A DEPTH-ENCODED PET DETECTOR

被引:38
|
作者
BARTZAKOS, P [1 ]
THOMPSON, CJ [1 ]
机构
[1] MCGILL UNIV,MED PHYS UNIT,MONTREAL H3A 2B4,QUEBEC,CANADA
关键词
D O I
10.1109/23.289382
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Radial blurring in high-resolution positron emission tomography (PET) can be reduced by measuring the PET photon's depth of interaction within a detector crystal. Simulations show that dividing the crystal into two equiprobable regions of interaction almost eliminates this blurring. This has been done using a black band placed around a 3 cm BGO crystal, 1 cm from the front. A PMT collects light from the rear of the crystal for energy and timing measurement. Another PMT collects light through a light guide from the top-front of the crystal to find the depth of interaction. Gamma rays interacting in the front of the crystal produce a strong, signal in the depth-sensing PMT, while interactions behind the band produce weaker (< 10%) signal. The intensity of the light recorded by the energy PMT is almost constant (within +/- 3%) for interactions at any depth in the crystal.
引用
收藏
页码:732 / 738
页数:7
相关论文
共 50 条
  • [41] Design of a depth of interaction detector with a PS-PMT for PET
    Murayama, H
    Ishibashi, H
    Uchida, H
    Omura, T
    Yamashita, T
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2000, 47 (03) : 1045 - 1050
  • [42] Computer simulation for design of depth encoding multicrystal detector for PET
    Haneishi, H
    Nishiya, S
    Miyake, Y
    Murayama, H
    Omura, T
    Ishibashi, H
    1999 IEEE NUCLEAR SCIENCE SYMPOSIUM - CONFERENCE RECORD, VOLS 1-3, 1999, : 1207 - 1211
  • [43] Optimization of a depth of interaction encoding PET block detector fora PET/MRI insert
    Selfridge, Aaron R.
    Cherry, Simon R.
    Judenhofer, Martin S.
    PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (23):
  • [44] A depth-encoding PET detector for high resolution PET using 1 mm SiPMs
    Du, Junwei
    Bai, Xiaowei
    Cherry, Simon R.
    PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (16):
  • [45] An LSO scintillator array for a PET detector module with depth of interaction measurement
    Huber, JS
    Moses, WW
    Andreaco, MS
    Petterson, O
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2001, 48 (03) : 684 - 688
  • [46] Depth identification accuracy of a three layer phoswich PET detector module
    Natl Inst of Health, Bethesda, United States
    IEEE Trans Nucl Sci, 3 II (485-490):
  • [47] Computer simulation of four-layer depth of interaction detector for PET
    Sato, M
    Haneishi, H
    Inadama, N
    Murayama, H
    2003 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-5, 2004, : 1913 - 1917
  • [48] Parametric positioning of a continuous crystal PET detector with depth of interaction decoding
    Ling, T.
    Burnett, T. H.
    Lewellen, T. K.
    Miyaoka, R. S.
    PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (07): : 1843 - 1863
  • [49] Resolution improvement in PET using detector depth of interaction information.
    Dahlbom, M
    MacDonald, LR
    Cherry, SR
    Eriksson, L
    Casey, ME
    JOURNAL OF NUCLEAR MEDICINE, 1997, 38 (05) : 162 - 162
  • [50] Study of two-layer tapered depth of interaction PET detector
    Behnamian, Hadi
    Yousefnejad, Sirous
    Shafiee, Mehdi
    Rafiei, Alireza
    APPLIED RADIATION AND ISOTOPES, 2021, 174