FAILURE OF THE RING POLYETHYLENE UNCEMENTED ACETABULAR CUP

被引:10
|
作者
HARPER, GD [1 ]
BULL, T [1 ]
COBB, AG [1 ]
BENTLEY, G [1 ]
机构
[1] INST ORTHOPAED,ROYAL NATL ORTHOPAED HOSP NHS TRUST,STANMORE HA7 4LP,MIDDX,ENGLAND
来源
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME | 1995年 / 77B卷 / 04期
关键词
D O I
10.1302/0301-620X.77B4.7615597
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We report a high rate of failure of the Ring polyethylene cementless cup caused largely by granulomatous osteolysis, We have reviewed 126 prostheses inserted from 1986 to 1992 at from 11 to 90 months after surgery, There was radiological evidence of osteolytic granulomas adjacent to the external surface of the cup in 32%, appearing on average at three years from operation, In a subgroup of 59 prostheses followed for at least four years the incidence of such changes was 54%. A total of 27 cups (22%) have required revision, 21 for granulomatous loosening at an average follow-up of five years. In the retrieved prostheses there was obvious polyethylene abrasion and histological examination confirmed the presence of polyethylene wear debris. We found no significant correlation of osteolysis with cup size, although smaller cups were predominant among those having revision.
引用
收藏
页码:557 / 561
页数:5
相关论文
共 50 条
  • [41] Uncemented hip revision cup as an alternative for T-type acetabular fractures: A cadaveric study
    Kussmaul, Adrian Cavalcanti
    Bruder, Jan
    Greiner, Axel
    Woiczinski, Matthias
    Thorwaechter, Christoph
    Dotzauer, Fabian
    Rubenbauer, Bianka
    Linhart, Christoph
    Boecker, Wolfgang
    Becker, Christopher A.
    ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2024, 110 (03)
  • [42] Wear of highly cross-linked polyethylene acetabular cup in Japan
    Oonishi, Hironobu
    Kim, Sok Chol
    Takao, Yasuhiro
    Kyomoto, Masayuki
    Iwamoto, Mikio
    Ueno, Masaru
    JOURNAL OF ARTHROPLASTY, 2006, 21 (07): : 944 - 949
  • [43] Polyethylene sterilized without irradiation in a polished uncemented acetabular component - A clinical and radiographic analysis
    Crockarell, John R., Jr.
    JOURNAL OF ARTHROPLASTY, 2008, 23 (06): : 911 - 915
  • [44] Allografting in uncemented acetabular revision
    Reikeraas, O
    Folleraas, G
    Winge, JF
    ORTHOPEDICS, 1998, 21 (11) : 1191 - 1194
  • [45] Acetabular cup displacement after polyethylene-cement interface failure: A complication of total hip replacement in seven dogs
    Hunter, S
    Dyce, J
    Butkus, L
    Olmstead, ML
    VETERINARY AND COMPARATIVE ORTHOPAEDICS AND TRAUMATOLOGY, 2003, 16 (02) : 99 - 104
  • [46] Accelerated polyethylene wear and early failure of the uncemented Reflection acetabular component: a ten to fifteen-year follow-up study
    Clement, Nicholas D.
    Mitchell, Martin
    Trayner, Morag E.
    Porter, Daniel E.
    Lawson, Graham M.
    Burnett, Richard
    INTERNATIONAL ORTHOPAEDICS, 2013, 37 (12) : 2337 - 2343
  • [47] A cemented cup with acetabular impaction bone grafting is more cost-effective than an uncemented cup in patients under 50 years
    Busch, Vincent J. J. F.
    Verschueren, Joost
    Adang, Eddy M.
    Lie, Stein A.
    Havelin, Leif I.
    Schreurs, Berend W.
    HIP INTERNATIONAL, 2016, 26 (01) : 43 - 49
  • [48] Uncemented acetabular components for arthritis after acetabular fracture
    Berry, DJ
    Halasy, M
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2002, (405) : 164 - 167
  • [49] Accelerated polyethylene wear and early failure of the uncemented Reflection acetabular component: a ten to fifteen-year follow-up study
    Nicholas D. Clement
    Martin Mitchell
    Morag E. Trayner
    Daniel E. Porter
    Graham M. Lawson
    Richard Burnett
    International Orthopaedics, 2013, 37 : 2337 - 2343
  • [50] AGGRESSIVE GRANULOMATOSIS FROM POLYETHYLENE FAILURE IN AN UNCEMENTED KNEE REPLACEMENT
    NOLAN, JF
    BUCKNILL, TM
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (01): : 23 - 24