FAILURE OF THE RING POLYETHYLENE UNCEMENTED ACETABULAR CUP

被引:10
|
作者
HARPER, GD [1 ]
BULL, T [1 ]
COBB, AG [1 ]
BENTLEY, G [1 ]
机构
[1] INST ORTHOPAED,ROYAL NATL ORTHOPAED HOSP NHS TRUST,STANMORE HA7 4LP,MIDDX,ENGLAND
来源
关键词
D O I
10.1302/0301-620X.77B4.7615597
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We report a high rate of failure of the Ring polyethylene cementless cup caused largely by granulomatous osteolysis, We have reviewed 126 prostheses inserted from 1986 to 1992 at from 11 to 90 months after surgery, There was radiological evidence of osteolytic granulomas adjacent to the external surface of the cup in 32%, appearing on average at three years from operation, In a subgroup of 59 prostheses followed for at least four years the incidence of such changes was 54%. A total of 27 cups (22%) have required revision, 21 for granulomatous loosening at an average follow-up of five years. In the retrieved prostheses there was obvious polyethylene abrasion and histological examination confirmed the presence of polyethylene wear debris. We found no significant correlation of osteolysis with cup size, although smaller cups were predominant among those having revision.
引用
收藏
页码:557 / 561
页数:5
相关论文
共 50 条
  • [1] Acute polyethylene fracture in an uncemented acetabular cup
    Gross, AE
    Dust, WN
    CANADIAN JOURNAL OF SURGERY, 1997, 40 (04) : 310 - 312
  • [2] THE RING UNCEMENTED POLYETHYLENE CUP IN THE ABNORMAL ACETABULUM
    NUNN, D
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1987, 69 (05): : 756 - 760
  • [3] CATASTROPHIC FAILURE OF THE POLYETHYLENE LINER OF UNCEMENTED ACETABULAR COMPONENTS
    BERRY, DJ
    BARNES, CL
    SCOTT, RD
    CABANELA, ME
    POSS, R
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1994, 76B (04): : 575 - 578
  • [4] Aseptic loosening of an uncemented Acetabular Cup
    Devarajan, G
    Gregori, A
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2000, 82A (08): : 1195 - 1196
  • [5] Severe polyethylene wear in uncemented acetabular cup system components - A report of 5 cases
    Patel, J
    Scott, JE
    Radford, WJP
    JOURNAL OF ARTHROPLASTY, 1999, 14 (05): : 635 - 636
  • [6] Aseptic loosening of an uncemented Acetabular Cup - Reply
    Peterson, MB
    Poulsen, IH
    Thomsen, J
    Solgaard, S
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2000, 82A (08): : 1196 - 1196
  • [7] POLYETHYLENE WEAR IN UNCEMENTED ACETABULAR COMPONENTS
    HERNANDEZ, JR
    KEATING, EM
    FARIS, PM
    MEDING, JB
    RITTER, MA
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1994, 76B (02): : 263 - 266
  • [8] Failure of the polyethylene uncemented acetabular cup fixed onto the bone: study of 32 Freeman cups at an average of 9 years follow up
    Boisgard, S
    Levai, JP
    Jaud, E
    Guyonnet, G
    Chatenet, T
    REVUE DE CHIRURGIE ORTHOPEDIQUE ET REPARATRICE DE L APPAREIL MOTEUR, 1998, 84 (08): : 700 - 704
  • [9] TOTAL HIP-REPLACEMENT WITH AN UNCEMENTED ACETABULAR CUP
    BODTKER, S
    KRAMHOFT, M
    STRANDBERG, C
    SOLGAARD, S
    ACTA ORTHOPAEDICA SCANDINAVICA, 1987, 58 (06): : 693 - 693
  • [10] Uncemented acetabular components with polyethylene or alumina liners
    Rainer Schmidt
    Gerd Willmann
    Rocco Paolo Pitto
    International Orthopaedics, 2003, 27 : 85 - 89