FATIGUE FAILURE OF NONCEMENTED POROUS-COATED IMPLANTS - A RETRIEVAL STUDY

被引:33
|
作者
COOK, SD
THOMAS, KA
机构
来源
关键词
D O I
10.1302/0301-620X.73B1.1991767
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The causes of mechanical failure of five noncemented porous-coated components were studied. There were two cobalt-chromium alloy and three titanium alloy implants which fractured after 12 to 48 months. The implants included one acetabular component, and one femoral condylar, one patellar and two tibial components. Examination of the fractured surfaces revealed fatigue to be the mechanism of failure in all cases. The porous coating and the processes required for its fabrication had resulted in weakening and reduction of substrate thickness. Additional factors were stress concentration due to limited, localised bone ingrowth, and some features of the design of the implants.
引用
收藏
页码:20 / 24
页数:5
相关论文
共 50 条
  • [41] Failure of a proximally porous-coated femoral prosthesis in revision total hip arthroplasty
    Woolson, ST
    Delaney, TJ
    JOURNAL OF ARTHROPLASTY, 1995, 10 : S22 - S28
  • [42] INITIAL STABILITY OF POROUS-COATED ACETABULAR IMPLANTS - THE EFFECT OF SCREW PLACEMENT, SCREW TIGHTNESS, DEFECT TYPE, AND OVERSIZE IMPLANTS
    HADJARI, MH
    HOLLIS, JM
    HOFMANN, OE
    FLAHIFF, CM
    NELSON, CL
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1994, (307) : 117 - 123
  • [43] FINITE-ELEMENT ANALYSIS OF CRESTAL BONE LOSS AROUND POROUS-COATED DENTAL IMPLANTS
    VAILLANCOURT, H
    PILLIAR, RM
    MCCAMMOND, D
    JOURNAL OF APPLIED BIOMATERIALS, 1995, 6 (04) : 267 - 282
  • [44] Improved fixation of porous-coated versus grit-blasted surface texture of hydroxyapatite-coated implants in dogs
    Overgaard, S
    Lind, M
    Rahbek, O
    Bunger, C
    Soballe, K
    ACTA ORTHOPAEDICA SCANDINAVICA, 1997, 68 (04): : 337 - 343
  • [45] FATIGUE PROPERTIES OF CARBON-COATED AND POROUS-COATED TI-6AL-4V ALLOY
    COOK, SD
    GEORGETTE, FS
    SKINNER, HB
    HADDAD, RJ
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1984, 18 (05): : 497 - 512
  • [46] CIPROFLOXACIN ATTACHMENT TO POROUS-COATED TITANIUM SURFACES
    DUNN, DS
    RAGHAVAN, S
    VOLZ, RG
    JOURNAL OF APPLIED BIOMATERIALS, 1994, 5 (04) : 325 - 331
  • [47] Porous-coated femoral fixation: The long and short of it
    McAuley, JP
    ORTHOPEDICS, 2000, 23 (09) : 917 - 918
  • [48] Ciprofloxacin attachment to porous-coated titanium surfaces
    Dunn, D.S.
    Raghavan, S.
    Volz, R.G.
    1600, John Wiley & Sons Inc, New York, NY, United States (05):
  • [49] Bone remodelling in the pores and around load bearing transchondral isoelastic porous-coated glassy carbon implants: Experimental study in rabbits
    Tarvainen, T
    Paronen, I
    Tunturi, T
    Rautavuori, J
    Tormala, P
    Patiala, H
    Rokkanen, P
    JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 1998, 9 (09) : 509 - 515
  • [50] Bone remodelling in the pores and around load bearing transchondral isoelastic porous-coated glassy carbon implants: Experimental study in rabbits
    T. Tarvainen
    I. Paronen
    T. Tunturi
    J. Rautavuori
    P. Tormala
    H. Patiala
    P. Rokkanen
    Journal of Materials Science: Materials in Medicine, 1998, 9 : 509 - 515