Targeting accuracy at couch kick for a frameless image guided radiosurgery system

被引:1
|
作者
Huang, Yimei [1 ]
Zhao, Bo [1 ]
Kim, Joshua [1 ]
Wen, Ning [1 ]
Chetty, Indrin J. [1 ]
Siddiqui, Salim [1 ]
机构
[1] Henry Ford Hlth Syst, Dept Radiat Oncol, 2799 W Grand Blvd, Detroit, MI 48202 USA
来源
JOURNAL OF RADIOSURGERY AND SBRT | 2018年 / 5卷 / 02期
关键词
frameless radiosurgery; trigeminal neuralgia; image-guided radiosurgery;
D O I
暂无
中图分类号
R61 [外科手术学];
学科分类号
摘要
Purpose/Objectives: Targeting accuracy at all possible couch angles needs to be carefully evaluated prior to initiating a frameless image-guided stereotactic radiosurgery program on a Linac for treating functional disorders such as trigeminal neuralgia. In this study, we report positioning accuracy with stereoscopic x-ray imaging over the complete range of couch rotation using anthropomorphic head phantoms. Materials/Methods: An anthropomorphic head phantom with three 5 mm tungsten BBs as hidden targets was CT simulated. A group of 7 arcs was planned with couch angles from 0 to 90 degrees in 15 degrees increments. A pair of stereoscopic x-ray images that auto-matches to planning CT according to bony anatomy was utilized to position each BB to machine isocenter at all planned couch angles. Targeting accuracy was measured by stereoscopic x-ray imaging of the BB itself, which provides the distance from the centroid of BB to the x-ray imaging isocenter. For each BB, the hidden target test was repeated 5 times at couch 0 degrees and 3 times at other couch angles, resulting in a total of 69 measurements, each with random initial setup deviation. Following the same workflow, a second anthropomorphic head phantom with two 5 mm BBs was utilized to evaluate localization accuracy at couch angles of 0 through 270 degrees in 15 degrees increments, resulting in another 18 measurements. Results: Residual setup deviation following image guidance in the first head phantom was 0.6 +/- 0.1, 0.4 +/- 0.1, and 0.4 +/- 0.1 mm, respectively, at the three BBs, and 0.5 +/- 0.1 mm overall (N=69). Comparable results of 0.4 +/- 0.1 mm (N=18) were achieved with the second head phantom. Combining results from both phantoms, the targeting accuracy was 0.5 +/- 0.1 mm (range 0.2-0.8 mm). No apparent correlation was observed between targeting accuracy and couch rotation. Conclusions: Accurate positioning within 1 mm can be achieved with stereoscopic x-ray imaging at any couch angle. Frameless image-guided stereotactic radiosurgery could achieve targeting accuracy similar to that of frame-based systems for high dose treatment of trigeminal neuralgia.
引用
收藏
页码:123 / 129
页数:7
相关论文
共 50 条
  • [21] EVALUATION OF IMAGE-GUIDED POSITIONING FOR FRAMELESS INTRACRANIAL RADIOSURGERY
    Lamba, Michael
    Breneman, John C.
    Warnick, Ronald E.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 74 (03): : 913 - 919
  • [22] Measurement of total clinical accuracy for an image guided radiosurgery system
    Wu, X
    Main, W
    Schwade, J
    MEDICAL PHYSICS, 2004, 31 (06) : 1901 - 1901
  • [23] Accuracy of remote translations of the couch during IGRT-guided radiosurgery
    Pozo, M.
    Calvo Ortega, J. F.
    Moragues Femenia, S.
    Casals Farran, J.
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S256 - S256
  • [24] 3D Image Guided Evaluation of Setup Accuracy for Patients undergoing Frameless Spine Stereotactic Radiosurgery
    Ionascu, D.
    Martin, S.
    Wloch, J.
    Matuszak, M.
    Grills, I. S.
    Yan, D.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 75 (03): : S572 - S573
  • [25] Novalis frameless image-guided noninvasive radiosurgery: Initial experience
    Wurm, Reinhard E.
    Erbel, Stephan
    Schwenkert, Isabel
    Gum, Franz
    Agaoglu, Daniel
    Schild, Reinhard
    Schlenger, Lorenz
    Scheffler, Dirk
    Brock, Mario
    Budach, Volker
    NEUROSURGERY, 2008, 62 (05) : A11 - A17
  • [26] Control of brain metastases using frameless image-guided radiosurgery
    Chen, Joseph C. T.
    Bugoci, Darlene M.
    Girvigian, Michael R.
    Miller, Michael J.
    Arellano, Alonso
    Rahimian, Javad
    NEUROSURGICAL FOCUS, 2009, 27 (06) : E6.1 - E6.7
  • [27] Patterns of patient movement during frameless image-guided radiosurgery
    Murphy, MJ
    Chang, SD
    Gibbs, IC
    Le, QT
    Hai, J
    Kim, D
    Martin, DP
    Adler, JR
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 55 (05): : 1400 - 1408
  • [28] Frameless LINAC Image Guided Radiosurgery for Primary and Recurrent Trigeminal Neuralgia
    Sckolnik, S. E.
    Jeffrey, K. M.
    Stea, B.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 90 : S683 - S683
  • [29] Accuracy Treatments With Frameless Stereotactic Radiosurgery
    Olivera Vegas, J.
    Rincon, M.
    Vasquez, W.
    Garcia, M. A.
    Perez, A. M.
    Vara, J.
    Luna, J.
    Prieto, I.
    Gomez Tejedor, S.
    Penedo, J. M.
    Esteban, D.
    Ilundain, A.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2016, 96 (02): : E77 - E77
  • [30] Accuracy of a patient positioning system for image-guided frameless stereotactic IMRT treatments
    Dohm, O.
    Hoenes, A.
    Paulsen, F.
    Alber, M.
    RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S288 - S288