FAILURE OF THE RING POLYETHYLENE UNCEMENTED ACETABULAR CUP

被引:10
|
作者
HARPER, GD [1 ]
BULL, T [1 ]
COBB, AG [1 ]
BENTLEY, G [1 ]
机构
[1] INST ORTHOPAED,ROYAL NATL ORTHOPAED HOSP NHS TRUST,STANMORE HA7 4LP,MIDDX,ENGLAND
来源
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME | 1995年 / 77B卷 / 04期
关键词
D O I
10.1302/0301-620X.77B4.7615597
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We report a high rate of failure of the Ring polyethylene cementless cup caused largely by granulomatous osteolysis, We have reviewed 126 prostheses inserted from 1986 to 1992 at from 11 to 90 months after surgery, There was radiological evidence of osteolytic granulomas adjacent to the external surface of the cup in 32%, appearing on average at three years from operation, In a subgroup of 59 prostheses followed for at least four years the incidence of such changes was 54%. A total of 27 cups (22%) have required revision, 21 for granulomatous loosening at an average follow-up of five years. In the retrieved prostheses there was obvious polyethylene abrasion and histological examination confirmed the presence of polyethylene wear debris. We found no significant correlation of osteolysis with cup size, although smaller cups were predominant among those having revision.
引用
收藏
页码:557 / 561
页数:5
相关论文
共 50 条
  • [21] FRACTURE OF THE POLYETHYLENE ACETABULAR CUP IN TOTAL HIP ARTHROPLASTIES
    STUCK, KJ
    FALAHEE, MH
    BRANDON, CJ
    JOURNAL OF THE CANADIAN ASSOCIATION OF RADIOLOGISTS-JOURNAL DE L ASSOCIATION CANADIENNE DES RADIOLOGISTES, 1988, 39 (01): : 65 - 67
  • [22] Polyethylene wear in the charnley offset bore acetabular cup
    IzquierdoAvino, RJ
    Siney, PD
    Wroblewski, BM
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1996, 78B (01): : 82 - 84
  • [23] Dissociation of the metal inlay from the polyethylene liner in an uncemented threaded cup
    Milosev, I
    Trebse, R
    Kovac, S
    Cör, A
    Campbell, P
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2005, 125 (02) : 134 - 141
  • [24] Dissociation of the metal inlay from the polyethylene liner in an uncemented threaded cup
    Ingrid Milošev
    Rihard Trebše
    Simon Kovač
    Andrej Cör
    Pat Campbell
    Archives of Orthopaedic and Trauma Surgery, 2005, 125 : 134 - 141
  • [25] Cup Press Fit in Uncemented THA Depends on Sex, Acetabular Shape, and Surgical Technique
    Garcia-Rey, Eduardo
    Garcia-Cimbrelo, Eduardo
    Cruz-Pardos, Ana
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2012, 470 (11) : 3014 - 3023
  • [26] Migration of the uncemented Harris-Galante acetabular cup - Results of the einbildroentgenanalyse (EBRA) method
    Hendrich, C
    Bahlmann, J
    Eulert, J
    JOURNAL OF ARTHROPLASTY, 1997, 12 (08): : 889 - 895
  • [27] Using acetabular fossa as a guide for anticipated inclination of uncemented cup in total hip replacement
    Li, Junwei
    Gao, Xu
    Yang, Guanghui
    Zhang, Yanru
    INTERNATIONAL JOURNAL OF CLINICAL AND EXPERIMENTAL MEDICINE, 2015, 8 (01): : 181 - 187
  • [28] Revision surgery of acetabular polyethylene wear - cup retention or revision?
    Yu, T. -Ch.
    Bioceramics and Alternative Bearings in Joint Arthroplasty, 2007, : 295 - 296
  • [29] UNCEMENTED ACETABULAR REPLACEMENT
    RING, PA
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1986, 68 (04): : 669 - 669
  • [30] Type of motion and lubricant in wear simulation of polyethylene acetabular cup
    Saikko, V
    Ahlroos, T
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 1999, 213 (H4) : 301 - 310