Corneal densitometry measurements comparison between anterior segment OCT and scheimpflug imaging

被引:0
|
作者
Lupardi, Enrico [1 ,2 ]
Moramarco, Antonio [1 ,2 ]
Cassini, Federico [1 ,2 ]
Febbraro, Simone [1 ,2 ]
Savini, Giacomo [3 ]
Fontana, Luigi [1 ,2 ]
机构
[1] Azienda Ospedaliero Univ Bologna, IRCCS, Bologna, Italy
[2] Alma Mater Studiorum Univ Bologna IT, Ophthalmol Unit, DIMEC, Bologna, Italy
[3] IRCCS Bietti Fdn, Rome, Italy
关键词
Corneal densitometry; Corneal transparency; Corneal opacity; AS-OCT; Scheimpflug Imaging; LIGHT-SCATTERING; ENDOTHELIAL DYSTROPHY; OBJECTIVE MEASUREMENT; HEALTHY CORNEAS; REPEATABILITY; BACKSCATTER; VALUES; CAMERA;
D O I
10.1007/s10792-024-03309-0
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PurposeTo evaluate and compare the repeatability of corneal densitometry (CD) measurements obtained using both an anterior-segment optical coherence tomography (AS-OCT) device and a Scheimpflug camera system, while also assessing the level of agreement. The study also sought to investigate the correlation of CD with age, gender, and central corneal thickness (CCT) in normal eyes.MethodsCD measurements were obtained using the Casia 2 and the Pentacam AXL Wave. Data were collected on Total Corneal Densitometry and 4 concentric corneal annular areas, these are referred to as zone 1, denoting the central area, through to zone 4, designating the outermost peripheral region. Repeatability was assessed using intra-session test-retest variability, coefficient of variation (CoV), and intraclass correlation coefficient (ICC). The agreement was evaluated using Bland-Altman plots. Correlation analysis was performed between CD, age, gender, and CCT.ResultsThe study included 96 healthy volunteers. The Casia 2 demonstrated high repeatability with ICC values exceeding 0.9 in all the corneal zones and lower CoV values compared to the Pentacam AXL Wave (ranging from 1.07% to 2.25% for Casia 2 and from 1.91% to 6.89% for Pentacam).95% LoA were within +/- 2 standard deviation from the average mean except from zone 1 (+/- 2.42).However, the measurements showed a consistent bias among all the corneal zones. CD values were positively correlated with age, except for zone 1 with the Pentacam (p = 0.083).ConclusionsThe findings suggest that the Casia 2 can be a reliable tool for assessing corneal transparency in healthy individuals, however its measurements are not interchangeable with those provided by the Pentacam. The AS-OCT device may be more sensitive in detecting subtle age-related changes in CD within the central zone.
引用
收藏
页数:10
相关论文
共 50 条
  • [21] Comparison of radius of anterior lens surface curvature measurements in vivo using the anterior segment optical coherence tomography and Scheimpflug imaging
    Liu, Zhenzhen
    Ruan, Xiaoting
    Wang, Wei
    Liu, Juzi
    Meng, Yijing
    Gu, Xiaoxun
    Fu, Jun
    Luo, Lixia
    Liu, Yizhi
    ANNALS OF TRANSLATIONAL MEDICINE, 2020, 8 (05)
  • [22] Comparison of anterior segment measurements with rotating Scheimpflug photography and partial coherence reflectometry
    Huang, Jinhai
    Pesudovs, Konrad
    Wen, Daizong
    Chen, Shihao
    Wright, Thomas
    Wang, Xiaoyu
    Li, Yini
    Wang, Qinmei
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2011, 37 (02): : 341 - 348
  • [23] Anterior segment OCT in corneal diseases and surgery
    Nubile, M.
    Calienno, R.
    Salgari, N.
    De Nicola, C.
    Lappa, A.
    Mastropasqua, A.
    ACTA OPHTHALMOLOGICA, 2016, 94
  • [24] Imaging of corneal dystrophies: Correlations between en face anterior segment OCT and in vivo confocal microscopy
    Ghouali, W.
    Hassani, R. Tahiri Joutei
    Liang, H.
    Dupont-Monod, S.
    Auclin, F.
    Baudouin, C.
    Labbe, A.
    JOURNAL FRANCAIS D OPHTALMOLOGIE, 2015, 38 (05): : 388 - 394
  • [25] Corneal Thickness Measurements Using Time-domain Anterior Segment OCT, Ultrasound, and Scheimpflug Tomographer Pachymetry Before and After Corneal Cross-linking for Keratoconus
    Mencucci, Rita
    Paladini, Iacopo
    Virgili, Gianni
    Giacomelli, Giovanni
    Menchini, Ugo
    JOURNAL OF REFRACTIVE SURGERY, 2012, 28 (08) : 562 - +
  • [26] Corneal graft imaging by anterior segment OCT during storage in eye banks
    Pataia, G.
    Gauthier, A. S.
    Campolmi, N.
    Trone, M. C.
    He, Z.
    Piselli, S.
    Forest, F.
    Acquart, S.
    Lepine, T.
    Gain, P.
    Thuret, G.
    ACTA OPHTHALMOLOGICA, 2013, 91
  • [27] Comparison of Central Corneal Thickness Measurements Using Ultrasonic Pachymetry, Anterior Segment OCT and Noncontact Specular Microscopy
    Scotto, Riccardo
    Bagnis, Alessandro
    Papadia, Marina
    Cutolo, Carlo Alberto
    Risso, Domenico
    Traverso, Carlo Enrico
    JOURNAL OF GLAUCOMA, 2017, 26 (10) : 860 - 865
  • [28] The precision and agreement of corneal thickness and keratometry measurements with SS-OCT versus Scheimpflug imaging
    Yune Zhao
    Ding Chen
    Giacomo Savini
    Qing Wang
    Hongfang Zhang
    Yili Jin
    Benhao Song
    Rui Ning
    Jinhai Huang
    Chenyang Mei
    Eye and Vision, 7
  • [29] COMPARISON OF CORNEAL PARAMETERS AND CORRESPONDING BIOMECHANICAL MEASUREMENTS USING DIFFERENT ANTERIOR SEGMENT IMAGING SYSTEMS
    Viswanathan, Deepa
    Kumar, Nikhil
    Males, John
    Graham, Stuart
    CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2012, 40 : 74 - 74
  • [30] The precision and agreement of corneal thickness and keratometry measurements with SS-OCT versus Scheimpflug imaging
    Zhao, Yune
    Chen, Ding
    Savini, Giacomo
    Wang, Qing
    Zhang, Hongfang
    Jin, Yili
    Song, Benhao
    Ning, Rui
    Huang, Jinhai
    Mei, Chenyang
    EYE AND VISION, 2020, 7 (01)