Corneal densitometry measurements comparison between anterior segment OCT and scheimpflug imaging

被引:0
|
作者
Lupardi, Enrico [1 ,2 ]
Moramarco, Antonio [1 ,2 ]
Cassini, Federico [1 ,2 ]
Febbraro, Simone [1 ,2 ]
Savini, Giacomo [3 ]
Fontana, Luigi [1 ,2 ]
机构
[1] Azienda Ospedaliero Univ Bologna, IRCCS, Bologna, Italy
[2] Alma Mater Studiorum Univ Bologna IT, Ophthalmol Unit, DIMEC, Bologna, Italy
[3] IRCCS Bietti Fdn, Rome, Italy
关键词
Corneal densitometry; Corneal transparency; Corneal opacity; AS-OCT; Scheimpflug Imaging; LIGHT-SCATTERING; ENDOTHELIAL DYSTROPHY; OBJECTIVE MEASUREMENT; HEALTHY CORNEAS; REPEATABILITY; BACKSCATTER; VALUES; CAMERA;
D O I
10.1007/s10792-024-03309-0
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PurposeTo evaluate and compare the repeatability of corneal densitometry (CD) measurements obtained using both an anterior-segment optical coherence tomography (AS-OCT) device and a Scheimpflug camera system, while also assessing the level of agreement. The study also sought to investigate the correlation of CD with age, gender, and central corneal thickness (CCT) in normal eyes.MethodsCD measurements were obtained using the Casia 2 and the Pentacam AXL Wave. Data were collected on Total Corneal Densitometry and 4 concentric corneal annular areas, these are referred to as zone 1, denoting the central area, through to zone 4, designating the outermost peripheral region. Repeatability was assessed using intra-session test-retest variability, coefficient of variation (CoV), and intraclass correlation coefficient (ICC). The agreement was evaluated using Bland-Altman plots. Correlation analysis was performed between CD, age, gender, and CCT.ResultsThe study included 96 healthy volunteers. The Casia 2 demonstrated high repeatability with ICC values exceeding 0.9 in all the corneal zones and lower CoV values compared to the Pentacam AXL Wave (ranging from 1.07% to 2.25% for Casia 2 and from 1.91% to 6.89% for Pentacam).95% LoA were within +/- 2 standard deviation from the average mean except from zone 1 (+/- 2.42).However, the measurements showed a consistent bias among all the corneal zones. CD values were positively correlated with age, except for zone 1 with the Pentacam (p = 0.083).ConclusionsThe findings suggest that the Casia 2 can be a reliable tool for assessing corneal transparency in healthy individuals, however its measurements are not interchangeable with those provided by the Pentacam. The AS-OCT device may be more sensitive in detecting subtle age-related changes in CD within the central zone.
引用
收藏
页数:10
相关论文
共 50 条
  • [41] Comparison of time domain anterior segment OCT with swept source OCT for angle closure imaging
    Mani, Baskaran
    Yau, Christine
    Ho, Sue-Wei
    Tun, Tin A.
    Perera, Shamira
    Aung, Tin
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2014, 55 (13)
  • [42] Repeatability, Reproducibility, and Agreement Between Three Different Scheimpflug Systems in Measuring Corneal and Anterior Segment Biometry
    Hernandez-Camarena, Julio C.
    Chirinos-Saldana, Patricia
    Navas, Alejandro
    Ramirez-Miranda, Arturo
    de la Mota, Angie
    Jimenez-Corona, Aida
    Graue-Hernandez, Enrique O.
    JOURNAL OF REFRACTIVE SURGERY, 2014, 30 (09) : 616 - U44
  • [43] Reproducibility and reliability of central corneal thickness determination in more and less profound corneal edema using ultrasound pachymetry, a Scheimpflug camera and anterior segment OCT
    Kuerten David
    Hossein-Zadeh Mandana
    Plange Niklas
    Walter Peter
    Fuest Matthias
    Graefe's Archive for Clinical and Experimental Ophthalmology, 2020, 258 : 351 - 358
  • [44] Imaging the Anterior Segment: High-Frequency Ultrasound and Anterior Segment OCT
    Allemann, Norma
    Coleman, D. Jackson
    Pavlin, Charles J.
    Huang, David
    JOURNAL OF OPHTHALMOLOGY, 2013, 2013
  • [45] Anterior segment imaging for glaucoma: OCT or UBM?
    Ishikawa, Hiroshi
    BRITISH JOURNAL OF OPHTHALMOLOGY, 2007, 91 (11) : 1420 - 1421
  • [46] Anterior segment vasculature imaging of OCT angiography
    Zhou, Xiao
    Zhang, Qinqin
    Wang, Ruikang K.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2019, 60 (09)
  • [47] Reproducibility and reliability of central corneal thickness determination in more and less profound corneal edema using ultrasound pachymetry, a Scheimpflug camera and anterior segment OCT
    David, Kuerten
    Mandana, Hossein-Zadeh
    Niklas, Plange
    Peter, Walter
    Matthias, Fuest
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2020, 258 (02) : 351 - 358
  • [48] A Comparison between Scheimpflug Imaging and Optical Coherence Tomography in Measuring Corneal Thickness
    Huang, Jinhai
    Ding, Xixia
    Savini, Giacomo
    Pan, Chao
    Feng, Yifan
    Cheng, Dan
    Hua, Yanjun
    Hu, Xiuli
    Wang, Qinmei
    OPHTHALMOLOGY, 2013, 120 (10) : 1951 - 1958
  • [49] Assessment and Comparison of Automated Anterior Chamber Metrics between Casia Swept-Source OCT vs Pentacam Scheimpflug Imaging
    Marion, Kenneth
    Dastiridou, Anna
    Francis, Brian
    Chopra, Vikas
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2016, 57 (12)
  • [50] Agreement between Scheimpflug photography and A-scan ultrasonography in anterior segment ocular measurements in children
    Tong, L
    Wong, EM
    Chan, YH
    Balakrishnan, V
    OPTOMETRY AND VISION SCIENCE, 2003, 80 (07) : 529 - 534