A Homogeneous Water-Equivalent Anthropomorphic Phantom for Dosimetric Verification of Radiotherapy Plans

被引:11
|
作者
Arjunan, Manikandan [1 ,2 ]
Sekaran, Sureka Chandra [1 ]
Sarkar, Biplab [3 ]
Manikandan, Sujatha [4 ]
机构
[1] Bharathiar Univ, Dept Med Phys, Coimbatore 641046, Tamil Nadu, India
[2] Nagarjuna Hosp, Dept Radiat Oncol, Vijayawada, Andhra Pradesh, India
[3] Fortis Hosp, Dept Radiat Oncol, New Delhi, India
[4] Govt Gen Hosp, Dept Radiotherapy, Guntur, Andhra Pradesh, India
关键词
Alent radiological depth; tissue equivalent phantom; water equivalent depth (WED) phantom;
D O I
10.4103/jmp.JMP_123_17
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Water is treated as radiological equivalent to human tissue. While this seems justified, there is neither mathematical proof nor sufficient experimental evidence that a water phantom can be treated as equivalent to human tissue. The aim of this work is to simulate and validate a water phantom that is tissue equivalent in terms of the dosimetric characteristics of both water and human tissue Dynamic, intensity-modulated radiotherapy plans for two head and neck, one brain, one pelvis, and three lung/mediastinum cases were chosen for this study. Using a treatment planning system (TPS) (Eclipse, Varian Medical System, Polo Alto, CA, USA) and Anisotropic Analytic Algorithm in a grid resolution of 5 mm x 5 mm, a patient-equivalent water phantom was calculated from all rays in the isocentric plane as an array of water equivalent depths (d(WE)). These rays were plotted versus isocentric separation and ray-tracing direction. Planar doses were compared between the isocentric plane in the patient computed tomography and the water equivalent phantom using gamma criteria of 2%-2 mm and 3%-3 mm. Except in one lung case, >95% gamma agreement was seen when using 3%-3 mm and >90% pass rate was seen when using 2%-2 mm. For head and neck cases, gamma-fail was restricted to the periphery. For mediastinum cases, gamma-fail was restricted to the lungs. This study demonstrates that a heterogeneous patient can be converted to a water phantom with comparable dosimetric characteristics and disagreements restricted to the lung area for both modulated and open beams. Potential sources of error include the d(WE) calculation and TPS dose computation.
引用
收藏
页码:100 / 105
页数:6
相关论文
共 50 条
  • [21] A customizable anthropomorphic phantom for dosimetric verification of 3D-printed lung, tissue, and bone density materials
    Tino, Rance Bolislis
    Yeo, Adam Unjin
    Brandt, Milan
    Leary, Martin
    Kron, Tomas
    MEDICAL PHYSICS, 2022, 49 (01) : 52 - 69
  • [22] Dosimetric verification of cancer patient?s treatment plan using an anthropomorphic, 3D-printed phantom
    Waluk, Kacper
    Pietrzak, Jakub
    APPLIED RADIATION AND ISOTOPES, 2023, 191
  • [23] Investigating LETd optimization strategies in carbon ion radiotherapy for pancreatic cancer: a dosimetric study using an anthropomorphic phantom
    Baltazar, Filipa
    Longarino, Friderike K.
    Stengl, Christina
    Liermann, Jakob
    Mein, Stewart
    Debus, Juergen
    Tessonnier, Thomas
    Mairani, Andrea
    MEDICAL PHYSICS, 2025, 52 (03) : 1746 - 1757
  • [24] 3D dosimetric validation of motion compensation concepts in radiotherapy using an anthropomorphic dynamic lung phantom
    Mann, P.
    Witte, M.
    Moser, T.
    Lang, C.
    Runz, A.
    Johnen, W.
    Berger, M.
    Biederer, J.
    Karger, C. P.
    PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (02): : 573 - 595
  • [25] DOSIMETRIC VERIFICATION OF INTENSITY MODULATED RADIOTHERAPY (IMRT) TREATMENT PLANS FOR PROSTATE CANCER PATIENTS
    Mrcela, Iva
    Gregov, Marin
    Matanic, Ante
    Budanec, Mirjana
    Murgic, Jure
    Jaksic, Blanka
    Prpic, Marin
    Secan, Angela Prgomet
    Frobe, Ana
    ACTA CLINICA CROATICA, 2022, 61 : 21 - 27
  • [26] Characteristics of the Water-Equivalent SP34 Plastic Phantom for Film Dosimetry in a Clinical Linear Accelerator
    Lee, S. S.
    Cho, G. S.
    Choi, S. H.
    Kang, K. U.
    An, D. H.
    Chang, H. S.
    Kim, K. T.
    Kim, K. B.
    MEDICAL PHYSICS, 2024, 51 (10) : 7901 - 7901
  • [27] Enhanced efficiency in helical tomotherapy quality assurance using a custom-designed water-equivalent phantom
    Goddu, S. Murty
    Mutic, Sasa
    Pechenaya, Olga L.
    Chaudhari, Summer R.
    Garcia-Ramirez, Jose
    Rangaraj, Dharanipathy
    Klein, Eric E.
    Yang, Deshan
    Grigsby, James
    Low, Daniel A.
    PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (19): : 5663 - 5674
  • [28] Theoretical and empirical investigations of KCl:Eu2+ for nearly water-equivalent radiotherapy dosimetry
    Zheng, Yuanshui
    Han, Zhaohui
    Driewer, Joseph P.
    Low, Daniel A.
    Li, H. Harold
    MEDICAL PHYSICS, 2010, 37 (01) : 146 - 153
  • [29] Dosimetric verification of IMRT head and neck plans using a 3D printed heterogeneous phantom
    Olasolo Alonso, J.
    Collado Chamorro, P.
    Sanz Freire, C. J.
    Diaz Pascual, V.
    Vazquez Galinanes, A.
    RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S993 - S994
  • [30] Evaluation of the water-equivalent characteristics of the SP34 plastic phantom for film dosimetry in a clinical linear accelerator
    Kim, Kyo-Tae
    Choi, Yona
    Cho, Gyu-Seok
    Jang, Won-Il
    Yang, Kwang-Mo
    Lee, Soon-Sung
    Bahng, Jungbae
    PLOS ONE, 2023, 18 (10):