Study Design, We evaluated a computer-assisted surgical tool for inserting pedicle screws. Objectives. This study reviewed the feasibility, usefulness, and accuracy of the proposed tool. Summary of Background Data, Reviews documented neurovascular damage caused by screw misplacement. Currently, screw hole position is assessed by radiologic means and curette palpation. Methods, Three sheep vertebrae and one artificial object were reconstructed three-dimensionally from computed tomography scan slices. At surgery, the surgeon's movements were displayed relative to the three-dimensional vertebrae on a computer screen. The tool was used to detect pedicles and to verify the position of drilled holes. In our laboratory, we calculated the system's accuracy by taking measurements on the artificial object. Results. All pedicles were identified with the computer. Five of the six drilled hole: positions were correctly represented. An accuracy of 4.5 mm +/- 1.1 mm RMS (root of the mean squared) and 1.6 degrees +/- 1.2 degrees were calculated. Conclusions, Results suggested the proposed system could be useful for pedicle detection and assessing the intravertebral location of a drilled hole. The proposed system could be used for many different orthopedic procedures where structures are hidden from the surgeon's view.
引用
收藏
页码:1208 / 1212
页数:5
相关论文
共 50 条
  • [31] Computer-assisted insertion of pedicle screws
    Li, QH
    Holdener, HJ
    Zamorano, L
    King, P
    Jiang, ZW
    Vinas, FC
    Nolte, L
    Visarius, H
    Diaz, F
    VISUALIZATION IN BIOMEDICAL COMPUTING, 1996, 1131 : 571 - 580
  • [32] Pullout resistance of conventional pedicle screw implantation in comparison to fluoroscopic computer-assisted technique
    Linhardt, O.
    Perlick, L.
    Renkawitz, T.
    Matussek, J.
    Plitz, W.
    Grifka, J.
    ZENTRALBLATT FUR CHIRURGIE, 2006, 131 (05): : 407 - 410
  • [33] Radiation dose for pedicle screw insertion -: Fluoroscopic method versus computer-assisted surgery
    Slomczykowski, M
    Roberto, M
    Schneeberger, P
    Ozdoba, C
    Vock, P
    SPINE, 1999, 24 (10) : 975 - 982
  • [34] Computer-assisted orthopaedic surgery (CAOS): From pedicle screw insertion to further applications
    Berlemann, U
    Langlotz, F
    Langlotz, U
    Nolte, LP
    ORTHOPADE, 1997, 26 (05): : 463 - 469
  • [35] Cervical pedicle screw insertion: Assessment of safety and accuracy with computer-assisted image guidance
    Kamimura, M
    Ebara, S
    Itoh, H
    Tateiwa, Y
    Kinoshita, T
    Takaoka, K
    JOURNAL OF SPINAL DISORDERS, 2000, 13 (03): : 218 - 224
  • [36] IMAGE-GUIDED COMPUTER-ASSISTED SPINE SURGERY - A PILOT-STUDY ON SPINAL PEDICLE FIXATION
    NOLTE, LP
    ZAMORANO, L
    JIANG, ZW
    WANG, QH
    LANGLOTZ, F
    DIAZ, FG
    JOURNAL OF NEUROSURGERY, 1995, 82 (02) : A343 - A344
  • [37] Iliac Screw Fixation Using Computer-Assisted Computer Tomographic Image Guidance: Technical Note
    Shin, John H.
    Hoh, Daniel J.
    Kalfas, Iain H.
    NEUROSURGERY, 2012, 70
  • [38] Feasibility of anterior pedicle screw fixation in lumbosacral spine: a radiographic and cadaveric study
    Xu, Wei-Xing
    Xu, Bin
    Ding, Wei-Guo
    Sheng, Hong-Feng
    Lu, Di
    Hu, Tian-Hong
    ANNALS OF TRANSLATIONAL MEDICINE, 2021, 9 (12)
  • [39] The application of a computer-assisted thermoplastic membrane navigation system in screw fixation of the sacroiliac joint - A clinical study
    Zheng Zhanle
    Zhang Yingze
    Hou Zhiyong
    Hao Jiandong
    Zhai Fushan
    Su Yanling
    Pan Jinshe
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2012, 43 (04): : 495 - 499
  • [40] Cervical pedicle screw insertion: Assessment of safety and accuracy with computer-assisted image guidance - Commentary
    Kamimura, M
    Ebara, S
    Itoh, H
    Tateiwa, Y
    Kinoshita, T
    Takaoka, K
    JOURNAL OF SPINAL DISORDERS, 2000, 13 (04): : 275 - 275