FATIGUE FAILURE OF NONCEMENTED POROUS-COATED IMPLANTS - A RETRIEVAL STUDY

被引:33
|
作者
COOK, SD
THOMAS, KA
机构
来源
关键词
D O I
10.1302/0301-620X.73B1.1991767
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The causes of mechanical failure of five noncemented porous-coated components were studied. There were two cobalt-chromium alloy and three titanium alloy implants which fractured after 12 to 48 months. The implants included one acetabular component, and one femoral condylar, one patellar and two tibial components. Examination of the fractured surfaces revealed fatigue to be the mechanism of failure in all cases. The porous coating and the processes required for its fabrication had resulted in weakening and reduction of substrate thickness. Additional factors were stress concentration due to limited, localised bone ingrowth, and some features of the design of the implants.
引用
收藏
页码:20 / 24
页数:5
相关论文
共 50 条
  • [1] INFLAMMATORY RESPONSE IN RETRIEVED NONCEMENTED POROUS-COATED IMPLANTS
    COOK, SD
    MCCLUSKEY, LC
    MARTIN, PC
    HADDAD, RJ
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1991, (264) : 209 - 222
  • [2] EFFECTS OF RADIATION ON FIXATION OF NONCEMENTED POROUS-COATED IMPLANTS IN A CANINE MODEL
    SUMNER, DR
    TURNER, TM
    PIERSON, RH
    KIENAPFEL, H
    URBAN, RM
    LIEBNER, EJ
    GALANTE, JO
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1990, 72A (10): : 1527 - 1533
  • [3] NONCEMENTED POROUS-COATED ANATOMIC TOTAL HIP-ARTHROPLASTY
    DODGE, BM
    FITZRANDOLPH, R
    COLLINS, DN
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1991, (269) : 16 - 24
  • [4] A COMPARISON OF 3 VARIETIES OF NONCEMENTED POROUS-COATED HIP-REPLACEMENT
    HADDAD, RJ
    COOK, SD
    BRINKER, MR
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1990, 72 (01): : 2 - 8
  • [5] CLINICAL AND ROENTGENOGRAPHIC EVALUATION OF NONCEMENTED POROUS-COATED ANATOMIC MEDULLARY LOCKING (AML) AND POROUS-COATED ANATOMIC (PCA) TOTAL HIP ARTHROPLASTIES
    HADDAD, RJ
    SKALLEY, TC
    COOK, SD
    BRINKER, MR
    CHERAMIE, J
    MEYER, R
    MISSRY, J
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1990, (258) : 176 - 182
  • [6] METALLURGICAL CONSIDERATIONS IN FABRICATION OF POROUS-COATED SURGICAL IMPLANTS
    WEATHERLY, GC
    YUE, S
    KILNER, T
    PILLIAR, R
    CIM BULLETIN, 1984, 77 (866): : 62 - 62
  • [7] THE CASE FOR POROUS-COATED HIP IMPLANTS - THE FEMORAL SIDE
    ENGH, CA
    GLASSMAN, AH
    SUTHERS, KE
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1990, (261) : 63 - 81
  • [8] CURRENT CONCEPTS REVIEW - BIOLOGICAL FIXATION OF POROUS-COATED IMPLANTS
    HADDAD, RJ
    COOK, SD
    THOMAS, KA
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1987, 69A (09): : 1459 - 1466
  • [9] EFFECTS OF INDOMETHACIN ON BIOLOGIC FIXATION OF POROUS-COATED TITANIUM IMPLANTS
    COOK, SD
    BARRACK, RL
    DALTON, JE
    THOMAS, KA
    BROWN, TD
    JOURNAL OF ARTHROPLASTY, 1995, 10 (03): : 351 - 358
  • [10] Osseointegration of hydroxyapatite porous-coated femoral implants in a canine model
    Karabatsos, B
    Myerthall, SL
    Fornasier, VL
    Binnington, A
    Maistrelli, GL
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2001, (392) : 442 - 449