A Homogeneous Water-Equivalent Anthropomorphic Phantom for Dosimetric Verification of Radiotherapy Plans

被引:11
|
作者
Arjunan, Manikandan [1 ,2 ]
Sekaran, Sureka Chandra [1 ]
Sarkar, Biplab [3 ]
Manikandan, Sujatha [4 ]
机构
[1] Bharathiar Univ, Dept Med Phys, Coimbatore 641046, Tamil Nadu, India
[2] Nagarjuna Hosp, Dept Radiat Oncol, Vijayawada, Andhra Pradesh, India
[3] Fortis Hosp, Dept Radiat Oncol, New Delhi, India
[4] Govt Gen Hosp, Dept Radiotherapy, Guntur, Andhra Pradesh, India
关键词
Alent radiological depth; tissue equivalent phantom; water equivalent depth (WED) phantom;
D O I
10.4103/jmp.JMP_123_17
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Water is treated as radiological equivalent to human tissue. While this seems justified, there is neither mathematical proof nor sufficient experimental evidence that a water phantom can be treated as equivalent to human tissue. The aim of this work is to simulate and validate a water phantom that is tissue equivalent in terms of the dosimetric characteristics of both water and human tissue Dynamic, intensity-modulated radiotherapy plans for two head and neck, one brain, one pelvis, and three lung/mediastinum cases were chosen for this study. Using a treatment planning system (TPS) (Eclipse, Varian Medical System, Polo Alto, CA, USA) and Anisotropic Analytic Algorithm in a grid resolution of 5 mm x 5 mm, a patient-equivalent water phantom was calculated from all rays in the isocentric plane as an array of water equivalent depths (d(WE)). These rays were plotted versus isocentric separation and ray-tracing direction. Planar doses were compared between the isocentric plane in the patient computed tomography and the water equivalent phantom using gamma criteria of 2%-2 mm and 3%-3 mm. Except in one lung case, >95% gamma agreement was seen when using 3%-3 mm and >90% pass rate was seen when using 2%-2 mm. For head and neck cases, gamma-fail was restricted to the periphery. For mediastinum cases, gamma-fail was restricted to the lungs. This study demonstrates that a heterogeneous patient can be converted to a water phantom with comparable dosimetric characteristics and disagreements restricted to the lung area for both modulated and open beams. Potential sources of error include the d(WE) calculation and TPS dose computation.
引用
收藏
页码:100 / 105
页数:6
相关论文
共 50 条
  • [1] A novel phantom for dosimetric verification of gated SIB radiotherapy treatment plans
    Soultan, D.
    Yock, A.
    Cornell, M.
    Murphy, J.
    Gill, B.
    Song, W.
    Moiseenko, V.
    Cervino, L.
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S831 - S832
  • [2] Three-dimensional thermal analysis of thermal characteristics in a water-equivalent phantom thermal analysis of water-equivalent phantom
    Fuse, Hiraku
    Tomita, Fumihiro
    Fujisaki, Tatsuya
    Yasue, Kenji
    Miyakawa, Shin
    Koori, Norikazu
    Abe, Shinji
    RADIATION PHYSICS AND CHEMISTRY, 2024, 217
  • [3] Design and Development of an Anthropomorphic Heterogeneous Female Pelvic (AHFP) Phantom for Dosimetric Verification of Advance Radiotherapy
    Yadav N.
    Singh M.
    Prasad Mishra S.
    Ansari M.S.
    Mishra A.
    Iranian Journal of Medical Physics, 2024, 21 (01) : 64 - 70
  • [4] Ice as a water-equivalent solid medium for brachytherapy dosimetric measurements
    Song, Haijun
    Chen, Zhe
    Yue, Ning
    Wu, Qingrong
    Yin, Fang-fang
    RADIATION AND ENVIRONMENTAL BIOPHYSICS, 2009, 48 (02) : 145 - 151
  • [5] Ice as a water-equivalent solid medium for brachytherapy dosimetric measurements
    Haijun Song
    Zhe Chen
    Ning Yue
    Qingrong Wu
    Fang-fang Yin
    Radiation and Environmental Biophysics, 2009, 48 : 145 - 151
  • [6] Design, fabrication, and validation of a polymethyl methacrylate head phantom for dosimetric verification of cranial radiotherapy treatment plans
    Shaiju, V. S.
    Kumar, Rajesh
    Phani, Debjani
    Rajasekhar, K., V
    Zacharia, George
    Bhasi, Saju
    Nair, Raghuram K.
    JOURNAL OF MEDICAL PHYSICS, 2020, 45 (02) : 66 - 70
  • [7] Characterization of Nylon-12 as a water-equivalent solid phantom material for dosimetric measurements in therapeutic photon and electron beams
    Ade, Nicholas
    van Eeden, D.
    du Plessis, F. C. P.
    APPLIED RADIATION AND ISOTOPES, 2020, 155 (155)
  • [8] Dosimetric impact of density variations in Solid Water 457 water-equivalent slabs
    Litzenberg, Dale W.
    Amro, Hanan
    Prisciandaro, Joann I.
    Acosta, Eduardo
    Gallagher, Ian
    Roberts, Don A.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2011, 12 (03): : 231 - 247
  • [9] Comparison of Patient Fluoroscopic Water-Equivalent Thicknesses to a Common Patient-Equivalent Phantom
    Fong, G.
    Wunderle, K.
    MEDICAL PHYSICS, 2019, 46 (06) : E251 - E251
  • [10] Dosimetric verification of calculation and measurements in radiotherapy using IMRT phantom
    Inal, Aysun
    Akkurt, Iskender
    SCIENTIFIC RESEARCH AND ESSAYS, 2010, 5 (10): : 1213 - 1215