FAILURE OF THE RING POLYETHYLENE UNCEMENTED ACETABULAR CUP

被引:10
|
作者
HARPER, GD [1 ]
BULL, T [1 ]
COBB, AG [1 ]
BENTLEY, G [1 ]
机构
[1] INST ORTHOPAED,ROYAL NATL ORTHOPAED HOSP NHS TRUST,STANMORE HA7 4LP,MIDDX,ENGLAND
来源
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME | 1995年 / 77B卷 / 04期
关键词
D O I
10.1302/0301-620X.77B4.7615597
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We report a high rate of failure of the Ring polyethylene cementless cup caused largely by granulomatous osteolysis, We have reviewed 126 prostheses inserted from 1986 to 1992 at from 11 to 90 months after surgery, There was radiological evidence of osteolytic granulomas adjacent to the external surface of the cup in 32%, appearing on average at three years from operation, In a subgroup of 59 prostheses followed for at least four years the incidence of such changes was 54%. A total of 27 cups (22%) have required revision, 21 for granulomatous loosening at an average follow-up of five years. In the retrieved prostheses there was obvious polyethylene abrasion and histological examination confirmed the presence of polyethylene wear debris. We found no significant correlation of osteolysis with cup size, although smaller cups were predominant among those having revision.
引用
收藏
页码:557 / 561
页数:5
相关论文
共 50 条
  • [31] Type of motion and lubricant in wear simulation of polyethylene acetabular cup
    Saikko, V.
    Ahlroos, T.
    Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 1999, 213 (04): : 301 - 310
  • [32] THE EFFECT OF POLYETHYLENE CUP PENETRATION ON ACETABULAR CEMENT FATIGUE LIFE
    Coultrup, Oliver J.
    Hunt, Chris
    Taylor, Mark
    PROCEEDINGS OF THE ASME SUMMER BIOENGINEERING CONFERENCE - 2009, PT A AND B, 2009, : 953 - 954
  • [33] Failure criteria for polyethylene acetabular cups
    Burger, N. D. L.
    de Vaal, P. L.
    Meyer, J. P.
    SOUTH AFRICAN JOURNAL OF SCIENCE, 2006, 102 (11-12) : 572 - 574
  • [34] UNCEMENTED ACETABULAR REPLACEMENT
    RING, PA
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 1983, 101 (04) : 225 - 229
  • [35] Surgical management of polyethylene wear and pelvic osteolysis with modular uncemented acetabular components
    Naudie, DDR
    Engh, CA
    JOURNAL OF ARTHROPLASTY, 2004, 19 (04): : 124 - 129
  • [36] Impingement of the iliopsoas muscle after revision of a loosened acetabular cup with an acetabular reinforcement ring
    Schuh, A
    Schraml, A
    Zeiler, G
    ZENTRALBLATT FUR CHIRURGIE, 2003, 128 (03): : 209 - 211
  • [37] Mechanism of failure in hips with an uncemented, all polyethylene socket
    Llinás, A
    Sarmiento, A
    Ebramzadeh, E
    Park, SH
    Campbell, P
    McKellop, HA
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1999, (362) : 145 - 155
  • [38] Different cup migration in rheumatoid arthritis and arthrosis - A radiographic analysis of 127 uncemented acetabular cups
    Dominkus, M
    Wanivenhaus, AH
    Morscher, M
    Powell, G
    Krismer, M
    Wolfl, G
    ACTA ORTHOPAEDICA SCANDINAVICA, 1998, 69 (05): : 455 - 462
  • [39] Finite Element Simulations for Investigating the Cause of Catastrophic Wear and/or Failure of Polyethylene Acetabular Cup Liner in Hip Prosthesis
    Cho, Changhee
    Mori, Toshiharu
    Kawasaki, Makoto
    COMPUTER METHODS, IMAGING AND VISUALIZATION IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2020, 36 : 539 - 549
  • [40] Use of uncemented Beznoska acetabular cup - Mid-term results at the orthopedic ward in pelhrimov
    Letocha, J.
    Behounek, J.
    Sikotak, M.
    ACTA CHIRURGIAE ORTHOPAEDICAE ET TRAUMATOLOGIAE CECHOSLOVACA, 2007, 74 (03) : 195 - 201