A Homogeneous Water-Equivalent Anthropomorphic Phantom for Dosimetric Verification of Radiotherapy Plans

被引:11
|
作者
Arjunan, Manikandan [1 ,2 ]
Sekaran, Sureka Chandra [1 ]
Sarkar, Biplab [3 ]
Manikandan, Sujatha [4 ]
机构
[1] Bharathiar Univ, Dept Med Phys, Coimbatore 641046, Tamil Nadu, India
[2] Nagarjuna Hosp, Dept Radiat Oncol, Vijayawada, Andhra Pradesh, India
[3] Fortis Hosp, Dept Radiat Oncol, New Delhi, India
[4] Govt Gen Hosp, Dept Radiotherapy, Guntur, Andhra Pradesh, India
关键词
Alent radiological depth; tissue equivalent phantom; water equivalent depth (WED) phantom;
D O I
10.4103/jmp.JMP_123_17
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Water is treated as radiological equivalent to human tissue. While this seems justified, there is neither mathematical proof nor sufficient experimental evidence that a water phantom can be treated as equivalent to human tissue. The aim of this work is to simulate and validate a water phantom that is tissue equivalent in terms of the dosimetric characteristics of both water and human tissue Dynamic, intensity-modulated radiotherapy plans for two head and neck, one brain, one pelvis, and three lung/mediastinum cases were chosen for this study. Using a treatment planning system (TPS) (Eclipse, Varian Medical System, Polo Alto, CA, USA) and Anisotropic Analytic Algorithm in a grid resolution of 5 mm x 5 mm, a patient-equivalent water phantom was calculated from all rays in the isocentric plane as an array of water equivalent depths (d(WE)). These rays were plotted versus isocentric separation and ray-tracing direction. Planar doses were compared between the isocentric plane in the patient computed tomography and the water equivalent phantom using gamma criteria of 2%-2 mm and 3%-3 mm. Except in one lung case, >95% gamma agreement was seen when using 3%-3 mm and >90% pass rate was seen when using 2%-2 mm. For head and neck cases, gamma-fail was restricted to the periphery. For mediastinum cases, gamma-fail was restricted to the lungs. This study demonstrates that a heterogeneous patient can be converted to a water phantom with comparable dosimetric characteristics and disagreements restricted to the lung area for both modulated and open beams. Potential sources of error include the d(WE) calculation and TPS dose computation.
引用
收藏
页码:100 / 105
页数:6
相关论文
共 50 条
  • [41] A system for three-dimensional dosimetric verification of treatment plans in intensity-modulated radiotherapy with heavy ions
    Karger, CP
    Jäkel, O
    Hartmann, GH
    Heeg, P
    MEDICAL PHYSICS, 1999, 26 (10) : 2125 - 2132
  • [42] A high DQE water-equivalent EPID employing an array of plastic-scintillating fibers for simultaneous imaging and dosimetry in radiotherapy
    Blake, Samuel J.
    Cheng, Zhangkai
    McNamara, Aimee
    Lu, Minghui
    Vial, Philip
    Kuncic, Zdenka
    MEDICAL PHYSICS, 2018, 45 (05) : 2154 - 2168
  • [43] Dosimetric Verification of Deterministic Acuros XB Radiation Transport Algorithm for IMRT and VMAT Plans with the RPC H&N Phantom
    Han, T.
    Mourtada, F.
    Kisling, K.
    Mikell, J.
    Followill, D.
    Howell, R.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [44] Dosimetric verification in inhomogeneous phantom geometries for the XiO radiotherapy treatment planning system with 6-MV photon beams
    Kohno R.
    Kitou S.
    Hirano E.
    Kameoka S.
    Goka T.
    Nishio T.
    Miyagishi T.
    Ariji T.
    Kawashima M.
    Ogino T.
    Radiological Physics and Technology, 2009, 2 (1) : 87 - 96
  • [45] Verification in the water phantom of the irradiation time calculation done by the algorithm used in intraoperative radiotherapy
    Slyk, Maria
    Litoborski, Marcin
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2010, 15 (05) : 132 - 137
  • [46] POLYETHYLENE-BASED WATER-EQUIVALENT PHANTOM MATERIAL FOR X-RAY DOSIMETRY AT TUBE VOLTAGES FROM 10 TO 100 KV
    HERMANN, KP
    GEWORSKI, L
    MUTH, M
    HARDER, D
    PHYSICS IN MEDICINE AND BIOLOGY, 1985, 30 (11): : 1195 - 1200
  • [47] Preliminary study on the use of an inhomogeneous anthropomorphic Fricke gel phantom and 3D magnetic resonance dosimetry for verification of IMRT treatment plans
    Gum, F
    Scherer, J
    Bogner, L
    Solleder, M
    Rhein, B
    Bock, M
    PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (07): : N67 - N77
  • [48] Experimental verification of a non-invasive method to monitor the lateral pencil beam position in an anthropomorphic phantom for carbon-ion radiotherapy
    Felix-Bautista, R.
    Gehrke, T.
    Ghesquiere-Dierickx, L.
    Reimold, M.
    Amato, C.
    Turecek, D.
    Jakubek, J.
    Ellerbrock, M.
    Martisikova, M.
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (17):
  • [49] Study of PRESAGE® dosimeter for end-to-end 3D radiotherapy verification using an anthropomorphic phantom with bespoke dosimeter insert
    Mohyedin, Muhammad Zamir
    Zin, Hafiz Mohd
    Abubakar, Auwal
    Rahman, Ahmad Taufek Abdul
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2024, 47 (03) : 955 - 966
  • [50] Description and dosimetric verification of the PEREGRINE Monte Carlo dose calculation system for photon beams incident on a water phantom
    Siantar, CLH
    Walling, RS
    Daly, TP
    Faddegon, B
    Albright, N
    Bergstrom, P
    Chuang, C
    Garrett, D
    House, RK
    Knapp, D
    Wieczorek, DJ
    Verhey, LJ
    MEDICAL PHYSICS, 2001, 28 (07) : 1322 - 1337