DETERMINATION OF TOTAL-BODY WATER IN UREMIC PATIENTS BY BIOELECTRICAL-IMPEDANCE

被引:32
|
作者
KONG, CH
THOMPSON, CM
LEWIS, CA
HILL, PD
THOMPSON, FD
机构
[1] MIDDLESEX HOSP,LONDON W1N 8AA,ENGLAND
[2] UNIV LONDON KINGS COLL HOSP,LONDON SE5 8RX,ENGLAND
关键词
BIOELECTRICAL IMPEDANCE; BODY COMPOSITION; DIALYSIS; TOTAL BODY WATER;
D O I
10.1093/ndt/8.8.716
中图分类号
R3 [基础医学]; R4 [临床医学];
学科分类号
1001 ; 1002 ; 100602 ;
摘要
The measurement of total body water by bioelectrical impedance in a group of renal patients was evaluated against the tritium dilution method. The effect of haemodialysis and the presence of peritoneal dialysate on the impedance were also investigated. The correlation between the two methods is r=0.90 with a residual standard deviation of 3.7. The standard deviation of the differences between the two methods against the means was 3.66 which means that total body water (TBW) estimated by the bioelectrical impedance (BEI) method may be 6.181 (X +/- 2 SD) above or 8.381 below the (H2O)-H-3 method. The BEI method overestimated the actual weight loss after haemodialysis (3.87 +/- 1.71 versus 2.43 +/- 1.81) but underestimated the volume of peritoneal dialysate in situ. The BEI method would not be appropriate for use in assessing total body water and monitoring acute volume changes in patients with renal failure who are on strict fluid restriction.
引用
收藏
页码:716 / 719
页数:4