FATIGUE FAILURE OF NONCEMENTED POROUS-COATED IMPLANTS - A RETRIEVAL STUDY

被引:33
|
作者
COOK, SD
THOMAS, KA
机构
来源
关键词
D O I
10.1302/0301-620X.73B1.1991767
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The causes of mechanical failure of five noncemented porous-coated components were studied. There were two cobalt-chromium alloy and three titanium alloy implants which fractured after 12 to 48 months. The implants included one acetabular component, and one femoral condylar, one patellar and two tibial components. Examination of the fractured surfaces revealed fatigue to be the mechanism of failure in all cases. The porous coating and the processes required for its fabrication had resulted in weakening and reduction of substrate thickness. Additional factors were stress concentration due to limited, localised bone ingrowth, and some features of the design of the implants.
引用
收藏
页码:20 / 24
页数:5
相关论文
共 50 条
  • [21] Novel low profile porous-coated implants - Studies in dug mandibular sites
    Pilliar, RM
    Deporter, DA
    Watson, PA
    Pharoah, M
    Valiquette, N
    JOURNAL OF DENTAL RESEARCH, 1997, 76 : 2154 - 2154
  • [22] MINERAL APPOSITION RATES OF HUMAN CANCELLOUS BONE AT THE INTERFACE OF POROUS-COATED IMPLANTS
    BLOEBAUM, RD
    BACHUS, KN
    MOMBERGER, NG
    HOFMANN, AA
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1994, 28 (05): : 537 - 544
  • [23] Porous-coated fixation: The experience
    Mallory, TH
    ORTHOPEDICS, 1996, 19 (09) : 736 - 738
  • [24] Acid-etched microtexture for enhancement of bone growth into porous-coated implants
    Hacking, SA
    Harvey, EJ
    Tanzer, M
    Krygier, JJ
    Bobyn, JD
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2003, 85B (08): : 1182 - 1189
  • [25] The outcome of rerevision of an extensively porous-coated stem with another extensively porous-coated stem
    Hamilton, William G.
    McAuley, James P.
    Tabaraee, Ehsan
    Engh, Charles A., Sr.
    JOURNAL OF ARTHROPLASTY, 2008, 23 (02): : 170 - 174
  • [26] Effect of noninvasive low intensity ultrasound on bone growth into porous-coated implants
    Tanzer, M
    Harvey, E
    Kay, A
    Morton, P
    Bobyn, JD
    JOURNAL OF ORTHOPAEDIC RESEARCH, 1996, 14 (06) : 901 - 906
  • [27] A HISTOMORPHOMETRIC AND HISTOLOGIC ANALYSIS OF THE IMPLANT INTERFACE IN 5 SUCCESSFUL, AUTOPSY-RETRIEVED, NONCEMENTED POROUS-COATED KNEE ARTHROPLASTIES
    VIGORITA, VJ
    MINKOWITZ, B
    DICHIARA, JF
    HIGHAM, PA
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1993, (293) : 211 - 218
  • [28] The odyssey of porous-coated fixation
    Engh, CA
    Hopper, RH
    JOURNAL OF ARTHROPLASTY, 2002, 17 (04): : 102 - 107
  • [29] POROUS-COATED FIXATION FOREVER
    ENGH, CA
    ORTHOPEDICS, 1994, 17 (09) : 775 - 776
  • [30] ENHANCED INGROWTH OF POROUS-COATED COCR IMPLANTS PLASMA-SPRAYED WITH TRICALCIUM PHOSPHATE
    CHAE, JC
    COLLIER, JP
    MAYOR, MB
    SURPRENANT, VA
    DAUPHINAIS, LA
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1992, 26 (01): : 93 - 102