A DEPTH-ENCODED PET DETECTOR

被引:38
|
作者
BARTZAKOS, P [1 ]
THOMPSON, CJ [1 ]
机构
[1] MCGILL UNIV,MED PHYS UNIT,MONTREAL H3A 2B4,QUEBEC,CANADA
关键词
D O I
10.1109/23.289382
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Radial blurring in high-resolution positron emission tomography (PET) can be reduced by measuring the PET photon's depth of interaction within a detector crystal. Simulations show that dividing the crystal into two equiprobable regions of interaction almost eliminates this blurring. This has been done using a black band placed around a 3 cm BGO crystal, 1 cm from the front. A PMT collects light from the rear of the crystal for energy and timing measurement. Another PMT collects light through a light guide from the top-front of the crystal to find the depth of interaction. Gamma rays interacting in the front of the crystal produce a strong, signal in the depth-sensing PMT, while interactions behind the band produce weaker (< 10%) signal. The intensity of the light recorded by the energy PMT is almost constant (within +/- 3%) for interactions at any depth in the crystal.
引用
收藏
页码:732 / 738
页数:7
相关论文
共 50 条
  • [31] Design of a depth of interaction detector with a PS-PMT for PET
    Murayama, H
    Ishibashi, H
    Uchida, H
    Omura, T
    Yamashita, T
    1999 IEEE NUCLEAR SCIENCE SYMPOSIUM - CONFERENCE RECORD, VOLS 1-3, 1999, : 1062 - 1066
  • [32] Design of a depth of interaction detector with a PS-PMT for PET
    Murayama, H.
    Ishibashi, H.
    Uchida, H.
    Omura, T.
    Yamashita, T.
    IEEE Transactions on Nuclear Science, 2000, 47 (3 III) : 1045 - 1050
  • [33] Monte Carlo simulation for depth encoding multicrystal detector for PET
    Haneishi, H
    Yamada, A
    Takagi, K
    Murayama, H
    2001 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORDS, VOLS 1-4, 2002, : 1993 - 1996
  • [34] Performance results of a prototype depth-encoding PET detector
    Meng, L.J.
    Ramsden, D.
    IEEE Transactions on Nuclear Science, 2000, 47 (3 III) : 1011 - 1017
  • [35] Evaluation of a prototype PET detector with depth-of-interaction encoding
    Burr, KC
    Ivan, A
    Castleberry, DE
    LeBlanc, JW
    Shah, KS
    Farrell, R
    2003 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-5, 2004, : 877 - 881
  • [36] DMice - a depth-of-interaction detector design for PET scanners
    Lewellen, TK
    Janes, M
    Miyaoka, RS
    2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 2388 - 2392
  • [37] Performance results of a prototype depth-encoding PET detector
    Meng, LJ
    Ramsden, D
    1999 IEEE NUCLEAR SCIENCE SYMPOSIUM - CONFERENCE RECORD, VOLS 1-3, 1999, : 929 - 933
  • [38] Performance results of a prototype depth-encoding PET detector
    Meng, LJ
    Ramsden, D
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2000, 47 (03) : 1011 - 1017
  • [39] A practical block detector for a depth-encoding PET camera
    Rogers, JG
    Moisan, C
    Hoskinson, EM
    Andreaco, MS
    Williams, CW
    Nutt, R
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1996, 43 (06) : 3240 - 3248
  • [40] A Depth of Interaction PET Detector Using Side Surface Readout
    Miyaoka, Robert S.
    Lehnert, Adrienne L.
    Hunter, William C. J.
    Ciarrocchi, Esther
    Bisogni, Maria G.
    Morrocchi, Matteo
    Del Guerra, Alberto
    2013 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2013,