DETERMINATION OF TOTAL-BODY WATER IN UREMIC PATIENTS BY BIOELECTRICAL-IMPEDANCE

被引:32
|
作者
KONG, CH
THOMPSON, CM
LEWIS, CA
HILL, PD
THOMPSON, FD
机构
[1] MIDDLESEX HOSP,LONDON W1N 8AA,ENGLAND
[2] UNIV LONDON KINGS COLL HOSP,LONDON SE5 8RX,ENGLAND
关键词
BIOELECTRICAL IMPEDANCE; BODY COMPOSITION; DIALYSIS; TOTAL BODY WATER;
D O I
10.1093/ndt/8.8.716
中图分类号
R3 [基础医学]; R4 [临床医学];
学科分类号
1001 ; 1002 ; 100602 ;
摘要
The measurement of total body water by bioelectrical impedance in a group of renal patients was evaluated against the tritium dilution method. The effect of haemodialysis and the presence of peritoneal dialysate on the impedance were also investigated. The correlation between the two methods is r=0.90 with a residual standard deviation of 3.7. The standard deviation of the differences between the two methods against the means was 3.66 which means that total body water (TBW) estimated by the bioelectrical impedance (BEI) method may be 6.181 (X +/- 2 SD) above or 8.381 below the (H2O)-H-3 method. The BEI method overestimated the actual weight loss after haemodialysis (3.87 +/- 1.71 versus 2.43 +/- 1.81) but underestimated the volume of peritoneal dialysate in situ. The BEI method would not be appropriate for use in assessing total body water and monitoring acute volume changes in patients with renal failure who are on strict fluid restriction.
引用
收藏
页码:716 / 719
页数:4
相关论文
共 50 条
  • [31] TOTAL-BODY GAMMAGRAPHY IN UREMIC PATIENTS
    CAVALLI, PL
    CAMUZZINI, GF
    GHEZZI, P
    PALANCA, R
    ROVERE, A
    TORTORE, P
    KIDNEY INTERNATIONAL, 1976, 9 (06) : 527 - 527
  • [32] COMPARISON OF 2 BIOELECTRICAL IMPEDANCE ANALYSIS MODELS FOR TOTAL-BODY WATER MEASUREMENT IN CHILDREN
    DANFORD, LC
    SCHOELLER, DA
    KUSHNER, RF
    ANNALS OF HUMAN BIOLOGY, 1992, 19 (06) : 603 - 607
  • [33] CHANGES IN TOTAL-BODY WATER MEASURED BY BIOELECTRICAL IMPEDANCE IN ALCOHOL-INDUCED MIGRAINE
    MARTELLETTI, P
    DELOREZZO, A
    SASSO, GF
    BARRA, P
    BRANCATI, A
    GIACOVAZZO, M
    CEPHALALGIA, 1991, 11 : 183 - 184
  • [34] VALIDATION OF BIOELECTRICAL-IMPEDANCE MEASUREMENTS AS A METHOD TO ESTIMATE BODY-WATER COMPARTMENTS
    LICHTENBELT, WDV
    WESTERTERP, KR
    WOUTERS, L
    LUIJENDIJK, SCM
    AMERICAN JOURNAL OF CLINICAL NUTRITION, 1994, 60 (02): : 159 - 166
  • [35] CAN BIOELECTRICAL IMPEDANCE ANALYSIS (BIA) AND TOTAL-BODY ELECTRICAL-CONDUCTIVITY (TOBEC) DETECT SMALL LOCALIZED CHANGES IN TOTAL-BODY WATER
    RALLISON, L
    KUSHNER, RF
    SCHOELLER, DA
    PENN, D
    CLINICAL RESEARCH, 1992, 40 (02): : A636 - A636
  • [36] MEASUREMENT OF EXTRACELLULAR AND TOTAL-BODY WATER OF RATS USING MULTIPLE FREQUENCY BIOELECTRICAL IMPEDANCE ANALYSIS
    CORNISH, BH
    WARD, LC
    THOMAS, BJ
    NUTRITION RESEARCH, 1992, 12 (4-5) : 657 - 666
  • [37] TOTAL-BODY WATER AND WHOLE-BODY IMPEDANCE
    OBERLANDER, J
    CROSBY, L
    GIANDOMENICO, A
    STEIN, T
    MIKUTA, J
    MULLEN, J
    CLINICAL RESEARCH, 1980, 28 (03): : A621 - A621
  • [38] BIOELECTRICAL IMPEDANCE ANALYSIS TO DETERMINE FAT-FREE MASS, TOTAL-BODY WATER AND BODY-FAT
    VANLOAN, MD
    SPORTS MEDICINE, 1990, 10 (04) : 205 - 217
  • [39] COMPARISON OF TOTAL-BODY WATER (TBW) DETERMINATION BY BIOELECTRICAL IMPEDANCE ANALYSIS (BIA), ANTHROPOMETRY, AND D2O DILUTION
    KUSHNER, RF
    SCHOELLER, DA
    BOWMAN, BB
    AMERICAN JOURNAL OF CLINICAL NUTRITION, 1984, 39 (04): : 658 - 658
  • [40] COMPARISON OF TOTAL-BODY WATER (TBW) DETERMINATION BY BIOELECTRICAL IMPEDANCE ANALYSIS (BIA), ANTHROPOMETRY, AND D2O DILUTION
    KUSHNER, RF
    SCHOELLER, DA
    BOWMAN, BB
    CLINICAL RESEARCH, 1984, 32 (02): : A630 - A630