EYE, 2010, 24 (07) : 1304 - 1306
  • [35] High resolution anterior segment OCT CASIA 2 for corneal and anterior segment diagnosis
    Dobrowolski, D.
    ACTA OPHTHALMOLOGICA, 2018, 96 : 139 - 139
  • [36] Anterior segment characteristics in normal and keratoconus eyes evaluated with a combined Scheimpflug/Placido corneal imaging device
    Safarzadeh, Masoud
    Nasiri, Nader
    JOURNAL OF CURRENT OPHTHALMOLOGY, 2016, 28 (03): : 106 - 111
  • [37] Repeatability of Corneal Densitometry Measurements using a Scheimpflug Camera in Healthy Normal Corneas
    Pakbin, Mojgan
    Khabazkhooba, Mehdi
    Pakravan, Mohammad
    Fotouhi, Akbar
    Jafarzadehpur, Ebrahim
    Aghamirsalim, Mohamadreza
    Hashemi, Hassan
    JOURNAL OF CURRENT OPHTHALMOLOGY, 2022, 34 (01): : 50 - 55
  • [38] Characteristics of Pre-Descemet Membrane Corneal Dystrophy by Three Different Imaging Modalities-In Vivo Confocal Microscopy, Anterior Segment Optical Coherence Tomography, and Scheimpflug Corneal Densitometry Analysis
    Malhotra, Chintan
    Jain, Arun Kumar
    Dwivedi, Sujata
    Chakma, Partha
    Rohilla, Vikash
    Sachdeva, Kulbhushan
    CORNEA, 2015, 34 (07) : 829 - 832
  • [39] Comparison of the corneal thickness as measured by Visante® anterior segment OCT versus ultrasound technique
    Wildner, K.
    Mueller, M.
    Dawczynski, J.
    Strobel, I.
    KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 2007, 224 (11) : 832 - 836
  • [40] Correlation between Anterior Segment OCT and Novel 360° Angle Imaging
    Yang, Xin
    Shi, Yue
    Marion, Kenneth
    Francis, Brian
    Sadda, Srinivas R.
    Chopra, Vikas
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2019, 60 (09)