Targeting accuracy at couch kick for a frameless image guided radiosurgery system

被引:1
|
作者
Huang, Yimei [1 ]
Zhao, Bo [1 ]
Kim, Joshua [1 ]
Wen, Ning [1 ]
Chetty, Indrin J. [1 ]
Siddiqui, Salim [1 ]
机构
[1] Henry Ford Hlth Syst, Dept Radiat Oncol, 2799 W Grand Blvd, Detroit, MI 48202 USA
来源
JOURNAL OF RADIOSURGERY AND SBRT | 2018年 / 5卷 / 02期
关键词
frameless radiosurgery; trigeminal neuralgia; image-guided radiosurgery;
D O I
暂无
中图分类号
R61 [外科手术学];
学科分类号
摘要
Purpose/Objectives: Targeting accuracy at all possible couch angles needs to be carefully evaluated prior to initiating a frameless image-guided stereotactic radiosurgery program on a Linac for treating functional disorders such as trigeminal neuralgia. In this study, we report positioning accuracy with stereoscopic x-ray imaging over the complete range of couch rotation using anthropomorphic head phantoms. Materials/Methods: An anthropomorphic head phantom with three 5 mm tungsten BBs as hidden targets was CT simulated. A group of 7 arcs was planned with couch angles from 0 to 90 degrees in 15 degrees increments. A pair of stereoscopic x-ray images that auto-matches to planning CT according to bony anatomy was utilized to position each BB to machine isocenter at all planned couch angles. Targeting accuracy was measured by stereoscopic x-ray imaging of the BB itself, which provides the distance from the centroid of BB to the x-ray imaging isocenter. For each BB, the hidden target test was repeated 5 times at couch 0 degrees and 3 times at other couch angles, resulting in a total of 69 measurements, each with random initial setup deviation. Following the same workflow, a second anthropomorphic head phantom with two 5 mm BBs was utilized to evaluate localization accuracy at couch angles of 0 through 270 degrees in 15 degrees increments, resulting in another 18 measurements. Results: Residual setup deviation following image guidance in the first head phantom was 0.6 +/- 0.1, 0.4 +/- 0.1, and 0.4 +/- 0.1 mm, respectively, at the three BBs, and 0.5 +/- 0.1 mm overall (N=69). Comparable results of 0.4 +/- 0.1 mm (N=18) were achieved with the second head phantom. Combining results from both phantoms, the targeting accuracy was 0.5 +/- 0.1 mm (range 0.2-0.8 mm). No apparent correlation was observed between targeting accuracy and couch rotation. Conclusions: Accurate positioning within 1 mm can be achieved with stereoscopic x-ray imaging at any couch angle. Frameless image-guided stereotactic radiosurgery could achieve targeting accuracy similar to that of frame-based systems for high dose treatment of trigeminal neuralgia.
引用
收藏
页码:123 / 129
页数:7
相关论文
共 50 条
  • [41] Geometric Accuracy of Frameless Based Image Guided Stereotactic Radiosurgery of Trigeminal Neuralgia using Brainlab's Exactrac 6-D Robotic System
    Rahimian, J.
    Girvigian, M. R.
    Chen, J. C. T.
    Rahimian, R.
    Miller, M. J.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 75 (03): : S676 - S676
  • [42] THE NEUROSTATION SYSTEM FOR IMAGE-GUIDED, FRAMELESS STEREOTAXY
    GERMANO, IM
    NEUROSURGERY, 1995, 37 (02) : 348 - 350
  • [43] The Impact of Imaging Frequency on Intrafraction Setup Corrections in Frameless Image Guided Stereotactic Radiosurgery
    Westerly, D.
    Stuhr, K.
    Altunbas, C.
    Diot, Q.
    Kavanagh, B.
    Miften, M.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [44] LINAC Frameless Image-Guided Radiosurgery for Initial Treatment of Typical Trigeminal Neuralgia
    Kim, Dong Gyu
    WORLD NEUROSURGERY, 2010, 74 (4-5) : 451 - 451
  • [45] FRAMELESS IMAGE-GUIDED INTRACRANIAL STEREOTACTIC RADIOSURGERY: CLINICAL OUTCOMES FOR BRAIN METASTASES
    Breneman, John C.
    Steinmetz, Ryan
    Smith, Aaron
    Lamba, Michael
    Warnick, Ronald E.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 74 (03): : 702 - 706
  • [46] Dosimetric Analysis of Flat versus Unflat Beams for Frameless Image guided Trigeminal Radiosurgery
    Bhaskar, V. N.
    Shankar, V.
    Haritha, C.
    Karuppusamy, A.
    Bhange, A.
    Maurya, S.
    Kumar, B. Ajai
    Sen, D.
    Mutha, S.
    Purohit, R.
    Chigurupalli, K.
    Vyas, H.
    Deepanjali, P.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2018, 102 (03): : E508 - E508
  • [47] Feasibility, safety, and outcome of frameless image-guided robotic radiosurgery for brain metastases
    Muacevic, Alexander
    Kufeld, Markus
    Wowra, Berndt
    Kreth, Friedrich-Wilhelm
    Tonn, Joerg-Christian
    JOURNAL OF NEURO-ONCOLOGY, 2010, 97 (02) : 267 - 274
  • [48] Feasibility, safety, and outcome of frameless image-guided robotic radiosurgery for brain metastases
    Alexander Muacevic
    Markus Kufeld
    Berndt Wowra
    Friedrich-Wilhelm Kreth
    Jörg-Christian Tonn
    Journal of Neuro-Oncology, 2010, 97 : 267 - 274
  • [49] Targeting accuracy of an image guided gating system for stereotactic body radiotherapy
    Tenn, SE
    Solberg, TD
    Medin, PM
    PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (23): : 5443 - 5462
  • [50] Factors affecting the accuracy of respiratory tracking of the image-guided robotic radiosurgery system
    Mitsuhiro Inoue
    Kohei Okawa
    Junichi Taguchi
    Yoshifumi Hirota
    Yohei Yanagiya
    Chie Kikuchi
    Michio Iwabuchi
    Taro Murai
    Hiromitsu Iwata
    Hiroya Shiomi
    Izumi Koike
    Koshi Tatewaki
    Seiji Ohta
    Japanese Journal of Radiology, 2019, 37 : 727 - 734